Is the early fragmentation of intrauterine devices caused by stress corrosion cracking?

被引:9
作者
Pereda, M. D. [2 ,3 ]
Farina, S. B. [1 ]
Fernandez Lorenzo, M. [2 ,3 ]
机构
[1] Univ Nacl Gral San Martin, Comis Nacl Energia Atom, Ctr Atom Constituyentes, Consejo Nacl Invest Cient & Tecn, RA-1650 Buenos Aires, DF, Argentina
[2] Natl Univ La Plata, INIFTA Inst Invest Fisicoquim Teor & Aplicadas, Fac Ciencias Exactas, La Plata, Buenos Aires, Argentina
[3] Natl Univ La Plata, Fac Ingn, RA-1900 La Plata, Argentina
关键词
Copper; Simulated uterine fluid; Intrauterine device; Corrosion; Stress cracking; COPPER CORROSION; UTERINE FLUIDS; BEHAVIOR; ALBUMIN; IONS; COMPLEXES; GLUCOSE; OXYGEN; PH;
D O I
10.1016/j.actbio.2009.04.033
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Copper wire is the main component of a type of intrauterine device used as a contraceptive. Its contraceptive effect is attributed to the copper ions released as a result of the dissolution of copper in the uterus. Even though 10-year intrauterine device life is estimated on the basis of the dissolution rate of copper measured in vivo and in vitro, some cases of breakdown or fragmentation of the copper wire after short periods of insertion (2-3 months) have been reported. Due to the possible existence of residual stresses as a consequence of the manufacturing process, stress corrosion cracking has been previously proposed as an explanation for the early ruptures. In the present work, the susceptibility of copper wires to stress corrosion cracking in simulated uterine fluids was investigated. Results indicate that early ruptures should not be attributed to stress corrosion cracking. They could be explained by considering the increase in corrosion rate under certain conditions (pH decrease during infections; changes in the concentration of organic components along the menstrual cycle; etc.) that reduces the wire section leading to the rupture of the specimen by overloading. (C) 2009 Acta Materialia Inc. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:3240 / 3246
页数:7
相关论文
共 27 条
[1]   Characterization of copper in uterine fluids of patients who use the copper T-380A intrauterine device [J].
Arancibia, V ;
Peña, C ;
Allen, HE ;
Lagos, G .
CLINICA CHIMICA ACTA, 2003, 332 (1-2) :69-78
[2]   Multi-metal binding site of serum albumin [J].
Bal, W ;
Christodoulou, J ;
Sadler, PJ ;
Tucker, A .
JOURNAL OF INORGANIC BIOCHEMISTRY, 1998, 70 (01) :33-39
[3]   Influence of oxygen, albumin and pH on copper dissolution in a simulated uterine fluid [J].
Bastidas, DM ;
Cano, E ;
Mora, EM .
EUROPEAN JOURNAL OF CONTRACEPTION AND REPRODUCTIVE HEALTH CARE, 2005, 10 (02) :123-130
[4]   Copper corrosion-simulated uterine solutions [J].
Bastidas, JM ;
Cano, E ;
Mora, N .
CONTRACEPTION, 2000, 61 (06) :395-399
[5]   Characterization of copper corrosion products originated in simulated uterine fluids and on packaged intrauterine devices [J].
Bastidas, JM ;
Mora, N ;
Cano, E ;
Polo, JL .
JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 2001, 12 (05) :391-397
[6]  
BEAVERS JA, 1996, STRESS CORROSION CRA, P211
[7]   HUMAN UTERINE FLUID, EXAMINED IN UNDILUTED SAMPLES FOR OSMOLARITY AND THE CONCENTRATIONS OF INORGANIC-IONS, ALBUMIN, GLUCOSE, AND UREA [J].
CASSLEN, B ;
NILSSON, B .
AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 1984, 150 (07) :877-881
[8]   Investigations of different carbohydrate anomers in copper(II) complexes with D-glucose, D-fructose, and D-galactose by Raman and EPR spectroscopy [J].
Cerchiaro, G ;
Sant'Ana, AC ;
Temperini, MLA ;
Ferreira, AMD .
CARBOHYDRATE RESEARCH, 2005, 340 (15) :2352-2359
[9]   RELEASE OF COPPER FROM COPPER-BEARING INTRAUTERINE CONTRACEPTIVE DEVICES [J].
CHANTLER, E ;
CRITOPH, F ;
ELSTEIN, M .
BRITISH MEDICAL JOURNAL, 1977, 2 (6082) :288-291
[10]   DEGRADATION OF THE COPPER-RELEASING INTRAUTERINE CONTRACEPTIVE DEVICE AND ITS SIGNIFICANCE [J].
CHANTLER, EN ;
SCOTT, K ;
FILHO, CI ;
ELSTEIN, M ;
FARAGHER, EB ;
LORIMER, GW ;
BROUGH, I .
BRITISH JOURNAL OF OBSTETRICS AND GYNAECOLOGY, 1984, 91 (02) :172-181