Does Use of a Powered Ankle-foot Prosthesis Restore Whole-body Angular Momentum During Walking at Different Speeds?

被引:34
作者
D'Andrea, Susan [1 ]
Wilhelm, Natalie [1 ]
Silverman, Anne K. [2 ]
Grabowski, Alena M. [3 ,4 ]
机构
[1] Providence VA Med Ctr, Dept Vet Affairs, Providence, RI USA
[2] Colorado Sch Mines, Golden, CO 80401 USA
[3] Univ Colorado, Boulder, CO 80309 USA
[4] Eastern Colorado Healthcare Syst, Dept Vet Affairs, Denver, CO USA
关键词
GAIT;
D O I
10.1007/s11999-014-3647-1
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Whole-body angular momentum (H) influences fall risk, is tightly regulated during walking, and is primarily controlled by muscle force generation. People with transtibial amputations using passive-elastic prostheses typically have greater H compared with nonamputees. (1) Do people with unilateral transtibial amputations using passive-elastic prostheses have greater sagittal and frontal plane H ranges of motion during walking compared with nonamputees and compared with using powered prostheses? (2) Does use of powered ankle-foot prostheses result in equivalent H ranges in all planes of motion compared with nonamputees during walking as a result of normative prosthetic ankle power generation? Eight patients with a unilateral transtibial amputation and eight nonamputees walked 0.75, 1.00, 1.25, 1.50, and 1.75 m/s while we measured kinematics and ground reaction forces. We calculated H for participants using their passive-elastic prosthesis and a powered ankle-foot prosthesis and for nonamputees at each speed. Patients using passive-elastic prostheses had 32% to 59% greater sagittal H ranges during the affected leg stance phase compared with nonamputees at 1.00 to 1.75 m/s (p < 0.05). Patients using passive-elastic prostheses had 5% and 9% greater sagittal H ranges compared with using powered prostheses at 1.25 and 1.50 m/s, respectively (p < 0.05). Participants using passive-elastic prostheses had 29% and 17% greater frontal H ranges at 0.75 and 1.50 m/s, respectively, compared with nonamputees (p < 0.05). Surprisingly, patients using powered prostheses had 26% to 50% greater sagittal H ranges during the affected leg stance phase compared with nonamputees at 1.00 to 1.75 m/s (p < 0.05). Patients using powered prostheses also had 26% greater frontal H range compared with nonamputees at 0.75 m/s (p < 0.05). People with a transtibial amputation may more effectively regulate H at two specific walking speeds when using powered compared with passive-elastic prostheses. Our results support the hypothesis that an ankle-foot prosthesis capable of providing net positive work during the stance phase of walking reduces sagittal plane H; future studies are needed to validate our biomechanical findings with larger numbers of patients and should determine whether powered prostheses can decrease the risk of falls in patients with a transtibial amputation.
引用
收藏
页码:3044 / 3054
页数:11
相关论文
共 25 条
  • [1] [Anonymous], 2002, SAGITTAL PLANE CHARA
  • [2] Au SK, 2008, IEEE ROBOT AUTOM MAG, V15, P52, DOI 10.1109/MRA.2008.927697
  • [3] Powered Ankle-Foot Prosthesis Improves Walking Metabolic Economy
    Au, Samuel K.
    Weber, Jeff
    Herr, Hugh
    [J]. IEEE TRANSACTIONS ON ROBOTICS, 2009, 25 (01) : 51 - 66
  • [4] Angular momentum of walking at different speeds
    Bennett, Bradford C.
    Russell, Shawn D.
    Sheth, Pradip
    Abel, Mark F.
    [J]. HUMAN MOVEMENT SCIENCE, 2010, 29 (01) : 114 - 124
  • [5] DEMPSTER P, 1995, MED SCI SPORT EXER, V27, P1692
  • [6] Control of a Powered Ankle-Foot Prosthesis Based on a Neuromuscular Model
    Eilenberg, Michael F.
    Geyer, Hartmut
    Herr, Hugh
    [J]. IEEE TRANSACTIONS ON NEURAL SYSTEMS AND REHABILITATION ENGINEERING, 2010, 18 (02) : 164 - 173
  • [7] Elftman H., 1939, Human Biology, V11, P529
  • [8] Angular momentum in human walking
    Herr, Hugh
    Popovic, Marko
    [J]. JOURNAL OF EXPERIMENTAL BIOLOGY, 2008, 211 (04) : 467 - 481
  • [9] Bionic ankle-foot prosthesis normalizes walking gait for persons with leg amputation
    Herr, Hugh M.
    Grabowski, Alena M.
    [J]. PROCEEDINGS OF THE ROYAL SOCIETY B-BIOLOGICAL SCIENCES, 2012, 279 (1728) : 457 - 464
  • [10] REPEATABILITY OF KINEMATIC, KINETIC, AND ELECTROMYOGRAPHIC DATA IN NORMAL ADULT GAIT
    KADABA, MP
    RAMAKRISHNAN, HK
    WOOTTEN, ME
    GAINEY, J
    GORTON, G
    COCHRAN, GVB
    [J]. JOURNAL OF ORTHOPAEDIC RESEARCH, 1989, 7 (06) : 849 - 860