The Relationship Between Increased Hip Range of Motion, Wear, and Locking Mechanism Failure in the Harris-Galante Acetabular Component

被引:7
作者
Imai, Hiroshi [1 ]
Mashima, Naohiko [1 ]
Takahashi, Toshiaki [1 ]
Yamamoto, Haruyasu [1 ]
机构
[1] Ehime Univ, Grad Sch Med, Dept Bone & Joint Surg, Toon, Ehime 7910295, Japan
关键词
total hip arthroplasty; acetabular component; polyethylene wear; range of motion; POLYETHYLENE LINER DISSOCIATION; FOLLOW-UP; RADIOGRAPHIC ANALYSIS; SCREW FIXATION; ARTHROPLASTY; REPLACEMENT; CEMENT; CUP;
D O I
10.1016/j.arth.2008.07.003
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We performed both clinical and radiographic evaluations of 178 patients (190 hips) who had undergone cementless total hip arthroplasties using Harris-Galante I/II porous Cups after an average 12-year follow-up period (range, 818 years). We revised 15 Harris-Galante I/II porous cups (7.8%), and the locking mechanism was broken in 10 revised cups (67%). There was a significant association between locking mechanism failure and linear polyethylene Wear. We observed a significant positive correlation between linear polyethylene wear and increased ranges of motion such as flexion, adduction, and external rotation at the last follow-up visit after the primary operation. Increased ranges of motion seen in Asians induced higher linear polyethylene wear and locking mechanism failure due to impingement of the neck and cup.
引用
收藏
页码:892 / 897
页数:6
相关论文
共 22 条
[1]   The Harris-Galante porous-coated acetabular component with screw fixation - An average ten-year follow-up study [J].
Clohisy, JC ;
Harris, WH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1999, 81A (01) :66-73
[2]  
DELEE JG, 1976, CLIN ORTHOP RELAT R, P20
[3]   Primary total hip arthroplasty with a porous-coated acetabular component - A concise follow-up of a previous report [J].
Della Valle, CJ ;
Berger, RA ;
Shott, S ;
Rosenberg, AG ;
Jacobs, JJ ;
Quigley, L ;
Galante, JO .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2004, 86A (06) :1217-1222
[4]  
Dunkley AB, 2000, CLIN ORTHOP RELAT R, P149
[5]  
GONZALEZ MH, 1988, CLIN ORTHOP RELAT R, P156
[6]   Hip, knee, and ankle kinematics of high range of motion activities of daily living [J].
Hemmerich, A ;
Brown, H ;
Smith, S ;
Marthandam, SSK ;
Wyss, UP .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2006, 24 (04) :770-781
[7]  
HODGKINSON JP, 1988, CLIN ORTHOP RELAT R, P105
[8]   NONPARAMETRIC-ESTIMATION FROM INCOMPLETE OBSERVATIONS [J].
KAPLAN, EL ;
MEIER, P .
JOURNAL OF THE AMERICAN STATISTICAL ASSOCIATION, 1958, 53 (282) :457-481
[9]   DISLOCATIONS AFTER TOTAL HIP-REPLACEMENT ARTHROPLASTIES [J].
LEWINNEK, GE ;
LEWIS, JL ;
TARR, R ;
COMPERE, CL ;
ZIMMERMAN, JR .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1978, 60 (02) :217-220
[10]   EFFECT OF FEMORAL-HEAD SIZE ON WEAR OF THE POLYETHYLENE ACETABULAR COMPONENT [J].
LIVERMORE, J ;
ILSTRUP, D ;
MORREY, B .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1990, 72A (04) :518-528