Clinical introduction of image lag correction for a cone beam CT system

被引:8
作者
Stankovic, Uros [1 ]
Ploeger, Lennert S. [1 ]
Sonke, Jan-Jakob [1 ]
van Herk, Marcel [1 ,2 ]
机构
[1] Netherlands Canc Inst, Dept Radiat Oncol, NL-1066 CX Amsterdam, Netherlands
[2] Univ Manchester, Inst Canc Sci, Manchester M20 4QL, Lancs, England
关键词
cone beam CT; digital X-ray imaging; flat panel detectors; image lag; image quality; FLAT-PANEL IMAGER; COMPUTED-TOMOGRAPHY; DETECTORS; SCATTER;
D O I
10.1118/1.4941015
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Image lag in the flat-panel detector used for Linac integrated cone beam computed tomography (CBCT) has a degrading effect on CBCT image quality. The most prominent visible artifact is the presence of bright semicircular structure in the transverse view of the scans, known also as radar artifact. Several correction strategies have been proposed, but until now the clinical introduction of such corrections remains unreported. In November 2013, the authors have clinically implemented a previously proposed image lag correction on all of their machines at their main site in Amsterdam. The purpose of this study was to retrospectively evaluate the effect of the correction on the quality of CBCT images and evaluate the required calibration frequency. Methods: Image lag was measured in five clinical CBCT systems (Elekta Synergy 4.6) using an in-house developed beam interrupting device that stops the x-ray beam midway through the data acquisition of an unattenuated beam for calibration. A triple exponential falling edge response was fitted to the measured data and used to correct image lag from projection images with an infinite response. This filter, including an extrapolation for saturated pixels, was incorporated in the authors' in-house developed clinical CBCT. reconstruction software. To investigate the short-term stability of the lag and associated parameters, a series of five image lag measurement over a period of three months was performed. For quantitative analysis, the authors have retrospectively selected ten patients treated in the pelvic region. The apparent contrast was quantified in polar coordinates for scans reconstructed using the parameters obtained from different dates with and without saturation handling. Results: Visually, the radar artifact was minimal in scans reconstructed using image lag correction especially when saturation handling was used. In patient imaging, there was a significant reduction of the apparent contrast from 43 +/- 16.7 to 15.5 +/- 11.9 HU without the saturation handling and to 9.6 +/- 12.1 HU with the saturation handling, depending on the date of the calibration. The image lag correction parameters were stable over a period of 3 months. The computational load was increased by approximately 10%, not endangering the fast in-line reconstruction. Conclusions: The lag correction was successfully implemented clinically and removed most image lag artifacts thus improving the image quality. Image lag correction parameters were stable for 3 months indicating low frequency of calibration requirements. (C) 2016 American Association of Physicists in Medicine.
引用
收藏
页码:1057 / 1064
页数:8
相关论文
共 16 条
  • [1] BEAM HARDENING IN X-RAY RECONSTRUCTIVE TOMOGRAPHY
    BROOKS, RA
    DICHIRO, G
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1976, 21 (03) : 390 - 398
  • [2] Analysis of the temporal response of computed tomography fluoroscopy
    Hsieh, J
    [J]. MEDICAL PHYSICS, 1997, 24 (05) : 665 - 675
  • [3] Flat-panel cone-beam computed tomography for image-guided radiation therapy
    Jaffray, DA
    Siewerdsen, JH
    Wong, JW
    Martinez, AA
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 53 (05): : 1337 - 1349
  • [4] Monte Carlo evaluation of scatter mitigation strategies in cone-beam CT
    Lazos, Dimitrios
    Williamson, Jeffrey F.
    [J]. MEDICAL PHYSICS, 2010, 37 (10) : 5456 - 5470
  • [5] An empirical method for lag correction in cone-beam CT
    Mail, N.
    Moseley, D. J.
    Siewerdsen, J. H.
    Jaffray, D. A.
    [J]. MEDICAL PHYSICS, 2008, 35 (11) : 5187 - 5196
  • [6] Comparison of ghosting effects for three commercial a-Si EPIDs
    McDermott, L. N.
    Nijsten, S. M. J. J. G.
    Sonke, J. -J.
    Partridge, M.
    van Herk, M.
    Mijnheer, B. J.
    [J]. MEDICAL PHYSICS, 2006, 33 (07) : 2448 - 2451
  • [7] Ploeger LS, 2014, RADIOTHER ONCOL, V111, pS101
  • [8] Lag measurement in an a-Se active matrix flat-panel imager
    Schroeder, C
    Stanescu, T
    Rathee, S
    Fallone, BG
    [J]. MEDICAL PHYSICS, 2004, 31 (05) : 1203 - 1209
  • [9] A ghost story: Spatio-temporal response characteristics of an indirect-detection flat-panel imager
    Siewerdsen, JH
    Jaffray, DA
    [J]. MEDICAL PHYSICS, 1999, 26 (08) : 1624 - 1641
  • [10] Cone-beam computed tomography with a flat-panel imager: Effects of image lag
    Siewerdsen, JH
    Jaffray, DA
    [J]. MEDICAL PHYSICS, 1999, 26 (12) : 2635 - 2647