Periprosthetic DXA after total hip arthroplasty with short vs. ultra-short custom-made femoral stems 37 patients followed for 3 years

被引:63
作者
Albanese, Carlina V. [1 ]
Santori, Francesco S. [2 ]
Pavan, Laura [3 ]
Learmonth, Ian D. [4 ]
Passariello, Roberto [1 ]
机构
[1] Univ Roma La Sapienza, Dept Radiol Sci, Rome, Italy
[2] San Pietro Fatebenefratelli Hosp, Dept Orthopaed, Rome, Italy
[3] Johnson & Johnson Med SpA, DePuy Div, Pratica Di Mare, RM, Italy
[4] Southmead Hosp, Avon Orthopaed Ctr, Bristol, Avon, England
关键词
BONE-MINERAL DENSITY; X-RAY ABSORPTIOMETRY; STRAIN DISTRIBUTION; PROXIMAL FEMUR; IN-VITRO; IMPLANT; PROSTHESIS; PRESERVATION; DENSITOMETRY; PRECISION;
D O I
10.3109/17453670903074467
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background and purpose Dual-energy X-ray absorptiometry (DXA) analysis of the 7 periprosthetic Gruen zones is the most commonly used protocol to evaluate bone remodeling after the implantation of conventional femoral stems. We assessed the value of DXA after cementless primary total hip arthroplasty (THA) by comparing the effect of progressive shortening of the stem of two femoral implants on periprosthetic bone remodeling using a specifically developed protocol of analysis with 5 periprosthetic regions of interest (ROIs). Patients and methods Bone mineral density (BMD) was evaluated in 37 patients in the plateau stage, 3 years after THA. Two femoral implants featuring conceptually new designs and surgical technique were tested: types I and 2, characterized by extremely short stem and virtual absence of distal stem, respectively. Results We found that progressive shortening of the femoral stem produces more proximal loading, which effectively preserves metaphyseal bone stock and increases periprosthetic BMD in the medial ROIs over time. In the type 2 group, higher absolute BMD values were observed in medial ROIs 4 and 5. No differences were found in ROIs 1, 2, and 3. Interpretation This study shows the flexibility of DXA in adapting the protocol of periprosthetic analysis to the specific requirements of new implant designs, and it shows its high sensitivity in evaluation of the biological response of bone to changes in implant shape.
引用
收藏
页码:291 / 297
页数:7
相关论文
共 37 条
[1]   Changes in proximal femoral strain after insertion of uncemented standard and customised femoral stems [J].
Aamodt, A ;
Lund-Larsen, J ;
Eine, J ;
Andersen, E ;
Benum, P ;
Husby, OS .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2001, 83B (06) :921-929
[2]   Bone remodelling in THA: A comparative DXA scan study between conventional implants and a new stemless femoral component. A preliminary report [J].
Albanese, CV ;
Rendine, M ;
De Palma, F ;
Impagliazzo, A ;
Falez, F ;
Postacchini, F ;
Villani, C ;
Passariello, R ;
Santori, FS .
HIP INTERNATIONAL, 2006, 16 :S9-S15
[3]   Pattern of periprosthetic bone remodeling around stable uncemented tapered hip stems: A prospective 84-month follow-up study and a median 156-month cross-sectional study with DXA [J].
Aldinger, PR ;
Sabo, D ;
Pritsch, M ;
Thomsen, M ;
Mau, H ;
Ewerbeck, V ;
Breusch, SJ .
CALCIFIED TISSUE INTERNATIONAL, 2003, 73 (02) :115-121
[4]   Continuous bone loss around a tapered uncemented femoral stem -: A long-term evaluation with DEXA [J].
Boden, Henrik S. G. ;
Skoldenberg, Olof G. ;
Salemyr, Mats F. ;
Lundberg, Hans-Jerker ;
Adolphson, Per Y. .
ACTA ORTHOPAEDICA, 2006, 77 (06) :877-885
[5]   Changes in bone mineral density in the proximal femur after cementless total hip arthroplasty - A five-year longitudinal study [J].
Brodner, W ;
Bitzan, P ;
Lomoschitz, F ;
Krepler, P ;
Jankovsky, R ;
Lehr, S ;
Kainberger, F ;
Gottsauner-Wolf, F .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2004, 86B (01) :20-26
[6]   ECTOPIC OSSIFICATION FOLLOWING TOTAL HIP-REPLACEMENT - INCIDENCE AND A METHOD OF CLASSIFICATION [J].
BROOKER, AF ;
BOWERMAN, JW ;
ROBINSON, RA ;
RILEY, LH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1973, A 55 (08) :1629-1632
[7]   ACCURACY OF DEXA MEASUREMENT OF BONE-MINERAL DENSITY AFTER TOTAL HIP-ARTHROPLASTY [J].
COHEN, B ;
RUSHTON, N .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1995, 77B (03) :479-483
[8]   Changes in strain distribution of loaded proximal femora caused by different types of cementless femoral stems [J].
Decking, Ralf ;
Puhl, Wolfhart ;
Simon, Ulrich ;
Claes, Lutz E. .
CLINICAL BIOMECHANICS, 2006, 21 (05) :495-501
[9]  
GIBBONS CE, 1997, INT ORTHOP, V25, P89
[10]   CLOSENESS OF FIT OF UNCEMENTED STEMS IMPROVES THE STRAIN DISTRIBUTION IN THE FEMUR [J].
HUA, J ;
WALKER, PS .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1995, 13 (03) :339-346