Evaluation of the automated coagulation analyzer CS-5100 and its utility in high throughput laboratories

被引:12
作者
Ratzinger, Franz [1 ]
Schmetterer, Klaus G. [1 ]
Haslacher, Helmuth [1 ]
Perkmann, Thomas [1 ]
Belik, Sabine [1 ]
Quehenberger, Peter [1 ]
机构
[1] Med Univ Vienna, Dept Lab Med, A-1090 Vienna, Austria
关键词
analyzer comparison; Bland-Altman plot; coefficient of variation; CS-5100; partial correlation coefficient; Passing-Bablok regression; PERFORMANCE-CHARACTERISTICS; ACL TOP; AGREEMENT; REGRESSION; HEMOLYSIS; QUALITY; CARE;
D O I
10.1515/cclm-2013-1094
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Automated analyzers are an important component of modern laboratories. As a representative of the newest generation of coagulation analyzers, the CS-5100 features several technical refinements including a pre-analytical assessment unit as well as multi-wavelength optical detection units. Therefore, the CS-5100 is supposed to rapidly and accurately perform a broad panel of coagulation tests. In the current study, the CS-5100 was evaluated regarding its precision and practicability in a clinical laboratory setting. Methods: The CS-5100 was evaluated regarding its intra and inter-assay precision using commercially available control samples. Results of patient samples, including hemolytic, icteric and lipemic specimens, measured on the CS-5100 were compared to reference analyzers, which are used in our accredited laboratory. Results: The coefficients of variation, assessed in the intra-and inter-assay precision analyses were below 5% representatively for most parameters. Results, obtained by the CS-5100 showed predominantly a high comparability to used reference analyzers, with correlation coefficients ranging from 0.857 to 0.990. Only minor ranged systemic or proportional differences were found in Passing-Bablok regression between the CS-5100 and reference analyzers regarding most of the tested parameters. Lipemic samples had a tendency to deteriorate correlation coefficients, but an overall effect of the sample's triglyceride level could be ruled out. In a routine setting, the analyzer reached a sample throughput rate of 160 tests per hour. Conclusions: The CS-5100 is able to rapidly and precisely measure patient samples. No considerable influence on test comparability was found for elevated levels of free hemoglobin, bilirubin or triglycerides.
引用
收藏
页码:1193 / 1202
页数:10
相关论文
共 26 条
[1]   Evaluation and performance characteristics of the automated coagulation analyzer ACL TOP [J].
Appert-Flory, Anny ;
Fischer, Florence ;
Jambou, Didier ;
Toulon, Pierre .
THROMBOSIS RESEARCH, 2007, 120 (05) :733-743
[2]  
Bilic-Zulle L, 2011, BIOCHEM MEDICA, V21, P49
[3]  
Bland JM, 1999, STAT METHODS MED RES, V8, P135, DOI 10.1177/096228029900800204
[4]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[5]   Laboratory evaluation of the Coasys® Plus C coagulation analyzer [J].
de Bie, Prim ;
Schornagel, Willem J. ;
van den Dool, Erik-Jan ;
Bakker, Bianca ;
van Dam, Willem ;
Heckman, Marion ;
Agar, Cetin ;
Sturk, Auguste ;
Stroobants, An K. .
THROMBOSIS RESEARCH, 2013, 131 (04) :357-362
[6]   Preanalytical and Postanalytical Variables: The Leading Causes of Diagnostic Error in Hemostasis? [J].
Favaloro, Emmanuel J. ;
Lippi, Giuseppe ;
Adcock, Dorothy M. .
SEMINARS IN THROMBOSIS AND HEMOSTASIS, 2008, 34 (07) :612-634
[7]   TESTS FOR RANK CORRELATION COEFFICIENTS .1. [J].
FIELLER, EC ;
HARTLEY, HO ;
PEARSON, ES .
BIOMETRIKA, 1957, 44 (3-4) :470-481
[8]   Evaluation of the automated coagulation analyzer Sysmex® CA-7000 [J].
Fischer, F ;
Appert-Flory, A ;
Jambou, D ;
Toulon, P .
THROMBOSIS RESEARCH, 2006, 117 (06) :721-729
[9]  
Fisher RA, 1914, BIOMETRIKA, V10, P507
[10]  
Flanders MM, 2002, CLIN CHEM, V48, P1622