Can a Single-Use and Patient-Specific Instrumentation Be Reliably Used in Primary Total Knee Arthroplasty? A Multicenter Controlled Study

被引:17
作者
Abane, Laurent [1 ]
Zaoui, Amine [1 ]
Anract, Philippe [1 ]
Lefevre, Nicolas [2 ]
Herman, Serge [2 ]
Hamadouche, Moussa [1 ]
机构
[1] CHU Cochin Port Royal, Clin Orthopaed Res Ctr, Dept Orthopaed & Reconstruct Surg, 27 Rue Faubourg St Jacques, F-75014 Paris, France
[2] Clin Sport, Paris, France
关键词
total knee arthroplasty; single use; instrumentation; patient specific; outcomes; COMPUTER-ASSISTED NAVIGATION; RANDOMIZED CONTROLLED-TRIAL; MECHANICAL AXIS ALIGNMENT; CONVENTIONAL INSTRUMENTATION; CORONAL ALIGNMENT; POSTOPERATIVE ALIGNMENT; COMPONENT ALIGNMENT; CLINICAL-TRIAL; CUTTING BLOCKS; REPLACEMENT;
D O I
10.1016/j.arth.2018.02.038
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: The aim of this controlled multicenter study is to evaluate the clinical and radiologic outcomes of primary total knee arthroplasty (TKA) using single-use fully disposable and patient-specific cutting guides (SU) and compare the results to those obtained with traditional patient-specific cutting guides (PSI) vs conventional instrumentation (CI). Methods: Seventy consecutive patients had their TKA performed using SU. They were compared to 140 historical patients requiring TKA that were randomized to have the procedure performed using PSI vs CI. The primary measure outcome was mechanical axis as measured on a standing long-leg radiograph using the hip-knee-ankle angle. Secondary outcome measures were Knee Society and Oxford knee scores, operative time, need for postoperative transfusion, and length of hospital stay. Results: The mean hip-knee-ankle value was 179.8 degrees (standard deviation [SD] 3.1 degrees), 179.2 degrees (SD 2.9 degrees), and 178.3 degrees (SD 2.5 degrees) in the CI, PSI and SU groups, respectively (P = .0082). Outliers were identified in 16 of 65 (24.6%), 15 of 67 (22.4%), and 14 of 70 (20.0%) knees in the CI, PSI, and SU group, respectively (P = .81). There was no significant difference in the clinical results (P = .29 and .19, respectively). Operative time, number of unit transfusion, and length of hospital stay were not significantly different between the 3 groups (P = .45,.31, and 0.98, respectively). Conclusion: The use of an SU in TKA provided similar clinical and radiologic results to those obtained with traditional PSI and CI. The potential economic advantages of single-use instrumentation in primary TKA require further investigation. (c) 2018 Elsevier Inc. All rights reserved.
引用
收藏
页码:2111 / 2118
页数:8
相关论文
共 50 条
  • [41] Patient-specific Instruments for Total Knee Arthroplasty
    Lachiewicz, Paul F.
    Henderson, Robert A.
    JOURNAL OF THE AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS, 2013, 21 (09) : 513 - 518
  • [42] Contact kinematics of patient-specific instrumentation versus conventional instrumentation for total knee arthroplasty
    Broberg, Jordan S.
    Naudie, Douglas D. R.
    Howard, James L.
    Vasarhelyi, Edward M.
    McCalden, Richard W.
    Teeter, Matthew G.
    KNEE, 2020, 27 (05) : 1501 - 1509
  • [43] Patient specific instrumentation in total knee arthroplasty: a state of the art
    Mattei, Lorenzo
    Pellegrino, Pietro
    Calo, Michel
    Bistolfi, Alessandro
    Castoldi, Filippo
    ANNALS OF TRANSLATIONAL MEDICINE, 2016, 4 (07)
  • [44] Stability and alignment do not improve by using patient-specific instrumentation in total knee arthroplasty: a randomized controlled trial
    Kosse, Nienke M.
    Heesterbeek, Petra J. C.
    Schimmel, Janneke J. P.
    van Hellemondt, Gijs G.
    Wymenga, Ate B.
    Defoort, Koen C.
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2018, 26 (06) : 1792 - 1799
  • [45] Patient-specific instrumentation in total knee arthroplasty: simpler, faster and more accurate than standard instrumentation—a randomized controlled trial
    João Vide
    Tânia Pinto Freitas
    Acácio Ramos
    Henrique Cruz
    João Paulo Sousa
    Knee Surgery, Sports Traumatology, Arthroscopy, 2017, 25 : 2616 - 2621
  • [46] Functional Outcome and Quality of Life after Patient-Specific Instrumentation in Total Knee Arthroplasty
    Chen, Jerry Yonggiang
    Chin, Pak Lin
    Tay, Darren Keng Jin
    Chia, Shi-Lu
    Lo, Ngai Nung
    Yeo, Seng Jin
    JOURNAL OF ARTHROPLASTY, 2015, 30 (10) : 1724 - 1728
  • [47] The impact of patient-specific instrumentation on unicompartmental knee arthroplasty: a prospective randomised controlled study
    Alvand, Abtin
    Khan, Tanvir
    Jenkins, Cathy
    Rees, Jonathan L.
    Jackson, William F.
    Dodd, Christopher A. F.
    Murray, David W.
    Price, Andrew J.
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2018, 26 (06) : 1662 - 1670
  • [48] Current Data Do Not Support Routine Use of Patient-Specific Instrumentation in Total Knee Arthroplasty
    Voleti, Pramod B.
    Hamula, Mathew J.
    Baldwin, Keith D.
    Lee, Gwo-Chin
    JOURNAL OF ARTHROPLASTY, 2014, 29 (09) : 1709 - 1712
  • [49] Restoring tibial obliquity for kinematic alignment in total knee arthroplasty: conventional versus patient-specific instrumentation
    Smolle, Maria Anna
    Koutp, Amir
    Clar, Clemens
    Leitner, Lukas
    Leithner, Andreas
    Sadoghi, Patrick
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2023, 143 (09) : 5867 - 5872
  • [50] Advantages of patient-specific cutting guides with disposable instrumentation in total knee arthroplasty: a case control study
    Kevin Moerenhout
    Behrang Allami
    Georgios Gkagkalis
    Olivier Guyen
    Brigitte M. Jolles
    Journal of Orthopaedic Surgery and Research, 16