Gradient moment smoothing: A new flow compensation technique for multi-shot echo-planar imaging

被引:10
作者
Slavin, GS [1 ]
Riederer, SJ [1 ]
机构
[1] MAYO CLIN & MAYO FDN, MAGNET RESONANCE LAB, DEPT DIAGNOST RADIOL, ROCHESTER, MN 55905 USA
关键词
echo-planar imaging; EPI; flow artifact; fast MR imaging; model; mathematical;
D O I
10.1002/mrm.1910380304
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This work identifies an additional source of phase error across k(y) in multi-shot echo-planar imaging resulting from flow or motion along the phase-encoding direction, A velocity-independent flow compensation technique, gradient moment smoothing, is presented that corrects this error by forcing the phase to have smooth quadratic behavior, The correction is implemented, without compromising scan time, by changing the first moment of a bipolar prephaser pulse on a shot-by-shot basis, In phantom and in vivo experiments, gradient moment smoothing effectively eliminates ghosting and signal loss due to phase-encoding flow, When used in conjunction with a ''flyback'' echo-planar readout, which compensates for flow in the frequency-encoding direction, gradient moment smoothing renders multi-shot echo-planar imaging relatively insensitive to in-plane flow. This can make multi-shot echo-planar imaging a viable technique for accurately imaging inplane flow and may desensitize it to the otherwise serious problem of in-plane motion.
引用
收藏
页码:368 / 377
页数:10
相关论文
共 21 条
[1]   ANALYSIS OF FLOW EFFECTS IN ECHO-PLANAR IMAGING [J].
BUTTS, K ;
RIEDERER, SJ .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1992, 2 (03) :285-293
[2]   ECHO-PLANAR IMAGING OF THE LIVER WITH A STANDARD MR-IMAGING SYSTEM [J].
BUTTS, K ;
RIEDERER, SJ ;
EHMAN, RL ;
FELMLEE, JP ;
GRIMM, RC .
RADIOLOGY, 1993, 189 (01) :259-264
[3]   THEORETICAL ASPECTS OF MOTION SENSITIVITY AND COMPENSATION IN ECHO-PLANAR IMAGING [J].
DUERK, JL ;
SIMONETTI, OP .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1991, 1 (06) :643-650
[4]   ECHO-PLANAR MR-IMAGING [J].
EDELMAN, RR ;
WIELOPOLSKI, P ;
SCHMITT, F .
RADIOLOGY, 1994, 192 (03) :600-612
[5]   ANALYSIS OF T2 LIMITATIONS AND OFF-RESONANCE EFFECTS ON SPATIAL-RESOLUTION AND ARTIFACTS IN ECHO-PLANAR IMAGING [J].
FARZANEH, F ;
RIEDERER, SJ ;
PELC, NJ .
MAGNETIC RESONANCE IN MEDICINE, 1990, 14 (01) :123-139
[6]  
FARZANEH F, 1989, P SMRI 7 ANN M, P70
[7]   PHASE ERRORS IN MULTISHOT ECHO-PLANAR IMAGING [J].
FEINBERG, DA ;
OSHIO, K .
MAGNETIC RESONANCE IN MEDICINE, 1994, 32 (04) :535-539
[8]   ECHO-PLANAR IMAGING WITH ASYMMETRIC GRADIENT MODULATION AND INNER-VOLUME EXCITATION [J].
FEINBERG, DA ;
TURNER, R ;
JAKAB, PD ;
VONKIENLIN, M .
MAGNETIC RESONANCE IN MEDICINE, 1990, 13 (01) :162-169
[9]   GRADIENT-ECHO SHIFTING IN FAST MRI TECHNIQUES (GRASE IMAGING) FOR CORRECTION OF FIELD INHOMOGENEITY ERRORS AND CHEMICAL-SHIFT [J].
FEINBERG, DA ;
OSHIO, K .
JOURNAL OF MAGNETIC RESONANCE, 1992, 97 (01) :177-183
[10]   ECHO-PLANAR HIGH-RESOLUTION FLOW VELOCITY MAPPING [J].
FIRMIN, DN ;
KLIPSTEIN, RH ;
HOUNSFIELD, GL ;
PALEY, MP ;
LONGMORE, DB .
MAGNETIC RESONANCE IN MEDICINE, 1989, 12 (03) :316-327