Efficacy of computerized infrared imaging analysis to evaluate mammographically suspicious lesions

被引:59
|
作者
Parisky, YR
Sardi, A
Hamm, R
Hughes, K
Esserman, L
Rust, S
Callahan, K
机构
[1] Univ So Calif, Norris Canc Ctr, Los Angeles, CA 90033 USA
[2] St Agnes Healthcare, Baltimore, MD 21229 USA
[3] Providence Hosp, Washington, DC 20017 USA
[4] Lahey Clin Northshore, Peabody, MA 01960 USA
[5] Mt Sinai Med Ctr, Miami Beach, FL 33140 USA
[6] Battelle Mem Inst, Columbus, OH 43201 USA
[7] Computerized Thermal Imaging, Lake Oswego, OR 97035 USA
关键词
D O I
10.2214/ajr.180.1.1800263
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The purpose of this clinical trial was to determine the efficacy of a dynamic computerized infrared imaging system for distinguishing between benign and malignant lesions in patients undergoing biopsy on the basis of mammographic findings. SUBJECTS AND METHODS. A 4-year clinical trial was conducted at five institutions using infrared imaging of patients for whom breast biopsy had been recommended. The data from a blinded subject set were obtained in 769 subjects with 875 biopsied lesions resulting in 187 malignant and 688 benign findings. The infrared technique records a series of sequential images that provides an assessment of the infrared information in a mammographically identified area. The suspicious area is localized on the infrared image by the radiologist using mammograms, and an index of suspicion is determined, yielding a negative or positive result. RESULTS. In the 875 biopsied lesions, the index of suspicion resulted in a 97% sensitivity, a 14% specificity, a 95% negative predictive value, and a 24% positive predictive value. Lesions that were assessed as false-negative by infrared analysis were microcalcifications, so an additional analysis was performed in a subset excluding lesions described only as microcalcification. In this restricted subset of 448 subjects with 479 lesions and 110 malignancies, the index of suspicion resulted in a 99% sensitivity, an 18% specificity, a 99% negative predictive value, and a 27% positive predictive value. Analysis of infrared imaging performance in all 875 biopsied lesions revealed that specificity was statistically improved in dense breast tissue compared with fatty breast tissue. CONCLUSION. Infrared imaging offers a safe noninvasive procedure that would be valuable as an adjunct to mammography in determining whether a lesion is benign or malignant.
引用
收藏
页码:263 / 269
页数:7
相关论文
共 50 条
  • [21] Infrared Technology as a Predictor of Pathologic Outcome in Patients with Suspicious Imaging
    Hughes, J.
    Careaga, E.
    Collett, A.
    Frazier, T. G.
    ANNALS OF SURGICAL ONCOLOGY, 2009, 16 : 44 - 44
  • [22] Infrared imaging of skin lesions
    McIntosh, LM
    Mansfield, JR
    Jackson, M
    Crowson, AN
    Mantsch, HH
    VIBRATIONAL SPECTROSCOPY-BASED SENSOR SYSTEMS, 2002, 4577 : 226 - 229
  • [23] Magnetic resonance imaging-guided biopsy of mammographically and clinically occult breast lesions
    Isabelle Bedrosian
    James Schlencker
    Francis R. Spitz
    Susan G. Orel
    Douglas L. Fraker
    Linda S. Callans
    Mitchell Schnall
    Carol Reynolds
    Brian J. Czerniecki
    Annals of Surgical Oncology, 2002, 9 : 457 - 461
  • [24] Magnetic resonance imaging-guided biopsy of mammographically and clinically occult breast lesions
    Bedrosian, I
    Schlencker, J
    Spitz, FR
    Orel, SG
    Fraker, DL
    Callans, LS
    Schnall, M
    Reynolds, C
    Czerniecki, BJ
    ANNALS OF SURGICAL ONCOLOGY, 2002, 9 (05) : 457 - 461
  • [25] Ensembling CNNs for dermoscopic analysis of suspicious skin lesions
    Nie, Yali
    Ferro, Matteo
    Sommella, Paolo
    Carratu, Marco
    Cacciapuoti, Sara
    Di Leo, Giuseppe
    Lundgren, Jan
    Fabbrocini, Gabriella
    2021 IEEE INTERNATIONAL SYMPOSIUM ON MEDICAL MEASUREMENTS AND APPLICATIONS (IEEE MEMEA 2021), 2021,
  • [26] Accuracy of MR imaging in the work-up of suspicious breast lesions: A diagnostic meta-analysis
    Hrung, JM
    Sonnad, SS
    Schwartz, JS
    Langlotz, CP
    ACADEMIC RADIOLOGY, 1999, 6 (07) : 387 - 397
  • [27] HIGH-RESOLUTION MR IMAGING OF SUSPICIOUS BREAST-LESIONS
    OREL, SG
    SCHNALL, MD
    HOLLAND, GA
    YEH, IT
    TROUPIN, RH
    RADIOLOGY, 1992, 185 : 246 - 246
  • [28] Effectiveness of breast electrical impedance imaging for clinically suspicious breast lesions
    Daglar, G.
    Senol, K.
    Yakut, Z., I
    Yuksek, Y. N.
    Tutuncu, T.
    Tez, M.
    Yesiltepe, C. H.
    BRATISLAVA MEDICAL JOURNAL-BRATISLAVSKE LEKARSKE LISTY, 2016, 117 (09): : 505 - 510
  • [29] A prospective pilot clinical trial evaluating the utility of a dynamic near-infrared imaging device for characterizing suspicious breast lesions
    Ronald X Xu
    Donn C Young
    Jimmy J Mao
    Stephen P Povoski
    Breast Cancer Research, 9
  • [30] A prospective pilot clinical trial evaluating the utility of a dynamic near-infrared imaging device for characterizing suspicious breast lesions
    Xu, Ronald X.
    Young, Donn C.
    Mao, Jimmy J.
    Povoski, Stephen P.
    BREAST CANCER RESEARCH, 2007, 9 (06):