The Measurement of the Air-Kerma Rate in Air and a Solid Phantom with Ionization Chambers for a 192Ir HDR Brachytherapy Source

被引:0
作者
Zeng, Jing [1 ]
Qu, Pengpeng [1 ]
Pang, Qingsong [2 ]
Wang, Ping [2 ]
机构
[1] Nankai Univ, Tianjin Cent Hosp Gynecol & Obstet, Dept Gynecol Oncol, Affiliated Hosp, Tianjin 300100, Peoples R China
[2] Tianjin Med Univ Canc Inst & Hosp, Natl Clin Res Ctr Canc, Tianjins Clin Res Ctr Canc, Key Lab Canc Prevent & Therapy, Tianjin 300060, Peoples R China
来源
CANCER MANAGEMENT AND RESEARCH | 2020年 / 12卷
关键词
brachytherapy; Ir-192; source; calibration; ionization chamber; TREATMENT PLANNING SYSTEMS; PROTOCOL;
D O I
10.2147/CMAR.S275378
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Introduction: This study aims to measure the air-kerma rate of 192-Ir-HDR-afterloading source with an ionization chamber in air and a solid cylindrical phantom separately and to compare the dose calibration by the American Association of Physicists in Medicine (AAPM) Task Group TG-43U1 formalism with the Abacus treatment planning system (TPS). Materials and Methods: The air-kerma rate of Ir-192 source was measured by an ionization chamber in air and a solid cylindrical phantom separately. For the interesting point position P (8cm, 90 degrees), the values of the dose were calculated with the TG-43U1 formula and compared with data from the Abacus TPS with single and multiple dwell positions, respectively. Results: The air-kerma rate percentage deviations between the detector measurements in air and the source certificate were -1.28%, -0.91%, -0.71%, and 0.33% at the distances of 25cm, 50cm, 75cm, and 100cm, respectively. For the measurement in solid cylindrical phantom, the percent deviation from the air-kerma rate certificate was 1.85%. The percentage deviations of the dose calibration between Abacus TPS and TG-43U1 formalism at P (8cm, 90 degrees) were -2.30%, 1.76%, and 2.10% with different distances (between the dwell positions) of 0cm, 0.5cm, and 1cm, respectively. Conclusion: The in-air technique was a new attempt for clinic routine measurement. Further studies are still necessary. As a treatment planning system, the Abacus TPS should apply the AAPM TG-43U1 formulism for the development required in the future.
引用
收藏
页码:10821 / 10828
页数:8
相关论文
共 50 条
[31]   EVALUATION OF SCATTER CONTRIBUTION AND DISTANCE ERROR BY ITERATIVE METHODS FOR STRENGTH DETERMINATION OF HDR 192IR BRACHYTHERAPY SOURCE [J].
Kumar, Sudhir ;
Srinivasan, Panchapakesan ;
Sharma, Sunil D. ;
Subbaiah, Kamatam V. ;
Mayya, Yelia S. .
MEDICAL DOSIMETRY, 2010, 35 (03) :230-237
[32]   Direct measurement of absorbed dose to water in HDR 192Ir brachytherapy: Water calorimetry, ionization chamber, Gafchromic film, and TG-43 [J].
Sarfehnia, Arman ;
Kawrakow, Iwan ;
Seuntjens, Jan .
MEDICAL PHYSICS, 2010, 37 (04) :1924-1932
[33]   Measurement of rectum dose by in vivo alanine/ESR dosimetry in gynecological 192Ir HDR brachytherapy [J].
Bravo-Miranda, Carlos ;
Rech, Amanda Burg ;
Oliveira, Harley Francisco ;
Bertucci, Edenyse ;
Cordova-Fraga, Teodoro ;
Baffa, Oswaldo .
RADIATION MEASUREMENTS, 2015, 75 :45-52
[34]   Film dosimetry calibration method for pulsed-dose-rate brachytherapy with an 192Ir source [J].
Schwob, Nathan ;
Orion, Itzhak .
MEDICAL PHYSICS, 2007, 34 (05) :1678-1683
[35]   Well-ionization chamber response relative to NIST air-kerma strength standard for prostate brachytherapy seeds [J].
Mitch, MG ;
Zimmerman, BE ;
Lamperti, PJ ;
Seltzer, SM ;
Coursey, BM .
MEDICAL PHYSICS, 2000, 27 (10) :2293-2296
[36]   Application of a Twin-Detector Method for the Determination of the Mean Photon Energy Em at Points of Measurement in a Water Phantom Surrounding a GammaMed HDR 192Ir Brachytherapy Source [J].
Chofor, N. ;
Nebah, F. ;
Harder, D. ;
Poppe, B. .
MEDICAL PHYSICS, 2014, 41 (06) :232-232
[37]   Corrections for photons attenuation and scattering for the extrapolation chamber for measurement of air-kerma strength of 125I brachytherapy source [J].
Liu, C. ;
Song, M. ;
Liu, Y. ;
Wang, H. ;
Wei, K. ;
Hang, Z. .
JOURNAL OF INSTRUMENTATION, 2021, 16 (03)
[38]   An innovative method for 192Ir HDR calibration by farmer chamber, V-film, and solid phantom [J].
Chang, Liyun ;
Ding, Hueisch-Jy ;
Ho, Sheng-Yow .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2011, 646 (01) :192-196
[39]   Dosimetric characteristics of the 192Ir high-dose-rate afterloading brachytherapy source [J].
Sadeghi, Mahdi ;
Taghdiri, Fatemah ;
Saidi, Pooneh .
JAPANESE JOURNAL OF RADIOLOGY, 2011, 29 (05) :324-329
[40]   Dosimetric characteristics of the 192Ir high-dose-rate afterloading brachytherapy source [J].
Mahdi Sadeghi ;
Fatemah Taghdiri ;
Pooneh Saidi .
Japanese Journal of Radiology, 2011, 29 :324-329