Purpose: To assess the value of percutaneous vacuum-assisted core biopsy to improve the diagnosis of non palpable mammographic abnormalities. Materials and methods: A total of 252 core biopsies using an 11G Mammotome(R) were performed in 249 patients. Stereotactic localization was performed in the prone position on a dedicated digital Fischer table. Results: Fifty-one, or 25%, of 200 clusters of microcalcifications corresponded to carcinomas: 126 benign lesions, 23 atypical hyperplasia and LCIS, 31 DCIS, 15 invasive ductal carcinomas, and 4 false negative biopsies. In these 4 last cases, surgery was performed because radiographs of the core biopsy showed no microcalcifications; carcinoma was confirmed at histology of the surgical specimen. Using the BI-RADS system, 7 lesions were category 3, 175 lesions were category 4, and 18 lesions were category 5. From a total of 52 masses, 31 were benign lesions, 2 were borderline lesions, and 19 were invasive carcinomas. From these, 5 lesions were category 3, 31 were category 4, and 16 were category 5. Diagnostic surgical biopsy was avoided in 161 cases (63%), in 152 cases for benign lesions including 151 lesions classified as category 4 lesions and in 9 cases for multifocal or recurrent malignant lesions. Conclusion: When technical pitfalls are avoided and when presence of microcalcifications in the core biopsy sample is verified, vacuum assisted core biopsy with Mammotome(R) 11G provides accurate diagnosis of non-palpable mammographic abnormalities.
引用
收藏
页码:1063 / 1071
页数:9
相关论文
共 50 条
  • [41] Impact of stereotactic 11-g vacuum-assisted breast biopsy on cost of diagnosis in Austria
    Gruber, R.
    Walter, E.
    Helbich, T. H.
    EUROPEAN JOURNAL OF RADIOLOGY, 2011, 77 (01) : 131 - 136
  • [42] Borderline breast lesions: comparison of malignancy underestimation rates with 14-gauge core needle biopsy versus 11-gauge vacuum-assisted device
    Londero, Viviana
    Zuiani, Chiara
    Linda, Anna
    Battigelli, Luisa
    Brondani, Giovanni
    Bazzocchi, Massimo
    EUROPEAN RADIOLOGY, 2011, 21 (06) : 1200 - 1206
  • [43] Benign intraductal papilloma: Diagnosis and removal at stereotactic vacuum-assisted directional biopsy guided by galactography
    Guenin, MA
    RADIOLOGY, 2001, 218 (02) : 576 - 579
  • [44] Benign breast lesions: Minimally invasive vacuum-assisted biopsy with 11-gauge needles - Patients acceptance and effect on follow-up imaging findings
    Huber, S
    Wagner, M
    Medl, M
    Czembirek, H
    RADIOLOGY, 2003, 226 (03) : 783 - 790
  • [45] Vacuum-assisted breast biopsy with 7-gauge, 8-gauge, 9-gauge, 10-gauge, and 11-gauge needles: how many specimens are necessary?
    Preibsch, Heike
    Baur, Astrid
    Wietek, Beate M.
    Kraemer, Bernhard
    Staebler, Annette
    Claussen, Claus D.
    Siegmann-Luz, Katja C.
    ACTA RADIOLOGICA, 2015, 56 (09) : 1078 - 1084
  • [46] Atypical ductal hyperplasia diagnosis by directional vacuum-assisted stereotactic biopsy of breast microcalcifications - Considerations for surgical excision
    Sneige, N
    Lim, SC
    Whitman, GJ
    Krishnamurthy, S
    Sahin, AA
    Smith, TL
    Stelling, CB
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2003, 119 (02) : 248 - 253
  • [47] Movement of a biopsy-site marker clip after completion of stereotactic directional vacuum-assisted breast biopsy: Case report
    Burnside, ES
    Sohlich, RE
    Sickles, EA
    RADIOLOGY, 2001, 221 (02) : 504 - 507
  • [48] Morphological parameters of lobular in situ neoplasia in stereotactic 11-gauge vacuum-assisted needle core biopsy do not predict the presence of malignancy on subsequent surgical excision
    Bianchi, Simonetta
    Bendinelli, Benedetta
    Castellano, Isabella
    Piubello, Quirino
    Renne, Giuseppe
    Cattani, Maria Grazia
    Di Stefano, Domenica
    Carrillo, Giovanna
    Laurino, Licia
    Bersiga, Alessandra
    Giardina, Carmela
    Dante, Stefania
    Di Loreto, Carla
    Quero, Carmela
    Antonacci, Concetta Maria
    Palli, Domenico
    HISTOPATHOLOGY, 2013, 63 (01) : 83 - 95
  • [49] Use of Bayesian Modeling to Estimate the Sensitivity of Stereotactic Directional Vacuum-Assisted Breast Biopsy When the Gold Standard is Incomplete
    Wang, Jane
    Kuo, Wen-Hung
    Pan, Shin-Liang
    Chang, King-Jen
    Teng, Yu-Chuan
    Ting-Fang, Tiffany
    ACADEMIC RADIOLOGY, 2009, 16 (11) : 1316 - 1322
  • [50] Accuracy of stereotactic vacuum-assisted breast biopsy with a 10-gauge hand-held system
    Salem, C.
    Sakr, R.
    Chopier, J.
    Marsault, C.
    Uzan, S.
    Darai, E.
    BREAST, 2009, 18 (03) : 178 - 182