The Measurement of the Air-Kerma Rate in Air and a Solid Phantom with Ionization Chambers for a 192Ir HDR Brachytherapy Source

被引:0
作者
Zeng, Jing [1 ]
Qu, Pengpeng [1 ]
Pang, Qingsong [2 ]
Wang, Ping [2 ]
机构
[1] Nankai Univ, Tianjin Cent Hosp Gynecol & Obstet, Dept Gynecol Oncol, Affiliated Hosp, Tianjin 300100, Peoples R China
[2] Tianjin Med Univ Canc Inst & Hosp, Natl Clin Res Ctr Canc, Tianjins Clin Res Ctr Canc, Key Lab Canc Prevent & Therapy, Tianjin 300060, Peoples R China
来源
CANCER MANAGEMENT AND RESEARCH | 2020年 / 12卷
关键词
brachytherapy; Ir-192; source; calibration; ionization chamber; TREATMENT PLANNING SYSTEMS; PROTOCOL;
D O I
10.2147/CMAR.S275378
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Introduction: This study aims to measure the air-kerma rate of 192-Ir-HDR-afterloading source with an ionization chamber in air and a solid cylindrical phantom separately and to compare the dose calibration by the American Association of Physicists in Medicine (AAPM) Task Group TG-43U1 formalism with the Abacus treatment planning system (TPS). Materials and Methods: The air-kerma rate of Ir-192 source was measured by an ionization chamber in air and a solid cylindrical phantom separately. For the interesting point position P (8cm, 90 degrees), the values of the dose were calculated with the TG-43U1 formula and compared with data from the Abacus TPS with single and multiple dwell positions, respectively. Results: The air-kerma rate percentage deviations between the detector measurements in air and the source certificate were -1.28%, -0.91%, -0.71%, and 0.33% at the distances of 25cm, 50cm, 75cm, and 100cm, respectively. For the measurement in solid cylindrical phantom, the percent deviation from the air-kerma rate certificate was 1.85%. The percentage deviations of the dose calibration between Abacus TPS and TG-43U1 formalism at P (8cm, 90 degrees) were -2.30%, 1.76%, and 2.10% with different distances (between the dwell positions) of 0cm, 0.5cm, and 1cm, respectively. Conclusion: The in-air technique was a new attempt for clinic routine measurement. Further studies are still necessary. As a treatment planning system, the Abacus TPS should apply the AAPM TG-43U1 formulism for the development required in the future.
引用
收藏
页码:10821 / 10828
页数:8
相关论文
共 50 条
  • [21] Ir-192 Calibration in Air with Farmer Chamber for HDR Brachytherapy
    Chang, Liyun
    Ho, Sheng-Yow
    Lee, Tsair-Fwu
    Ding, Hueisch-Jy
    Chen, Pang-Yu
    JOURNAL OF MEDICAL AND BIOLOGICAL ENGINEERING, 2016, 36 (02) : 145 - 152
  • [22] A technique for calibrating a high dose rate 192Ir brachytherapy source
    Fourie, O. L.
    Crabtree, T. G.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2012, 35 (01) : 85 - 92
  • [23] Experimental derivation of wall correction factors for ionization chambers used in high dose rate 192Ir source calibration
    Maréchal, MH
    de Almeida, CE
    Ferreira, IH
    Sibata, CH
    MEDICAL PHYSICS, 2002, 29 (01) : 1 - 5
  • [24] Simulation evaluation of NIST air-kerma rate calibration standard for electronic brachytherapy
    Hiatt, Jessica R.
    Rivard, Mark J.
    Hughes, H. Grady
    MEDICAL PHYSICS, 2016, 43 (03) : 1119 - 1129
  • [25] The non-uniformity correction factor for the cylindrical ionization chambers in dosimetry of an HDR Ir-192 brachytherapy source
    Majumdar, Bishnu
    Patel, Narayan Prasad
    Vijayan, V.
    JOURNAL OF MEDICAL PHYSICS, 2006, 31 (02) : 83 - 88
  • [26] Reference air kerma rate calibration system for high dose rate Ir-192 brachytherapy sources in Taiwan
    Chu, Wei-Han
    Yuan, Ming-Chen
    Lee, Jeng-Hung
    Lin, Yi-Chun
    RADIATION PHYSICS AND CHEMISTRY, 2017, 140 : 361 - 364
  • [27] High-Dose-Rate 192Ir Brachytherapy Dose Verification: A Phantom Study
    Nikoofar, Alireza
    Hoseinpour, Zohreh
    Mahdavi, Seied Rabi
    Hasanzadeh, Hadi
    Tavirani, Mostafa Rezaei
    IRANIAN JOURNAL OF CANCER PREVENTION, 2015, 8 (03)
  • [28] Enhancement and validation of Geant4 Brachytherapy application on clinical HDR 192Ir source
    Ababneh, Eshraq
    Dababneh, Saed
    Qatarneh, Sharif
    Wadi-Ramahi, Shada
    RADIATION PHYSICS AND CHEMISTRY, 2014, 103 : 57 - 66
  • [29] Dosimetry revisited for the HDR 192Ir brachytherapy source model mHDR-v2
    Granero, Domingo
    Vijande, Javier
    Ballester, Facundo
    Rivard, Mark J.
    MEDICAL PHYSICS, 2011, 38 (01) : 487 - 494
  • [30] EVALUATION OF SCATTER CONTRIBUTION AND DISTANCE ERROR BY ITERATIVE METHODS FOR STRENGTH DETERMINATION OF HDR 192IR BRACHYTHERAPY SOURCE
    Kumar, Sudhir
    Srinivasan, Panchapakesan
    Sharma, Sunil D.
    Subbaiah, Kamatam V.
    Mayya, Yelia S.
    MEDICAL DOSIMETRY, 2010, 35 (03) : 230 - 237