Wall correction factors, Pwall, for parallel-plate ionization chambers

被引:42
作者
Buckley, Lesley A. [1 ]
Rogers, D. W. O.
机构
[1] Carleton Univ, Ottawa Carleton Inst Phys, Ottawa, ON K1S 5B6, Canada
[2] Natl Res Council Canada, Ottawa, ON K1A 0R6, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
Monte Carlo; dosimetry; correction factor;
D O I
10.1118/1.2199988
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The EGSnrc Monte Carlo user-code CSnrc is used to calculate wall correction factors, P-wall, for parallel-plate ionization chambers in photon and electron beams. A set Of P-wall values, computed at the reference depth in water, is presented for several commonly used parallel-plate chambers. These values differ from the standard assumption of unity used by dosimetry protocols by up to 1.7% for clinical electron beams. Calculations also show that P-wall is strongly dependent on the depth of measurement and can vary by as much as 6% for a 6 MeV beam in moving from a depth of d(ref) to a depth of R-50, In photon beams, where there is limited information available regarding P-wall for parallel-plate chambers, CSnrc calculations show P-wall values of up to 2.4% at the reference depth over a range of photon energies. The P-wall values for photon beams are in good agreement with previous estimates of the wall correction but have much lower statistical uncertainties and cover a wider range of photon beam energies. (C) 2006 American Association of Physicists in Medicine.
引用
收藏
页码:1788 / 1796
页数:9
相关论文
共 37 条
[1]   THE CALIBRATION AND USE OF PLANE-PARALLEL IONIZATION CHAMBERS FOR DOSIMETRY OF ELECTRON-BEAMS - AN EXTENSION OF THE 1983 AAPM PROTOCOL REPORT OF AAPM RADIATION-THERAPY COMMITTEE TASK GROUP NO-39 [J].
ALMOND, PR ;
ATTIX, FH ;
HUMPHRIES, LJ ;
KUBO, H ;
NATH, R ;
GOETSCH, S ;
ROGERS, DWO .
MEDICAL PHYSICS, 1994, 21 (08) :1251-1260
[2]   AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams [J].
Almond, PR ;
Biggs, PJ ;
Coursey, BM ;
Hanson, WF ;
Huq, MS ;
Nath, R ;
Rogers, DWO .
MEDICAL PHYSICS, 1999, 26 (09) :1847-1870
[3]   IONIZATION-CHAMBER DOSIMETRY FOR PHOTON AND ELECTRON-BEAMS - THEORETICAL CONSIDERATIONS [J].
ALMOND, PR ;
SVENSSON, H .
ACTA RADIOLOGICA-THERAPY PHYSICS BIOLOGY, 1977, 16 (02) :177-186
[4]  
[Anonymous], PIRS702 NAT RES COUN
[5]   PENELOPE - AN ALGORITHM FOR MONTE-CARLO SIMULATION OF THE PENETRATION AND ENERGY-LOSS OF ELECTRONS AND POSITRONS IN MATTER [J].
BARO, J ;
SEMPAU, J ;
FERNANDEZVAREA, JM ;
SALVAT, F .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 1995, 100 (01) :31-46
[6]   Wall correction factors, Pwall, for thimble ionization chambers [J].
Buckley, LA ;
Rogers, DWO .
MEDICAL PHYSICS, 2006, 33 (02) :455-464
[7]   CSnrc: Correlated sampling Monte Carlo calculations using EGSnrc [J].
Buckley, LA ;
Kawrakow, I ;
Rogers, DWO .
MEDICAL PHYSICS, 2004, 31 (12) :3425-3435
[8]  
Ding G, 1995, PIRS, V439
[9]   Measurement of Prepl Pwall factors in electron beams and in a 60Co beam for plane-parallel chambers [J].
Ding, GX ;
Cygler, JE .
MEDICAL PHYSICS, 1998, 25 (08) :1453-1457
[10]   THE EFFECT OF THICKNESS OF THE WATERPROOFING SHEATH ON THE CALIBRATION OF PHOTON AND ELECTRON-BEAMS [J].
GILLIN, MT ;
KLINE, RW ;
NIROOMANDRAD, A ;
GRIMM, DF .
MEDICAL PHYSICS, 1985, 12 (02) :234-236