Performance of clinically available deep learning image reconstruction in computed tomography: a phantom study

被引:21
作者
Kawashima, Hiroki [1 ]
Ichikawa, Katsuhiro [1 ]
Takata, Tadanori [2 ]
Mitsui, Wataru [2 ]
Ueta, Hiroshi [2 ]
Yoneda, Norihide [3 ]
Kobayashi, Satoshi [1 ]
机构
[1] Kanazawa Univ, Fac Hlth Sc, Inst Med Pharmaceut & Hlth Sci, Kanazawa, Ishikawa, Japan
[2] Kanazawa Univ Hosp, Radiol Div, Kanazawa, Ishikawa, Japan
[3] Kanazawa Univ, Dept Radiol, Grad Sch Med Sci, Kanazawa, Ishikawa, Japan
关键词
computed tomography; image reconstruction; deep learning; dose reduction; TASK-BASED PERFORMANCE; ITERATIVE RECONSTRUCTION; ABDOMINAL CT; REDUCTION TECHNIQUES; QUALITY; NOISE; RESOLUTION; MDCT; DETECTABILITY; INDEX;
D O I
10.1117/1.JMI.7.6.063503
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To assess the physical performance of deep learning image reconstruction (DLIR) compared with those of filtered back projection (FBP) and iterative reconstruction (IR) and to estimate the dose reduction potential of the technique. Approach: A cylindrical water bath phantom with a diameter of 300 mm including two rods composed of acrylic and soft tissue-equivalent material was scanned using a clinical computed tomography (CT) scanner at four dose levels (CT dose index of 20, 15, 10, and 5 mGy). Phantom images were reconstructed using FBP, DLIR, and IR. The in-plane and z axis task transfer functions (TTFs) and in-plane noise power spectrum (NPS) were measured. The dose reduction potential was estimated by evaluating the system performance function calculated from TTF and NPS. The visibilities of a bar pattern phantom placed in the same water bath phantom were compared. Results: The use of DLIR resulted in a notable decrease in noise magnitude. The shift in peak NPS frequency was reduced compared with IR. Preservation of in-plane TTF was superior using DLIR than using IR. The estimated dose reduction potentials of DLIR and IR were 39% to 54% and 19% to 29%, respectively. However, the z axis resolution was decreased with DLIR by 6% to 21% compared with FBP. The bar pattern visibilities were approximately consistent with the TTF results in both planes. Conclusions: The in-plane edge-preserving noise reduction performance of DLIR is superior to that of IR. Moreover, DLIR enables approximately half-dose acquisitions with no deterioration in noise texture in cases that permit some z axis resolution reduction. (C) 2020 Society of Photo-Optical Instrumentation Engineers (SPIE).
引用
收藏
页数:12
相关论文
共 35 条
[1]   Deep learning reconstruction improves image quality of abdominal ultra-high-resolution CT [J].
Akagi, Motonori ;
Nakamura, Yuko ;
Higaki, Toru ;
Narita, Keigo ;
Honda, Yukiko ;
Zhou, Jian ;
Yu, Zhou ;
Akino, Naruomi ;
Awai, Kazuo .
EUROPEAN RADIOLOGY, 2019, 29 (11) :6163-6171
[2]  
[Anonymous], 2019, PERF EV COMP TOM SYS
[3]  
[Anonymous], 2006, J ICRU
[4]   Validation of deep-learning image reconstruction for coronary computed tomography angiography: Impact on noise, image quality and diagnostic accuracy [J].
Benz, Dominik C. ;
Benetos, Georgios ;
Rampidis, Georgios ;
von Felten, Elia ;
Bakula, Adam ;
Sustar, Aleksandra ;
Kudura, Ken ;
Messerli, Michael ;
Fuchs, Tobias A. ;
Gebhard, Catherine ;
Pazhenkottil, Aju P. ;
Kaufmann, Philipp A. ;
Buechel, Ronny R. .
JOURNAL OF CARDIOVASCULAR COMPUTED TOMOGRAPHY, 2020, 14 (05) :444-451
[5]   Application of the noise power spectrum in modern diagnostic MDCT: Part I. Measurement of noise power spectra and noise equivalent quanta [J].
Boedeker, K. L. ;
Cooper, V. N. ;
McNitt-Gray, M. F. .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (14) :4027-4046
[6]   Assessment of volumetric noise and resolution performance for linear and nonlinear CT reconstruction methods [J].
Chen, Baiyu ;
Christianson, Olav ;
Wilson, Joshua M. ;
Samei, Ehsan .
MEDICAL PHYSICS, 2014, 41 (07)
[7]   Image quality comparison of two adaptive statistical iterative reconstruction (ASiR, ASiR-V) algorithms and filtered back projection in routine liver CT [J].
Chen, Li-Hong ;
Jin, Chao ;
Li, Jian-Ying ;
Wang, Ge-Liang ;
Jia, Yong-Jun ;
Duan, Hai-Feng ;
Pan, Ning ;
Guo, Jianxin .
BRITISH JOURNAL OF RADIOLOGY, 2018, 91 (1088)
[8]   An Improved Index of Image Quality for Task-based Performance of CT Iterative Reconstruction across Three Commercial Implementations [J].
Christianson, Olav ;
Chen, Joseph J. S. ;
Yang, Zhitong ;
Saiprasad, Ganesh ;
Dima, Alden ;
Filliben, James J. ;
Peskin, Adele ;
Trimble, Christopher ;
Siegel, Eliot L. ;
Samei, Ehsan .
RADIOLOGY, 2015, 275 (03) :725-734
[9]   Methods for Clinical Evaluation of Noise Reduction Techniques in Abdominopelvic CT [J].
Ehman, Eric C. ;
Yu, Lifeng ;
Manduca, Armando ;
Hara, Amy K. ;
Shiung, Maria M. ;
Jondal, Dayna ;
Lake, David S. ;
Paden, Robert G. ;
Blezek, Daniel J. ;
Bruesewitz, Michael R. ;
McCollough, Cynthia H. ;
Hough, David M. ;
Fletcher, Joel G. .
RADIOGRAPHICS, 2014, 34 (04) :849-862
[10]   Technical advances in multi-slice spiral CT [J].
Fuchs, T ;
Kachelriess, M ;
Kalender, WA .
EUROPEAN JOURNAL OF RADIOLOGY, 2000, 36 (02) :69-73