The effect of operational speed on the fracture of nickel-titanium rotary instruments

被引:46
|
作者
Kitchens, G. Gray, Jr. [1 ]
Liewebr, Frederick R. [1 ]
Moon, Peter C. [1 ]
机构
[1] Virginia Commonwealth Univ, Richmond, VA USA
关键词
fracture; rotary; speed;
D O I
10.1016/j.joen.2006.09.004
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
The purpose of this study was to compare the number of rotations to fracture of nickel-titanium (NiTi) rotary files operated at different speeds and different angles. We used 60 Profile NiTi rotary files, size 25 (30 each of 0.04 and 0.06 taper), operated at speeds of 350 or 600 rpm at angles of 25, 28, and 33.5 degrees. The time to fracture and number of rotations to fracture were recorded and calculated. A significant (p < 0.001) difference was found in the number of rotations to fracture according to taper and angle. Files of 0.06 taper fractured more readily than files of 0.04 taper. Increasing the angle at which the file was rotated decreased the number of rotations to fracture for both tapers; 0.04-taper files were more affected by an increase in the angle than the 0.06-taper files. However, the number of rotations to fracture was not related to the speed at which the files were operated.
引用
收藏
页码:52 / 54
页数:3
相关论文
共 50 条
  • [41] Effect From Surface Treatment of Nickel-Titanium Rotary Files on the Fracture Resistance
    Kim, Bo Hoon
    Ha, Jung-Hong
    Lee, Woo Cheol
    Kwak, Sang-Won
    Kim, Hyeon-Cheol
    SCANNING, 2015, 37 (01) : 82 - 87
  • [42] Effect of Environment on Fatigue Failure of Controlled Memory Wire Nickel-Titanium Rotary Instruments
    Shen, Ya
    Qian, Wei
    Abtin, Houman
    Gao, Yuan
    Haapasalo, Markus
    JOURNAL OF ENDODONTICS, 2012, 38 (03) : 376 - 380
  • [43] Effect of nickel-titanium alloys on root canal preparation and on mechanical properties of rotary instruments
    Vivan, Rodrigo Ricci
    Sanches Marques Costa, Vanessa Abreu
    Conti, Leticia Citelli
    Duque, Jussaro Alves
    Bramante, Clovis Monteiro
    So, Marcus Vinicius Reis
    Alcalde, Murilo Priori
    Hungaro Duarte, Marco Antonio
    BRAZILIAN ORAL RESEARCH, 2022, 36
  • [44] Effect of clinical use on the cyclic fatigue resistance of ProTaper nickel-titanium rotary instruments
    Ounsi, Hani F.
    Salameh, Ziad
    Al-Shalan, Thakib
    Ferrari, Marco
    Grandini, Simone
    Pashley, David H.
    Tay, Franklin R.
    JOURNAL OF ENDODONTICS, 2007, 33 (06) : 737 - 741
  • [45] Factors influencing the fracture of rotary nickel titanium instruments
    Saber, Shehab El-Din Mohamed
    ENDO-ENDODONTIC PRACTICE TODAY, 2008, 2 (04): : 273 - 283
  • [46] Deformation and fracture of K3 rotary nickel-titanium endodontic instruments after clinical use
    Shen, S. M.
    Deng, M.
    Wang, P. P.
    Chen, X. M.
    Zheng, L. W.
    Li, H. L.
    INTERNATIONAL ENDODONTIC JOURNAL, 2016, 49 (11) : 1088 - 1094
  • [47] Cutting efficiency of five different types of rotary nickel-titanium instruments
    Schaefer, Edgar
    Oitzinger, Mirjam
    JOURNAL OF ENDODONTICS, 2008, 34 (02) : 198 - 200
  • [48] Cyclic Fatigue of Different Nickel-Titanium Rotary Instruments: A Comparative Study
    Testarelli, L.
    Grande, N. M.
    Plotino, G.
    Lendini, M.
    Pongione, G.
    De Paolis, G.
    Rizzo, F.
    Milana, V.
    Gambarini, G.
    OPEN DENTISTRY JOURNAL, 2009, 3 : 55 - 58
  • [49] Accuracy of the Nouvag torque control motor for nickel-titanium rotary instruments
    Yared, G
    Kulkarni, GK
    ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTICS, 2004, 97 (04): : 499 - 501
  • [50] Evaluation of dynamic and static torsional resistances of nickel-titanium rotary instruments
    Abu-Tahun, Ibrahim H.
    Ha, Jung-Hong
    Kwak, Sang Won
    Kim, Hyeon-Cheol
    JOURNAL OF DENTAL SCIENCES, 2018, 13 (03) : 207 - 212