Overcoming Electromagnetic Interference by LVADs on ICD Function by Shielding the ICD Programmer Wand and Extension Cable

被引:16
作者
Biviano, Angelo [1 ]
Mancini, Donna [1 ]
Naka, Yoshifumi [1 ]
Arellano, Jerry [1 ]
Garan, Hasan [1 ]
机构
[1] Columbia Univ, Med Ctr, NewYork Presbyterian Hosp, Div Cardiol,Dept Med, New York, NY 10027 USA
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2009年 / 32卷 / 07期
关键词
ICD; LVAD; electromagnetic interference; VENTRICULAR ASSIST DEVICE; IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR; CONGESTIVE-HEART-FAILURE; ARRHYTHMIAS; CARDIOMYOPATHY; OUTCOMES; SUPPORT; THERAPY;
D O I
10.1111/j.1540-8159.2009.02419.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Patients with end-stage cardiomyopathy and congestive heart failure are increasingly undergoing implantation with left ventricular assist devices (LVADs). In addition, implantable cardioverter-defibrillator (ICD)therapy has been proven to be an important part of the treatment for cardiomyopathy/ congestive heart failure. Previous reports have noted a potential and dramatic electromagnetic interference from LVADs on ICDs that cause impaired telemetry communication between the ICD and ICD programmer. Such interference has necessitated explantation and generator replacement in order to resume communication between the ICD and programmer. We report two patients with advanced congestive heart failure and ICD programming impairment caused by a HeartMate II LVAD (Thoratec Corporation, Pleasanton, CA, USA) that was overcome by placing aluminum shielding around the ICD programmer wand and steel shielding around the extension cable during ICD interrogation. ( PACE 2009; 32: 945-948)
引用
收藏
页码:945 / 948
页数:4
相关论文
共 17 条
[1]   Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure [J].
Bardy, GH ;
Lee, KL ;
Mark, DB ;
Poole, JE ;
Packer, DL ;
Boineau, R ;
Domanski, M ;
Troutman, C ;
Anderson, J ;
Johnson, G ;
McNulty, SE ;
Clapp-Channing, N ;
Davidson-Ray, LD ;
Fraulo, ES ;
Fishbein, DP ;
Luceri, RM ;
Ip, JH .
NEW ENGLAND JOURNAL OF MEDICINE, 2005, 352 (03) :225-237
[2]   Ventricular arrhythmias during left ventricular assist device support [J].
Bedi, Maninder ;
Kormos, Robert ;
Winowich, Steve ;
McNamara, Dennis M. ;
Mathier, Michael A. ;
Murali, Srinivas .
AMERICAN JOURNAL OF CARDIOLOGY, 2007, 99 (08) :1151-1153
[3]  
Draft guidance for industry and FDA staff, 2007, DRAFT GUID IND FDA S
[4]   Prolonged repolarization after ventricular assist device support is associated with arrhythmias in humans with congestive heart failure [J].
Harding, JD ;
Piacentino, V ;
Rothman, S ;
Chambers, S ;
Jessup, M ;
Margulies, KB .
JOURNAL OF CARDIAC FAILURE, 2005, 11 (03) :227-232
[5]   Outcomes of left ventricular assist device implantation as destination therapy in the post-REMATCH era - Implications for patient selection [J].
Lietz, Katherine ;
Long, James W. ;
Kfoury, Abdallah G. ;
Slaughter, Mark S. ;
Silver, Marc A. ;
Milano, Carmelo A. ;
Rogers, Joseph G. ;
Naka, Yoshifumi ;
Mancini, Donna ;
Miller, Leslie W. .
CIRCULATION, 2007, 116 (05) :497-505
[6]  
Long James W, 2005, Congest Heart Fail, V11, P133, DOI 10.1111/j.1527-5299.2005.04540.x
[7]  
Matthews JC, 2007, J CARDIOVASC ELECTR, V18, P1107, DOI 10.1111/j.1540-8167.2007.00805.x
[8]  
MEHTA R, 2007, HEART RHYTHM, V5, pS284
[9]   Device interactions in patients with advanced cardiomyopathy [J].
Mehta, Rohit ;
Doshi, Amit A. ;
Hasan, Ayesha K. ;
Love, Charles J. ;
Pizzuto, Marg ;
Sai-Sudhakar, Chittoor ;
Chan, David P. .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2008, 51 (16) :1613-1614
[10]  
Mehta R, 2007, J CARDIOVASC ELECTR, V18, pE27, DOI 10.1111/j.1540-8167.2007.00805.x