Safety of MRI in patients with implanted deep brain stimulation devices

被引:90
作者
Tagliati, Michele [1 ]
Jankovic, Joseph [2 ]
Pagan, Fernando [3 ]
Susatia, Frandy [4 ]
Isaias, Ioannis U. [1 ]
Okun, Michael S. [4 ]
机构
[1] Mt Sinai Sch Med, Dept Neurol, New York, NY 10029 USA
[2] Baylor Coll Med, Dept Neurol, Houston, TX 77030 USA
[3] Georgetown Univ Hosp, Dept Neurol, Washington, DC 20007 USA
[4] Univ Florida, Dept Neurol, Gainesville, FL USA
关键词
MAGNETIC-RESONANCE; IN-VITRO; DISEASE;
D O I
10.1016/j.neuroimage.2009.04.044
中图分类号
Q189 [神经科学];
学科分类号
071006 ;
摘要
Objective: To survey the safety of MRI in PD patients implanted with DBS devices. Background: MRI in patients with DBS implants is useful to Confirm electrode placement, optimize programming and investigating complications. However, several medical centers do not perform MRI studies in DBS patients because of safety concerns. The safety profile of MRI in DBS patients has not been documented in large clinical series. Methods: 42 NPF Centers of Excellence (COEs) were asked to complete a questionnaire on MRI use and DBS. Results: Investigators from 40 of 42 (95%) NPF COEs completed the survey and 23 (58%) reported that they were Currently performing brain MRI in DBS patients, while 3 (7.5%) had done it in the past. The 17 COEs currently not performing post-operative MRI for DBS listed the following reasons: 1) industry guidelines and/or warnings (53%); 2) decision deferred to Outside department (29%); 3) liability/risk/safety (18%); 4) no active DBS program (18%): 5) no available MRI (12%); and 6) insurance and reimbursement concerns (6%). A total of 3304 PD patients with one or more DBS leads had a brain MRI scan, and 177 DBS patients had MRI of other body regions. In one case MRI was associated with an IPG failure without neurological sequelae after IPG replacement. No other complications were reported. Conclusions: these data provide evidence for a favorable risk/benefit ratio for brain MRI in patients with DBS implants. Further studies will need to address whether a re-assessment of more restrictive recommendations (i.e. very low SAR values) may be warranted. (C) 2009 Elsevier Inc. All rights reserved.
引用
收藏
页码:T53 / T57
页数:5
相关论文
共 16 条
[1]   Performing functional magnetic resonance imaging in patients with Parkinson's disease treated with deep brain stimulation [J].
Arantes, Paula R. ;
Cardoso, Ellison F. ;
Barreiros, Maria A. ;
Teixeira, Manoel J. ;
Goncalves, Marcia R. ;
Barbosa, Egberto R. ;
Sukwinder, Sukhi Shergill ;
Leite, Claudia C. ;
Amaro, Edson, Jr. .
MOVEMENT DISORDERS, 2006, 21 (08) :1154-1162
[2]   REDUCTION OF MAGNETIC RESONANCE IMAGING-RELATED HEATING IN DEEP BRAIN STIMULATION LEADS USING A LEAD MANAGEMENT DEVICE [J].
Baker, Kenneth B. ;
Tkach, Jean ;
Hall, John D. ;
Nyenhuis, John A. ;
Shellock, Frank G. ;
Rezai, Ali R. .
NEUROSURGERY, 2005, 57 (04) :392-396
[3]   Bilateral neuro stimulation systems used for deep brain stimulation: in vitro study of MRI-related heating at 1.5 T and implications for clinical imaging of the brain [J].
Bhidayasiri, R ;
Bronstein, JM ;
Sinha, S ;
Krahl, SE ;
Ahn, S ;
Behnke, EJ ;
Cohen, MS ;
Frysinger, R ;
Shellock, FG .
MAGNETIC RESONANCE IMAGING, 2005, 23 (04) :549-555
[4]   Deep brain stimulation in neurologic disorders [J].
Halpern, Casey ;
Hurtig, Howard ;
Jaggi, Jurg ;
Grossman, Murray ;
Won, Michelle ;
Baltuch, Gordon .
PARKINSONISM & RELATED DISORDERS, 2007, 13 (01) :1-16
[5]   Prevention of mechanical failures in implanted spinal cord stimulation systems [J].
Henderson, JM ;
Oakley, JC ;
Schade, CM ;
Sasaki, J ;
Caraway, DL .
NEUROSURGERY, 2005, 57 (02) :414-414
[6]   Functional magnetic resonance imaging during deep brain stimulation:: A pilot study in four patients with Parkinson's disease [J].
Jech, R ;
Urgosík, D ;
Tintera, J ;
Nebuzelsky, A ;
Krásensky, J ;
Liscák, R ;
Roth, J ;
Ruzicka, E .
MOVEMENT DISORDERS, 2001, 16 (06) :1126-1132
[7]   Magnetic resonance imaging of implanted deep brain stimulators: Experience in a large series [J].
Larson, Paul S. ;
Richardson, R. Mark ;
Starr, Philip A. ;
Martin, Alastair J. .
STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2008, 86 (02) :92-100
[8]   Deep brain stimulation for Parkinson's disease [J].
Limousin, Patricia ;
Martinez-Torres, Irene .
NEUROTHERAPEUTICS, 2008, 5 (02) :309-319
[9]   DBS and diathermy interaction induces severe CNS damage [J].
Nutt, JG ;
Anderson, VC ;
Peacock, JH ;
Hammerstad, JP ;
Burchiel, KJ .
NEUROLOGY, 2001, 56 (10) :1384-1386
[10]  
Patterson Terry, 2007, Am J Electroneurodiagnostic Technol, V47, P93