Can a Single-Use and Patient-Specific Instrumentation Be Reliably Used in Primary Total Knee Arthroplasty? A Multicenter Controlled Study

被引:17
作者
Abane, Laurent [1 ]
Zaoui, Amine [1 ]
Anract, Philippe [1 ]
Lefevre, Nicolas [2 ]
Herman, Serge [2 ]
Hamadouche, Moussa [1 ]
机构
[1] CHU Cochin Port Royal, Clin Orthopaed Res Ctr, Dept Orthopaed & Reconstruct Surg, 27 Rue Faubourg St Jacques, F-75014 Paris, France
[2] Clin Sport, Paris, France
关键词
total knee arthroplasty; single use; instrumentation; patient specific; outcomes; COMPUTER-ASSISTED NAVIGATION; RANDOMIZED CONTROLLED-TRIAL; MECHANICAL AXIS ALIGNMENT; CONVENTIONAL INSTRUMENTATION; CORONAL ALIGNMENT; POSTOPERATIVE ALIGNMENT; COMPONENT ALIGNMENT; CLINICAL-TRIAL; CUTTING BLOCKS; REPLACEMENT;
D O I
10.1016/j.arth.2018.02.038
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: The aim of this controlled multicenter study is to evaluate the clinical and radiologic outcomes of primary total knee arthroplasty (TKA) using single-use fully disposable and patient-specific cutting guides (SU) and compare the results to those obtained with traditional patient-specific cutting guides (PSI) vs conventional instrumentation (CI). Methods: Seventy consecutive patients had their TKA performed using SU. They were compared to 140 historical patients requiring TKA that were randomized to have the procedure performed using PSI vs CI. The primary measure outcome was mechanical axis as measured on a standing long-leg radiograph using the hip-knee-ankle angle. Secondary outcome measures were Knee Society and Oxford knee scores, operative time, need for postoperative transfusion, and length of hospital stay. Results: The mean hip-knee-ankle value was 179.8 degrees (standard deviation [SD] 3.1 degrees), 179.2 degrees (SD 2.9 degrees), and 178.3 degrees (SD 2.5 degrees) in the CI, PSI and SU groups, respectively (P = .0082). Outliers were identified in 16 of 65 (24.6%), 15 of 67 (22.4%), and 14 of 70 (20.0%) knees in the CI, PSI, and SU group, respectively (P = .81). There was no significant difference in the clinical results (P = .29 and .19, respectively). Operative time, number of unit transfusion, and length of hospital stay were not significantly different between the 3 groups (P = .45,.31, and 0.98, respectively). Conclusion: The use of an SU in TKA provided similar clinical and radiologic results to those obtained with traditional PSI and CI. The potential economic advantages of single-use instrumentation in primary TKA require further investigation. (c) 2018 Elsevier Inc. All rights reserved.
引用
收藏
页码:2111 / 2118
页数:8
相关论文
共 50 条
  • [31] A comparison of conventional and patient-specific instruments in total knee arthroplasty
    Daniilidis, Kiriakos
    Tibesku, Carsten O.
    INTERNATIONAL ORTHOPAEDICS, 2014, 38 (03) : 503 - 508
  • [32] Patient-specific computed tomography based instrumentation in total knee arthroplasty: a prospective randomized controlled study
    Kotela, Andrzej
    Kotela, Ireneusz
    INTERNATIONAL ORTHOPAEDICS, 2014, 38 (10) : 2099 - 2107
  • [33] Comparison of clinical outcomes of VISIONAIRE patient-specific instrumentation with conventional instrumentation in total knee arthroplasty: a systematic literature review and meta-analysis
    Tibesku, Carsten O. O.
    Haas, Steven B. B.
    Saunders, Christopher
    Harwood, David A. A.
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2023, 143 (07) : 4379 - 4393
  • [34] Patient-Specific Instrumentation in Total Knee Arthroplasty Provides No Improvement in Component Alignment
    Stronach, Benjamin M.
    Pelt, Christopher E.
    Erickson, Jill A.
    Peters, Christopher L.
    JOURNAL OF ARTHROPLASTY, 2014, 29 (09) : 1705 - 1708
  • [35] Reducing Blood Loss in Bilateral Total Knee Arthroplasty with Patient-Specific Instrumentation
    Rathod, Parthiv A.
    Deshmukh, Ajit J.
    Cushner, Fred D.
    ORTHOPEDIC CLINICS OF NORTH AMERICA, 2015, 46 (03) : 343 - +
  • [36] Rotation in total knee arthroplasty: no difference between patient-specific and conventional instrumentation
    Parratte, Sebastien
    Blanc, Guillaume
    Boussemart, Thomas
    Ollivier, Matthieu
    Le Corroller, Thomas
    Argenson, Jean-Noel
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2013, 21 (10) : 2213 - 2219
  • [37] CT-based patient-specific instrumentation for total knee arthroplasty in over 700 cases: single-use instruments are as accurate as standard instruments
    Stefan Gaukel
    Raphael N. Vuille-dit-Bille
    Michel Schläppi
    Peter P. Koch
    Knee Surgery, Sports Traumatology, Arthroscopy, 2022, 30 : 447 - 455
  • [38] Unsatisfactory Accuracy as Determined by Computer Navigation of VISIONAIRE Patient-Specific Instrumentation for Total Knee Arthroplasty
    Lustig, Sebastien
    Scholes, Corey J.
    Oussedik, Sam I.
    Kinzel, Vera
    Coolican, Myles R. J.
    Parker, David A.
    JOURNAL OF ARTHROPLASTY, 2013, 28 (03) : 469 - 473
  • [39] Patient-specific instrumentation improved three-dimensional accuracy in total knee arthroplasty: a comparative radiographic analysis of 1257 total knee arthroplasties
    Pauzenberger, Leo
    Munz, Martin
    Brandl, Georg
    Frank, Julia K.
    Heuberer, Philipp R.
    Laky, Brenda
    Schwameis, Eva
    Anderl, Werner
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2019, 14 (01)
  • [40] Patient-Specific Instrumentation Does Not Improve Accuracy in Total Knee Arthroplasty
    Shen, Chong
    Tang, Zhi-Hong
    Hu, Jun-Zu
    Zou, Guo-Yao
    Xiao, Rong-Chi
    Yan, Dong-Xue
    ORTHOPEDICS, 2015, 38 (03) : E178 - E188