Performance Parameters of Inductively Coupled Plasma Optical Emission Spectrometry and Graphite Furnace Atomic Absorption Spectrometry Techniques for Pd and Pt Determination in Automotive Catalysts

被引:21
|
作者
Senila, Marin [1 ]
Cadar, Oana [1 ]
Senila, Lacrimioara [1 ]
Boeringer, Sarah [2 ]
Seaudeau-Pirouley, Karine [3 ]
Ruiu, Andrea [4 ]
Lacroix-Desmazes, Patrick [4 ]
机构
[1] Natl Inst Res & Dev Optoelect Bucharest, Res Inst Analyt Instrumentat, Cluj Napoca 400293, Romania
[2] Fraunhofer Inst Chem Technol ICT, Dept Environm Engn, Grp React & Separat Tech, D-76327 Pfinztal, Germany
[3] Innovat Fluides Supercrit IFS, Batiment INEED,BP16109, F-26300 Alixan, France
[4] Univ Montpellier, CNRS, ENSCM, Inst Charles Gerhardt Montpellier ICGM, F-34095 Montpellier, France
基金
欧盟地平线“2020”;
关键词
palladium; platinum; ICP-OES; catalysts; method validation; microwave digestion; PLATINUM-GROUP METALS; ONLINE SEPARATION; PALLADIUM; PRECONCENTRATION; VALIDATION; RECOVERY; RHODIUM; DIGESTION; TRACES; FOODS;
D O I
10.3390/ma13225136
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
Palladium (Pd) and platinum (Pt) are extensively used as catalysts in the petrochemical and automotive industries, and due to high demand for them on the market, their recycling from spent supported catalysts is clearly needed. To assess the content of Pd and Pt in catalysts in order to establish their commercial value or to evaluate the recovery efficiency of technologies used for recycling, reliable analytical methods for determination of these elements are required. Spectrometric methods, such as inductively coupled plasma optical emission spectrometry (ICP-OES) and graphite furnace atomic absorption spectrometry (GFAAS) are powerful tools that can be employed for the determination of Pd and Pt in various sample matrices. However, these methods allow only the injection of liquid samples. In this regard, the digestion of solid sample by microwave-assisted acid extraction procedures at high pressures and temperatures is often used. In this study, a microwave acid digestion method was optimized for the extraction of Pd and Pt from spent catalysts, using a four-step program, at a maximum 200 degrees C. The resulting solutions were analyzed using ICP-OES, at two different wavelengths for each metal (Pd at 340.458 and 363.470 nm, and Pt at 265.945 and 214.423 nm, respectively) and using GFAAS (Pd at 247.64 nm, Pt at 265.94 nm). Five types of spent catalyst were analyzed and the standard deviations of repeatability for five parallel samples were less than predicted relative standard deviations (PRSD%) calculated using Horvitz's equation for all the analyzed samples.
引用
收藏
页码:1 / 13
页数:13
相关论文
共 50 条
  • [1] PRECONCENTRATION OF PALLADIUM, PLATINUM AND RHODIUM BY ONLINE SORBENT EXTRACTION FOR GRAPHITE-FURNACE ATOMIC-ABSORPTION SPECTROMETRY AND INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY
    LEE, ML
    TOLG, G
    BEINROHR, E
    TSCHOPEL, P
    ANALYTICA CHIMICA ACTA, 1993, 272 (02) : 193 - 203
  • [2] Preconcentration of platinum by online sorption for graphite furnace atomic absorption spectrometry and inductively coupled plasma atomic emission spectrometry:: A comparative study
    Cano Pavon, J. M.
    de Torres, A. Garcia
    Rojas, F. Sanchez
    Ojeda, C. Bosch
    SPECTROSCOPY LETTERS, 2007, 40 (01) : 27 - 39
  • [3] DETERMINATION OF MACRONUTRIENTS IN SPICES BY INDUCTIVELY COUPLED PLASMA-OPTICAL EMISSION SPECTROMETRY
    Khan, Naeem
    Choi, Ji Yeon
    Nho, Eun Yeong
    Habte, Girum
    Jamila, Nargis
    Hong, Joon Ho
    Ryu, Keun Yeoung
    Park, Kyung Su
    Kim, Kyong Su
    ANALYTICAL LETTERS, 2014, 47 (14) : 2394 - 2405
  • [4] Direct precise determination of Pd, Pt and Rh in spent automobile catalysts solution by high-resolution continuum source graphite furnace atomic absorption spectrometry
    Eskina, Vasilina V.
    Dalnova, Olga A.
    Filatova, Daria G.
    Baranovskaya, Vasilisa B.
    Karpov, Yuri A.
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2020, 165 (165)
  • [5] SELENIUM DETERMINATION IN PLASMA/SERUM BY INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY (ICP-MS): COMPARISON WITH GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROMETRY (GF-AAS)
    Janasik, Beata
    Trzcinka-Ochocka, Malgorzata
    Brodzka, Renata
    MEDYCYNA PRACY, 2011, 62 (05) : 489 - 498
  • [6] Cloud point extraction for the determination of trace amounts of Pt(IV) by graphite furnace atomic absorption spectrometry
    Mortada, Wael I.
    Hassanien, Mohammed M.
    El-Asmy, Ahmed A.
    ANALYTICAL METHODS, 2013, 5 (02) : 530 - 535
  • [7] Determination of Pt, Pd and Rh in Brassica Napus using solid sampling electrothermal vaporization inductively coupled plasma optical emission spectrometry
    Nischkauer, Winfried
    Herincs, Esther
    Puschenreiter, Markus
    Wenzel, Walter
    Limbeck, Andreas
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2013, 89 : 60 - 65
  • [8] Preconcentration of trace Pd(II) on crosslinked chitosan and determination by graphite furnace atomic absorption spectrometry
    Qian Sha-hua
    Xiang Luo-jing
    Deng Hong-bing
    Xiao Mei
    Lin Han
    Li Xue-qin
    SPECTROSCOPY AND SPECTRAL ANALYSIS, 2007, 27 (03) : 592 - 594
  • [9] Solid Phase Extraction and Graphite furnace Atomic Absorption Spectrometry for the Determination of Palladium
    Huang, Zhangjie
    Chang, Jun
    Feng, Shaoping
    Yang, Zhongqi
    ASIAN JOURNAL OF CHEMISTRY, 2010, 22 (03) : 2018 - 2026
  • [10] Determination of Chlorine in Gasoline by Inductively Coupled Plasma Atomic Emission Spectrometry
    Zhao Yan
    Chen Xiao-yan
    Xu Dong-yu
    Zhang Shi-yuan
    Chen Ze-yong