A compact high resolution electrospray ionization ion mobility spectrometer

被引:31
|
作者
Reinecke, T. [1 ]
Kirk, A. T. [1 ]
Ahrens, A. [1 ]
Raddatz, C. -R. [1 ]
Thoben, C. [1 ]
Zimmermann, S. [1 ]
机构
[1] Leibniz Univ Hannover, Inst Elect Engn & Measurement Technol, Appelstr 9A, D-30167 Hannover, Germany
关键词
Ion mobility spectrometry; Electrospray; Water analysis; Three grid ion shutter; GAS;
D O I
10.1016/j.talanta.2015.12.006
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Electrospray is a commonly used ionization method for the analysis of liquids. An electrospray is a dispersed nebular of charged droplets produced under the influence of a strong electrical field. Subsequently, ions are produced in a complex process initiated by evaporation of neutral solvent molecules from these droplets. We coupled an electrospray ionization source to our previously described high resolution ion mobility spectrometer with 75 mm drift tube length and a drift voltage of 5 kV. When using a tritium source for chemical gas phase ionization, a resolving power of R=100 was reported for this setup. We replaced the tritium source and the field switching shutter by an electrospray needle, a desolvation region with variable length and a three-grid shutter for injecting ions into the drift region. Preliminary measurements with tetraalkylammonium halides show that the current configuration with the electrospray ionization source maintains the resolving power of R=100. In this work, we present the characterization of our setup. One major advantage of our setup is that the desolvation region can be heated separately from the drift region so that the temperature in the drift region stays at room temperature even up to desolvation region temperatures of 100 degrees C. We perform parametric studies for the investigation of the influence of temperature on solvent evaporation with different ratios of water and methanol in the solvent for different analyte substances. Furthermore, the setup is operated in negative mode and spectra of bentazon with different solvents are presented. (C) 2015 Elsevier B.V. All rights reserved.
引用
收藏
页码:1 / 6
页数:6
相关论文
共 50 条
  • [31] A pulsed corona discharge switchable high resolution ion mobility spectrometer-mass spectrometer
    Hill, CA
    Thomas, CLP
    ANALYST, 2003, 128 (01) : 55 - 60
  • [32] Cluster Measurements at CLOUD using a High Resolution Ion Mobility Spectrometer - Mass Spectrometer Combination
    Adamov, Alexey
    Junninen, Heikki
    Duplissy, Jonathan
    Sipila, Mikko
    Kulmala, Markku
    NUCLEATION AND ATMOSPHERIC AEROSOLS, 2013, 1527 : 350 - 353
  • [33] A novel surface ionization source for ion mobility spectrometer
    Guo Hui-yong
    He Xiu-li
    Gao Xiao-guang
    Jia Jian
    Li Jian-ping
    ANALYTICA CHIMICA ACTA, 2007, 587 (01) : 137 - 141
  • [34] Pulsed-ionization miniature ion mobility spectrometer
    Xu, J
    Whitten, WB
    Ramsey, JM
    ANALYTICAL CHEMISTRY, 2003, 75 (16) : 4206 - 4210
  • [35] A PHOTO-IONIZATION SOURCE FOR AN ION MOBILITY SPECTROMETER
    EATHERTON, RL
    BAIM, MA
    HILL, HH
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 1983, 185 (MAR): : 22 - ANYL
  • [36] High resolving power electrospray drift tube ion mobility spectrometer with heated desolvation tube
    Boillat, Marc-Aurele
    Hauser, Peter C.
    ANALYTICA CHIMICA ACTA, 2025, 1338
  • [37] Development of an ion mobility spectrometer for use in an atmospheric pressure ionization ion mobility spectrometer/mass spectrometer instrument for fast screening analysis
    Sysoev, A
    Adamov, A
    Vildanoja, J
    Ketoja, RA
    Kostiainen, R
    Kotiaho, T
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2004, 18 (24) : 3131 - 3139
  • [38] Ion fragmentation in an electrospray ionization mass spectrometer interface with different gases
    Schneider, BS
    Douglas, DJ
    Chan, DDY
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2001, 15 (04) : 249 - 257
  • [39] High-resolution electrospray ionization orthogonal-injection time-of-flight mass spectrometer
    Dodonov, AF
    Kozlovski, VI
    Soulimenkov, IV
    Raznikov, VV
    Loboda, AV
    Zhen, Z
    Horwath, T
    Wollnik, H
    EUROPEAN JOURNAL OF MASS SPECTROMETRY, 2000, 6 (06) : 481 - 490
  • [40] High-resolution tissue imaging on an orbitrap mass spectrometer by desorption electrospray ionization mass spectrometry
    Manicke, N. E.
    Dill, A. L.
    Ifa, D. R.
    Cooks, R. G.
    JOURNAL OF MASS SPECTROMETRY, 2010, 45 (02): : 223 - 226