High-energy neutron dosimetry with superheated drop detectors

被引:15
作者
d'Errico, F
Agosteo, S
Sannikov, AV
Silari, M
机构
[1] Yale Univ, Sch Med, Dept Therapeut Radiol, New Haven, CT 06520 USA
[2] Politecn Milan, Dipartimento Ingn Nucl, Milan, Italy
[3] Inst High Energy Phys, Protvino 142284, Russia
[4] CERN, CH-1211 Geneva, Switzerland
关键词
D O I
10.1093/oxfordjournals.rpd.a005929
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
A systematic analysis of the response of dichlorodifluoromethane superheated drop detectors was performed in the 46-133 MeV energy range. Experiments with quasi-monoenergetic neutron beams were performed at the Universite Catholique de Leuvain-la-Neuve. Belgium and the Svedberg Laboratory, Sweden, while tests in a broad field were performed at CERN. To determine the response of the detectors to the high-energy beams, the spectra of incident neutrons were folded over functions modelled after the cross sections for the production of heavy ions from the detector elements. The cross sections for fluorine and chlorine were produced in this work by means of the Monte Carlo high-energy transport code HADRON based on the cascade exciton model of nuclear interactions. The new response data permit the interpretation of measurements at high-energy accelerators and on high-altitude commercial flights, where a 30-50% under-response had been consistently recorded with respect to neutron dose equivalent. The introduction of a I em lead shell around the detectors effectively compensates most of the response defect.
引用
收藏
页码:529 / 532
页数:4
相关论文
共 21 条
[1]   Improved response of bubble detectors to high energy neutrons [J].
Agosteo, S ;
Silari, M ;
Ulrici, L .
RADIATION PROTECTION DOSIMETRY, 2000, 88 (02) :149-155
[2]   SUPERHEATED DROP DETECTOR [J].
APFEL, RE .
NUCLEAR INSTRUMENTS & METHODS, 1979, 162 (1-3) :603-608
[3]  
APFEL RE, 1993, RADIAT PROT DOSIM, V47, P551
[4]   In-phantom spectra and dose distributions from a high-energy neutron therapy beam [J].
Benck, S ;
d'Errico, F ;
Denis, JM ;
Meulders, JP ;
Nath, R ;
Pitcher, EJ .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2002, 476 (1-2) :127-131
[5]  
BIRATTARI C, 1992, RADIAT PROT DOSIM, V44, P193
[6]   Nuclear data for accelerator-driven systems [J].
Chadwick, MB ;
Hughes, HG ;
Little, RC ;
Pitcher, EJ ;
Young, PG .
PROGRESS IN NUCLEAR ENERGY, 2001, 38 (1-2) :179-219
[7]   A FACILITY FOR STUDIES OF NEUTRON-INDUCED REACTIONS IN THE 50-200 MEV RANGE [J].
CONDE, H ;
HULTQVIST, S ;
OLSSON, N ;
RONNQVIST, T ;
ZORRO, R ;
BLOMGREN, J ;
TIBELL, G ;
HAKANSSON, A ;
JONSSON, O ;
LINDHOLM, A ;
NILSSON, L ;
RENBERG, PU ;
BROCKSTEDT, A ;
EKSTROM, P ;
OSTERLUND, M ;
BRADY, FP ;
SZEFLINSKI, Z .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1990, 292 (01) :121-128
[9]   SUPERHEATED-DROP (BUBBLE) NEUTRON DETECTORS AND THEIR COMPLIANCE WITH ICRP-60 [J].
DERRICO, F ;
ALBERTS, WG .
RADIATION PROTECTION DOSIMETRY, 1994, 54 (3-4) :357-360
[10]   Neutron ambient dosimetry with superheated drop (bubble) detectors [J].
dErrico, F ;
Alberts, WG ;
Dietz, E ;
Gualdrini, G ;
Kurkdjian, J ;
Noccioni, P ;
Siebert, BRL .
RADIATION PROTECTION DOSIMETRY, 1996, 65 (1-4) :397-400