A method for multichannel dosimetry with EBT3 radiochromic films

被引:33
作者
Perez Azorin, Jose Fernando [1 ]
Ramos Garcia, Luis Isaac [2 ]
Marti-Climent, Josep M. [3 ]
机构
[1] Hosp Univ Cruces, Dept Med Phys, Baracaldo 48903, Bizkaia, Spain
[2] Univ Navarra, Dept Oncol, Univ Navarra Clin, Pamplona 31008, Navarre, Spain
[3] Univ Navarra, Dept Nucl Med, Univ Navarra Clin, Pamplona 31008, Navarre, Spain
关键词
radiochromic films; multichannel dosimetry; IMRT verification; QUANTITATIVE-EVALUATION; GAFCHROMIC EBT2; CALIBRATION; SCANNERS; THERAPY; PHOTON; RANGE;
D O I
10.1118/1.4871622
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: An improved method for multichannel dosimetry is presented. This method explicitly takes into account the information provided by the unexposed image of the film. Methods: The method calculates the dose by applying a couple of perturbations to the scanned dose, one dependent and the other independent on the color channel. The method has been compared with previous multichannel and two single channel methods (red and green) against measurements using two different tests: first, five percentage depth dose profiles covering a wide range of doses; second, the dose map perpendicular to the beam axis for a 15 x 15 cm(2) square field. Finally, the results of 30 IMRT quality assurances tests are presented. All tests have been evaluated using the gamma analysis. Results: The coefficient of variation was found to be similar for all methods in a wide range of doses. The results of the proposed method are more in agreement with the experimental measurements and with the treatment planning system. Furthermore, the differences in the mean gamma pass rates are statistically significant. Conclusions: The improved multichannel dosimetric method is able to remove many of the common disturbances usually present in radiochromic films and improves the gamma analysis results compared with the other three methods. (C) 2014 American Association of Physicists in Medicine.
引用
收藏
页数:10
相关论文
共 29 条
[1]   Energy dependence and dose response of Gafchromic EBT2 film over a wide range of photon, electron, and proton beam energies [J].
Arjomandy, Bijan ;
Tailor, Ramesh ;
Anand, Aman ;
Sahoo, Narayan ;
Gillin, Michael ;
Prado, Karl ;
Vicic, Milos .
MEDICAL PHYSICS, 2010, 37 (05) :1942-1947
[2]   Dosimetric characterization and use of GAFCHROMIC EBT3 film for IMRT dose verification [J].
Borca, Valeria Casanova ;
Pasquino, Massimo ;
Russo, Giuliana ;
Grosso, Pierangelo ;
Cante, Domenico ;
Sciacero, Piera ;
Girelli, Giuseppe ;
La Porta, Maria Rosa ;
Tofani, Santi .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (02) :158-171
[3]   Dosimetric properties of improved GafChromic films for seven different digitizers [J].
Devic, S ;
Seuntjens, J ;
Hegyi, G ;
Podgorsak, EB ;
Soares, CG ;
Kirov, AS ;
Ali, I ;
Williamson, JF ;
Elizondo, A .
MEDICAL PHYSICS, 2004, 31 (09) :2392-2401
[4]   Precise radiochromic film dosimetry using a flat-bed document scanner [J].
Devic, S ;
Seuntjens, J ;
Sham, E ;
Podgorsak, EB ;
Schmidtlein, CR ;
Kirov, AS ;
Soares, CG .
MEDICAL PHYSICS, 2005, 32 (07) :2245-2253
[5]   Sensitivity of linear CCD array based film scanners used for film dosimetry [J].
Devic, Slobodan ;
Wang, Yi-Zhen ;
Tomic, Nada ;
Podgorsak, Ervin B. .
MEDICAL PHYSICS, 2006, 33 (11) :3993-3996
[6]   Radiochromic film dosimetry: Past, present, and future [J].
Devic, Slobodan .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2011, 27 (03) :122-134
[7]   Optimizing the dynamic range extension of a radiochromic film dosimetry system [J].
Devic, Slobodan ;
Tomic, Nada ;
Soares, Christopher G. ;
Podgorsak, Ervin B. .
MEDICAL PHYSICS, 2009, 36 (02) :429-437
[8]   A method of increasing the film intrinsic robustness of radiochromic film dosimetry [J].
Ebongue, A. Ngwa ;
Srivastava, R. P. ;
Njock, M. G. Kwato ;
De Wagter, C. .
PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (22) :N445-N456
[9]   Clinical use of EBT model Gafchromic™ film in radiotherapy [J].
Fiandra, Christian ;
Ricardi, Umberto ;
Ragona, Riccardo ;
Anglesio, Silvia ;
Giglioli, Francesca Romana ;
Calamia, Elisa ;
Lucio, Francesco .
MEDICAL PHYSICS, 2006, 33 (11) :4314-4319
[10]   Comparison of Gafchromic EBT2 and EBT3 for patient-specific quality assurance: Cranial stereotactic radiosurgery using volumetric modulated arc therapy with multiple noncoplanar arcs [J].
Fiandra, Christian ;
Fusella, Marco ;
Giglioli, Francesca Romana ;
Filippi, Andrea Riccardo ;
Mantovani, Cristina ;
Ricardi, Umberto ;
Ragona, Riccardo .
MEDICAL PHYSICS, 2013, 40 (08)