Prediction of the chromatographic signal in gradient elution ion chromatography

被引:8
|
作者
Bolanca, Tomislav [1 ]
Stefanovic, Stefica Cerjan [1 ]
Ukic, Sime [1 ]
Rogosic, Marko [1 ]
Lusa, Melita [1 ]
机构
[1] Univ Zagreb, Fac Chem Engn & Technol, Zagreb 10000, Croatia
关键词
Ion chromatography; Peak shape modeling; Retention modeling; ARTIFICIAL NEURAL NETWORKS; ANION CHROMATOGRAPHY; RETENTION MODELS; HYDROXIDE ELUENTS; MULTIPLE ELUENT; PEAK FUNCTIONS; OPTIMIZATION; SEPARATION; DECONVOLUTION; PHOSPHATE;
D O I
10.1002/jssc.200900231
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This study describes the development of a signal prediction model in gradient elution ion chromatography. The proposed model is based on a retention model and generalized logistic peak shape function which guarantees simplicity of the model and its easy implementation in method development process. Extensive analysis of the model predictive ability has been performed for ion chromatographic determination of bromate, nitrite, bromide, iodide, and perchlorate, using KOH solutions as eluent. The developed model shows good predictive ability (average relative error of gradient predictions 1.94%). The developed model offers short calculation times as well as low experimental effort (only nine isocratic runs are used for modeling).
引用
收藏
页码:2877 / 2884
页数:8
相关论文
共 50 条
  • [21] Effects of peak compression in gradient elution of liquid chromatography
    Hao Weiqiang
    Liu Lijuan
    Shen Qiaoyin
    CHINESE JOURNAL OF CHROMATOGRAPHY, 2021, 39 (01) : 10 - 14
  • [22] The ultimate band compression factor in gradient elution chromatography
    Gritti, Fabrice
    Guiochon, Georges
    JOURNAL OF CHROMATOGRAPHY A, 2008, 1178 (1-2) : 79 - 91
  • [23] Some insights on the description of gradient elution in reversed-phase liquid chromatography
    Jose Baeza-Baeza, Juan
    Celia Garcia-Alvarez-Coque, Maria
    JOURNAL OF SEPARATION SCIENCE, 2014, 37 (17) : 2269 - 2277
  • [24] PREDICTION OF CHROMATOGRAPHIC RETENTION OF METAL-COMPLEXES FROM STABILITY-CONSTANTS MEASURED BY ION CHROMATOGRAPHY
    ALUMAA, P
    PENTSUK, J
    CHROMATOGRAPHIA, 1994, 38 (9-10) : 566 - 570
  • [25] Probing the kinetic performance limits for ion chromatography. II. Gradient conditions for small ions
    Causon, Tim J.
    Hilder, Emily F.
    Shellie, Robert A.
    Haddad, Paul R.
    JOURNAL OF CHROMATOGRAPHY A, 2010, 1217 (31) : 5063 - 5068
  • [26] Measurement of optimal flow rate in gradient elution liquid chromatography
    Foster, Samuel W.
    Wright, Nicholas
    Grinias, James P.
    Blumberg, Leonid M.
    JOURNAL OF CHROMATOGRAPHY A, 2021, 1659
  • [27] Optimization of gradient elution conditions in multicomponent preparative liquid chromatography
    Shan, YC
    Seidel-Morgenstern, A
    JOURNAL OF CHROMATOGRAPHY A, 2005, 1093 (1-2) : 47 - 58
  • [28] Computer aided optimization of multilinear gradient elution in liquid chromatography
    Hao, Weiqiang
    Li, Bo
    Deng, Yuying
    Chen, Qiang
    Liu, Lijuan
    Shen, Qiaoyin
    JOURNAL OF CHROMATOGRAPHY A, 2021, 1635 (1635)
  • [29] Determination of Pyridinium Ionic Liquid Cations by Reversed Phase Ion-Pair Chromatography Using Gradient Elution
    Meng, Lingmin
    Yu, Hong
    Liu, Yuzhen
    CHROMATOGRAPHIA, 2011, 73 (3-4) : 367 - 371
  • [30] Theoretical background of short chromatographic layers Optimization of gradient elution in short columns
    Yamamoto, S
    Kita, A
    JOURNAL OF CHROMATOGRAPHY A, 2005, 1065 (01) : 45 - 50