Magneto-Optical Tracking of Flexible Laparoscopic Ultrasound: Model-Based Online Detection and Correction of Magnetic Tracking Errors

被引:51
作者
Feuerstein, Marco [1 ]
Reichl, Tobias [2 ]
Vogel, Jakob [2 ]
Traub, Joerg [2 ]
Navab, Nassir [2 ]
机构
[1] Nagoya Univ, Dept Media Sci, Grad Sch Informat Sci, Nagoya, Aichi 4648603, Japan
[2] Tech Univ Munich, Chair Comp Aided Med Procedures & Augmented Real, D-85748 Garching, Germany
关键词
Electromagnetic tracking; hybrid tracking; image-guided surgery; laparoscopic surgery; optical tracking; CALIBRATION; NAVIGATION; SURGERY; ULTRASONOGRAPHY; SYSTEM;
D O I
10.1109/TMI.2008.2008954
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
Electromagnetic tracking is currently one of the most promising means of localizing flexible endoscopic instruments such as flexible laparoscopic ultrasound transducers. However, electromagnetic tracking is also susceptible to interference from ferro-magnetic material, which distorts the magnetic field and leads to tracking errors. This paper presents new methods for real-time on-line detection and reduction of dynamic electromagnetic tracking errors when localizing a flexible laparoscopic ultrasound transducer. We use a hybrid tracking setup to combine optical tracking of the transducer shaft and electromagnetic tracking of the flexible transducer tip. A novel approach of modeling the poses of the transducer tip in relation to the transducer shaft allows us to reliably detect and significantly reduce electromagnetic tracking errors. For detecting errors of more than 5 rum, we achieved a sensitivity and specificity of 91% and 93%, respectively. Initial 3-D rms error of 6.91 mm were reduced to 3.15 mm.
引用
收藏
页码:951 / 967
页数:17
相关论文
共 45 条
  • [1] [Anonymous], P INT C ROB AUT
  • [2] A tracker alignment framework for augmented reality
    Baillot, Y
    Julier, SJ
    Brown, D
    Livingston, MA
    [J]. SECOND IEEE AND ACM INTERNATIONAL SYMPOSIUM ON MIXED AND AUGMENTED REALITY, PROCEEDINGS, 2003, : 142 - 150
  • [3] Calibration of tracking systems in a surgical environment
    Birkfellner, W
    Watzinger, F
    Wanschitz, F
    Ewers, R
    Bergmann, H
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1998, 17 (05) : 737 - 742
  • [4] Climent J., 2004, ELECT LETT COMPUTER, V4, P21, DOI DOI 10.5565/REV/ELCVIA.70
  • [5] Hand-eye calibration using dual quaternions
    Daniilidis, K
    [J]. INTERNATIONAL JOURNAL OF ROBOTICS RESEARCH, 1999, 18 (03) : 286 - 298
  • [6] DEBUCK S, LECT NOTES COMPUTER, V4791, P467
  • [7] Doignon C., 2006, European Conference on Computer Vision, P314
  • [8] Doucet A, 2001, STAT ENG IN, P3
  • [9] A new visualization technique for laparoscopic ultrasonography
    Ellsmere, J
    Stoll, M
    Wells, W
    Kikinis, R
    Vosburgh, K
    Kane, R
    Brooks, D
    Rattner, D
    [J]. SURGERY, 2004, 136 (01) : 84 - 92
  • [10] FEUERSTEIN M, P INT C MED IM COMP, V4791, P458