Modeling micropatterned antigen-antibody binding kinetics in a microfluidic chip

被引:114
作者
Hu, Guoqing
Gao, Yali
Li, Dongqing
机构
[1] Vanderbilt Univ, Dept Mech Engn, Nashville, TN 37235 USA
[2] Univ Toronto, Dept Mech & Ind Engn, Toronto, ON M5S 3G8, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
microfluidics; numerical simulation; reaction kinetics; immunoassay; electrokinetically driven flow;
D O I
10.1016/j.bios.2006.06.017
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
The reaction kinetics of antigen-antibody binding in the electrokinetically controlled microfluidic heterogeneous immunoassays has been investigated by numerical simulations. A two-dimensional computational model was employed to include the mass transport (convection and diffusion) and binding reaction between the antigen in the bulk flow and the immobilized antibody at the channel surface. The influence of the bulk velocity, the concentrations of the antibody and antigen, and the geometry of the microchips was studied for a variation of conditions and the guidance for designing of microfluidic immunoassay was provided. The model also shows that electrokinetically driven immunoassays have better reaction kinetics than pressure-driven ones, resulting from the plug-like velocity profile. Finally, a multi-patch immunoassay chip was analyzed and the reaction kinetics was optimized by rearranging the reaction patches at the channel surfaces. (c) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:1403 / 1409
页数:7
相关论文
共 33 条
[1]   Micro total analysis systems. 2. Analytical standard operations and applications [J].
Auroux, PA ;
Iossifidis, D ;
Reyes, DR ;
Manz, A .
ANALYTICAL CHEMISTRY, 2002, 74 (12) :2637-2652
[2]   Microfluidic immunosensor systems [J].
Bange, A ;
Halsall, HB ;
Heineman, WR .
BIOSENSORS & BIOELECTRONICS, 2005, 20 (12) :2488-2503
[3]   Convection during incubation of microplate solid phase immunoassay: Effects on assay response and variation [J].
Beumer, T ;
Haarbosch, P ;
Carpay, W .
ANALYTICAL CHEMISTRY, 1996, 68 (08) :1375-1380
[4]   QUANTITATIVE-ANALYSIS OF PROTEIN-INTERACTION WITH LIGANDS .2. ANALYSIS OF MACROMOLECULAR INTERACTIONS USING IMMOBILIZED LIGANDS [J].
CHAIKEN, I ;
ROSE, S ;
KARLSSON, R .
ANALYTICAL BIOCHEMISTRY, 1992, 201 (02) :197-210
[5]  
Chiem NH, 1998, CLIN CHEM, V44, P591
[6]   Patterned delivery of immunoglobulins to surfaces using microfluidic networks [J].
Delamarche, E ;
Bernard, A ;
Schmid, H ;
Michel, B ;
Biebuyck, H .
SCIENCE, 1997, 276 (5313) :779-781
[7]   Electrokinetically driven microfluidic chips with surface-modified chambers for heterogeneous immunoassays [J].
Dodge, A ;
Fluri, K ;
Verpoorte, E ;
de Rooij, NF .
ANALYTICAL CHEMISTRY, 2001, 73 (14) :3400-3409
[8]   Estimating rate constants in a convection-diffusion system with a boundary reaction [J].
Edwards, DA .
IMA JOURNAL OF APPLIED MATHEMATICS, 1999, 63 (01) :89-112
[9]   THE ESTIMATION OF THYROXINE IN HUMAN PLASMA BY AN ELECTROPHORETIC TECHNIQUE [J].
EKINS, RP .
CLINICA CHIMICA ACTA, 1960, 5 (04) :453-459
[10]   An electrokinetically-controlled immunoassay for simultaneous detection of multiple microbial antigens [J].
Gao, YL ;
Hu, GQ ;
Lin, FYH ;
Sherman, PM ;
Li, DQ .
BIOMEDICAL MICRODEVICES, 2005, 7 (04) :301-312