A biomechanical evaluation of methods of distal humerus fracture fixation using locking compression plates versus conventional reconstruction plates

被引:187
作者
Korner, J
Diederichs, G
Arzdorf, M
Lill, H
Josten, C
Schneider, E
Linke, B
机构
[1] Johannes Gutenberg Univ Mainz, Dept Trauma & Reconstruct Surg, D-55101 Mainz, Germany
[2] AO Res Inst, Davos, Switzerland
[3] Univ Leipzig, Dept Trauma & Reconstruct Surg, Leipzig, Germany
关键词
distal humerus; osteosynthesis; fracture; biomechanics; LCP;
D O I
10.1097/00005131-200405000-00004
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objectives: To examine the biomechanical behavior of 2 techniques of double-plate osteosynthesis for fractures of the adult distal humerus using conventional reconstruction plates and locking compression plates. Design: Basic science Study. Setting: Experimental in vitro study. Patients/Participants: Forty fresh-frozen human distal humeri specimens. Intervention: Four matched groups with 10 humeri each, median age 74 years (46-95), were created using similar bone mineral density values. Two standard configurations of double-plate osteosynthesis (dorsal or 90degrees configuration) with either conventional reconstruction plates or locking compression plates were studied for biomechanical properties of the constructs. A fracture model with a 5-mm supracondylar osteotomy gap simulating metaphyseal comminution (AO type 13-A3.3) was used. Main Outcome Measurement: Stiffness testing of the constructs in anterior/posterior bending, torsion, and axial compression loading. Evaluation of alterations of the bone-implant interface and failure patterns under cyclic loading and strength testing. Results: The study demonstrates that primary stiffness in anterior/posterior bending and torsional loading is significantly increased by using locking compression plates in a 90degrees configuration (P < 0.05) as compared with dorsally applied plates. The differences between the different plate types are insignificant if applied in the same configuration. It is demonstrated that none of the tested implants failed under cyclic loading within the number of cycles expected for 3 months of use. The bone-implant interface is less likely to fail during strength testing with locking compression plates. Conclusion: The biomechanical behavior of the osteosynthesis depends more on plate configuration than plate type. Advantages of locking compression plates are only significant if compared with dorsal plate application techniques. Nevertheless, locking compression plates are helpful supplementary tools for achieving primary stable fracture fixation. This might be of considerable clinical relevance in patients with diminished bone mineral quality or in the presence of metaphyseal comminution.
引用
收藏
页码:286 / 293
页数:8
相关论文
共 50 条
  • [1] PHYSIOLOGICAL CONSIDERATIONS OF MUSCLE FORCE THROUGH THE ELBOW JOINT
    AN, KN
    KAUFMAN, KR
    CHAO, EYS
    [J]. JOURNAL OF BIOMECHANICS, 1989, 22 (11-12) : 1249 - 1256
  • [2] AN KN, 1984, J BIOMECH ENG-T ASME, V106, P364, DOI 10.1115/1.3138507
  • [3] BAUER GJ, 1997, AKTUEL TRAUMATOL, V27, P125
  • [4] ROLE OF MECHANICAL LOADING IN THE PROGRESSIVE OSSIFICATION OF A FRACTURE CALLUS
    BLENMAN, PR
    CARTER, DR
    BEAUPRE, GS
    [J]. JOURNAL OF ORTHOPAEDIC RESEARCH, 1989, 7 (03) : 398 - 407
  • [5] Boyer Martin I, 2003, Instr Course Lect, V52, P591
  • [6] Corcaci R, 2002, Rev Med Chir Soc Med Nat Iasi, V106, P178
  • [7] A BIOMECHANICAL ANALYSIS OF PROPHYLACTIC FIXATION FOR PATHOLOGICAL FRACTURES OF THE DISTAL 3RD OF THE HUMERUS
    DAMRON, TA
    HEINER, JP
    FREUND, EMN
    DAMRON, LA
    MCCABE, R
    VANDERBY, R
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1994, 76A (06) : 839 - 847
  • [8] Fornasieri C, 1997, REV CHIR ORTHOP, V83, P237
  • [9] Frigg R, 2001, INJURY, V32, pSB63
  • [10] Total elbow replacement for complex fractures of the distal humerus - An option for the elderly patient
    Gambirasio, R
    Riand, N
    Stern, R
    Hoffmeyer, P
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2001, 83B (07): : 974 - 978