The impact of patient table on size-specific dose estimate (SSDE)

被引:28
|
作者
Anam, Choirul [1 ,2 ]
Haryanto, Freddy [2 ]
Widita, Rena [2 ]
Arif, Idam [2 ]
Dougherty, Geoff [3 ]
McLean, Donald [4 ]
机构
[1] Diponegoro Univ, Dept Phys, Fac Math & Nat Sci, Semarang 50275, Central Java, Indonesia
[2] Bandung Inst Technol, Dept Phys, Fac Math & Nat Sci, Ganesha 10, Bandung 40132, West Java, Indonesia
[3] CSUCI, Appl Phys & Med Imaging, Camarillo, CA 93012 USA
[4] Canberra Hosp, Med Phys & Radiat Engn, Garran, ACT 2605, Australia
关键词
Water equivalent diameter (D-W); Size-specific dose estimates (SSDE); Volume CT dose index (CTDIVOL); Patient dose; Patient table; ATTENUATION-BASED ESTIMATION; RAY COMPUTED-TOMOGRAPHY; PEDIATRIC CHEST CT; INDEX; IMPLEMENTATION; DIAMETER; MEDICINE; PURPOSE; CANCER; RISKS;
D O I
10.1007/s13246-016-0497-z
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The use of parameters water equivalent diameter (D-W) and size-specific dose estimate (SSDE) are becoming increasingly established as a recognised method to relate patient dose from a CT examination to the dose indicator volume CT dose index (CTDIVOL). However, the role of the attenuation due to the patient table in these estimations requires careful consideration and is the subject of this study. The aim of this study is to investigate the impact of a minimal part of the patient table when calculating the D-W and SSDE. We investigated 164 patients who had undergone CT examinations for the pelvis, abdomen, thorax and head. We subsequently calculated D-W and SSDE using two methods: one using a small circular region of interest (ROI) including a minimal part of the patient table and the other using a ROI fitted to the patient border alone. The results showed that the water equivalent diameter calculated with the table included in the ROI (D-W,D- t) is greater, compared to that without the consideration of the patient table (D-W,D- nt), by 1.5-6.2% depending on the anatomy being imaged. On the other hand, the SSDE calculated with inclusion of the patient table (SSDEt) is smaller than otherwise (SSDEnt) by 1.0-5.5% again depending on the anatomy being imaged. The effect of the patient table on D-W and SSDE in the thorax CT examination was statistically significant, but its effect on D-W and SSDE in the other examinations of head, pelvis and abdomen was relatively small and not statistically significant.
引用
收藏
页码:153 / 158
页数:6
相关论文
共 50 条
  • [21] Estimating Patient Dose from X-ray Tube Output Metrics: Automated Measurement of Patient Size from CT Images Enables Large-scale Size-specific Dose Estimates
    Ikuta, Ichiro
    Warden, Graham I.
    Andriole, Katherine P.
    Khorasani, Ramin
    Sodickson, Aaron
    RADIOLOGY, 2014, 270 (02) : 472 - 480
  • [22] Influence of cone beam CT (CBCT) scan parameters on size specific dose estimate (SSDE): a Monte Carlo study
    Abuhaimed, Abdullah
    Martin, Colin J.
    Demirkaya, Omer
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (11):
  • [23] Can Patient's Body Weight Represent Body Diameter for Pediatric Size-Specific Dose Estimate in Thoracic and Abdominal Computed Tomography?
    Kritsaneepaiboon, Supika
    Eng-chuan, Suwadee
    Yoykaew, Saowapark
    JOURNAL OF CLINICAL IMAGING SCIENCE, 2019, 9 : 1 - 8
  • [24] Size-Specific Dose Estimates Based on Water-Equivalent Diameter and Effective Diameter in Computed Tomography Coronary Angiography
    Xu, Jian
    Wang, Xiangquan
    Xiao, Huawei
    Xu, Jianguo
    MEDICAL SCIENCE MONITOR, 2019, 25 : 9299 - 9305
  • [25] CNN application for automated determination of the patient's size to obtain the size-specific dose estimated in CT
    Hernandez-Davila, Erik R.
    Torres-Garcia, Eugenio
    Aranda-Lara, Liliana
    Roldan-Valadez, Ernesto
    Isaac-Olive, Keila
    Flores-Reyes, Mario
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2024, 10 (03)
  • [26] A SIMPLE METHOD FOR CALIBRATING PIXEL VALUES OF THE CT LOCALIZER RADIOGRAPH FOR CALCULATING WATER-EQUIVALENT DIAMETER AND SIZE-SPECIFIC DOSE ESTIMATE
    Anam, Choirul
    Fujibuchi, Toshioh
    Toyoda, Takatoshi
    Sato, Naoki
    Haryanto, Freddy
    Widita, Rena
    Arif, Idam
    Dougherty, Geoff
    RADIATION PROTECTION DOSIMETRY, 2018, 179 (02) : 158 - 168
  • [27] Evolved size-specific dose estimates for patient-specific organ doses from chest CT scans based on hybrid patient size vectors
    Shao, Wencheng
    Yang, Ke
    Lou, Lizhi
    Lin, Xin
    Qu, Liangyong
    Zhuo, Weihai
    Liu, Haikuan
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2025,
  • [28] Simplifying Size-Specific Radiation Dose Estimates in Pediatric CT
    Khawaja, Ranish Deedar Ali
    Singh, Sarabjeet
    Vettiyil, Beth
    Lim, Ruth
    Gee, Michael
    Westra, Sjirk
    Kalra, Mannudeep K.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2015, 204 (01) : 167 - 176
  • [29] An improved method for automated calculation of the water-equivalent diameter for estimating size-specific dose in CT
    Anam, Choirul
    Mahdani, Fahmi Rosydiansyah
    Dewi, Winda Kusuma
    Sutanto, Heri
    Triadyaksa, Pandji
    Haryanto, Freddy
    Dougherty, Geoff
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (09): : 313 - 323
  • [30] A Segmentation-based Automated Calculation of Patient Size and Size-specific Dose Estimates in Pediatric Computed Tomography Scans
    Abdulkadir, Muhammad Kabir
    Osman, Noor Diyana
    Achuthan, Anusha
    Nasirudin, Radin A.
    Ahmad, Muhammad Zabidi
    Zain, Noor Hasyima Mat
    Shuaib, Ibrahim Lutfi
    JOURNAL OF MEDICAL PHYSICS, 2024, 49 (03) : 456 - 463