The Measurement of the Air-Kerma Rate in Air and a Solid Phantom with Ionization Chambers for a 192Ir HDR Brachytherapy Source

被引:0
作者
Zeng, Jing [1 ]
Qu, Pengpeng [1 ]
Pang, Qingsong [2 ]
Wang, Ping [2 ]
机构
[1] Nankai Univ, Tianjin Cent Hosp Gynecol & Obstet, Dept Gynecol Oncol, Affiliated Hosp, Tianjin 300100, Peoples R China
[2] Tianjin Med Univ Canc Inst & Hosp, Natl Clin Res Ctr Canc, Tianjins Clin Res Ctr Canc, Key Lab Canc Prevent & Therapy, Tianjin 300060, Peoples R China
关键词
brachytherapy; Ir-192; source; calibration; ionization chamber; TREATMENT PLANNING SYSTEMS; PROTOCOL;
D O I
10.2147/CMAR.S275378
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Introduction: This study aims to measure the air-kerma rate of 192-Ir-HDR-afterloading source with an ionization chamber in air and a solid cylindrical phantom separately and to compare the dose calibration by the American Association of Physicists in Medicine (AAPM) Task Group TG-43U1 formalism with the Abacus treatment planning system (TPS). Materials and Methods: The air-kerma rate of Ir-192 source was measured by an ionization chamber in air and a solid cylindrical phantom separately. For the interesting point position P (8cm, 90 degrees), the values of the dose were calculated with the TG-43U1 formula and compared with data from the Abacus TPS with single and multiple dwell positions, respectively. Results: The air-kerma rate percentage deviations between the detector measurements in air and the source certificate were -1.28%, -0.91%, -0.71%, and 0.33% at the distances of 25cm, 50cm, 75cm, and 100cm, respectively. For the measurement in solid cylindrical phantom, the percent deviation from the air-kerma rate certificate was 1.85%. The percentage deviations of the dose calibration between Abacus TPS and TG-43U1 formalism at P (8cm, 90 degrees) were -2.30%, 1.76%, and 2.10% with different distances (between the dwell positions) of 0cm, 0.5cm, and 1cm, respectively. Conclusion: The in-air technique was a new attempt for clinic routine measurement. Further studies are still necessary. As a treatment planning system, the Abacus TPS should apply the AAPM TG-43U1 formulism for the development required in the future.
引用
收藏
页码:10821 / 10828
页数:8
相关论文
共 50 条
[21]   On source models for 192Ir HDR brachytherapy dosimetry using model based algorithms [J].
Pantelis, Evaggelos ;
Zourari, Kyveli ;
Zoros, Emmanouil ;
Lahanas, Vasileios ;
Karaiskos, Pantelis ;
Papagiannis, Panagiotis .
PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (11) :4235-4246
[22]   Ir-192 Calibration in Air with Farmer Chamber for HDR Brachytherapy [J].
Chang, Liyun ;
Ho, Sheng-Yow ;
Lee, Tsair-Fwu ;
Ding, Hueisch-Jy ;
Chen, Pang-Yu .
JOURNAL OF MEDICAL AND BIOLOGICAL ENGINEERING, 2016, 36 (02) :145-152
[23]   A technique for calibrating a high dose rate 192Ir brachytherapy source [J].
Fourie, O. L. ;
Crabtree, T. G. .
AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2012, 35 (01) :85-92
[24]   Experimental derivation of wall correction factors for ionization chambers used in high dose rate 192Ir source calibration [J].
Maréchal, MH ;
de Almeida, CE ;
Ferreira, IH ;
Sibata, CH .
MEDICAL PHYSICS, 2002, 29 (01) :1-5
[25]   Simulation evaluation of NIST air-kerma rate calibration standard for electronic brachytherapy [J].
Hiatt, Jessica R. ;
Rivard, Mark J. ;
Hughes, H. Grady .
MEDICAL PHYSICS, 2016, 43 (03) :1119-1129
[26]   The non-uniformity correction factor for the cylindrical ionization chambers in dosimetry of an HDR Ir-192 brachytherapy source [J].
Majumdar, Bishnu ;
Patel, Narayan Prasad ;
Vijayan, V. .
JOURNAL OF MEDICAL PHYSICS, 2006, 31 (02) :83-88
[27]   Reference air kerma rate calibration system for high dose rate Ir-192 brachytherapy sources in Taiwan [J].
Chu, Wei-Han ;
Yuan, Ming-Chen ;
Lee, Jeng-Hung ;
Lin, Yi-Chun .
RADIATION PHYSICS AND CHEMISTRY, 2017, 140 :361-364
[28]   High-Dose-Rate 192Ir Brachytherapy Dose Verification: A Phantom Study [J].
Nikoofar, Alireza ;
Hoseinpour, Zohreh ;
Mahdavi, Seied Rabi ;
Hasanzadeh, Hadi ;
Tavirani, Mostafa Rezaei .
IRANIAN JOURNAL OF CANCER PREVENTION, 2015, 8 (03)
[29]   Dosimetry revisited for the HDR 192Ir brachytherapy source model mHDR-v2 [J].
Granero, Domingo ;
Vijande, Javier ;
Ballester, Facundo ;
Rivard, Mark J. .
MEDICAL PHYSICS, 2011, 38 (01) :487-494
[30]   Enhancement and validation of Geant4 Brachytherapy application on clinical HDR 192Ir source [J].
Ababneh, Eshraq ;
Dababneh, Saed ;
Qatarneh, Sharif ;
Wadi-Ramahi, Shada .
RADIATION PHYSICS AND CHEMISTRY, 2014, 103 :57-66