Two-point water-fat imaging with partially-opposed phase (POP) acquisition: An asymmetric Dixon method

被引:95
作者
Xiang, Qing-San
机构
[1] British Columbia Childrens Hosp, Dept Radiol, Vancouver, BC V6H 3V4, Canada
[2] Univ British Columbia, Dept Phys & Astron, Vancouver, BC V6H 3V4, Canada
关键词
water-fat imaging; Dixon method; asymmetric sampling; phase correction; regional iterative phasor extraction (RIPE);
D O I
10.1002/mrm.20984
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A novel two-point water-fat imaging method is introduced. In addition to the in-phase acquisition, water and fat magnetization vectors are sampled at partially-opposed-phase (POP) rather than exactly antiparallel as in the original Dixon method. This asymmetric sampling encodes more valuable phase information for identifying water and fat. From the magnitudes of the two complex images, a big and a small chemical component are first robustly obtained pixel by pixel and then used to form two possible error phasor candidates. The true error phasor is extracted from the two error phasor candidates through a simple procedure of regional iterative phasor extraction (RIPE). Finally, least-squares solutions of water and fat are obtained after the extracted error phasor is smoothed and removed from the complex images. For noise behavior, the effective number of signal averages NSA* is typically in the range of 1.87-1.96, very close to the maximum possible value of 2. Compared to earlier approaches, the proposed method is more efficient in data acquisition and straightforward in processing, and the final results are more robust. At both 1.5T and 0.3T, well separated and identified in vivo water and fat images covering a broad range of anatomical regions have been obtained, supporting the clinical utility of the method.
引用
收藏
页码:572 / 584
页数:13
相关论文
共 40 条
[1]  
AHN C B, 1986, Magnetic Resonance Imaging, V4, P110, DOI 10.1016/0730-725X(86)90921-5
[2]  
Akkerman E., 1995, P SOC MAGN RES EUR S, P649
[3]   Chemical shift imaging with spectrum modeling [J].
An, L ;
Xiang, QS .
MAGNETIC RESONANCE IN MEDICINE, 2001, 46 (01) :126-130
[4]  
AN L, 1998, THESIS U BRIT COLUMB
[5]  
AN L, 1998, P 6 ANN M ISMRM SYDN, P1866
[6]  
AN L, 1996, P ISMRM 4 ANN M SAN, P1541
[7]  
Bernstein MA, 2004, HDB MRI PULSE SEQUEN, P616
[8]   Fat suppression using direct phase encoding: Musculoskeletal applications using MR imaging [J].
Carson, BW ;
Xiang, QS .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1999, 173 (01) :230-233
[9]   Understanding phase maps in MRI: A new cutline phase unwrapping method [J].
Chavez, S ;
Xiang, QS ;
An, L .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2002, 21 (08) :966-977
[10]   AN INTRODUCTION TO THE ISING-MODEL [J].
CIPRA, BA .
AMERICAN MATHEMATICAL MONTHLY, 1987, 94 (10) :937-959