Evaluation of the eclipse electron Monte Carlo dose calculation for small fields

被引:16
作者
Xu, Zhigang [1 ]
Walsh, Sarah E. [1 ]
Telivala, Tejas P. [1 ]
Meek, Allen G. [1 ]
Yang, Guozhen [2 ]
机构
[1] SUNY Stony Brook, Med Ctr, Dept Radiat Oncol, Stony Brook, NY 11794 USA
[2] Mem Sloan Kettering Canc Ctr, Commack, NY USA
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2009年 / 10卷 / 03期
关键词
electron beam; Monte Carlo; small fields; MU calculation; extended SSD; RADIOCHROMIC FILM;
D O I
10.1120/jacmp.v10i3.2834
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Varian Medical Systems (Palo Alto, CA) has implemented the Monte Carlo electron dose calculation algorithm (eMC) in the Eclipse treatment planning system. Previous algorithms for electron treatment planning were limited in their calculation ability for small field depth doses and monitor units. An old rule of thumb to approximate the limiting cutout size for an electron field was determined by the lateral scatter equilibrium and approximated by E (MeV)/2.5 in centimeters of water. In this study, we compared eMC calculations and measurements of depth doses, isodose distributions, and monitor units for several different energy and small field cutout size combinations at different SSDs. Measurements were made using EBT film (International Specialty Products, Wayne, NJ) and a PinPoint ion chamber (PTW-New York Corp., Hicksville, NY). Our results indicate that the eMC algorithm can accurately predict depth doses, isodose distributions, and monitor units (within 2.5%) for field sizes as small as 3.0 cm diameter for energies in the 6 to 20 MeV range at 100 cm SSD. Therefore, the previous energy dependent rule of thumb does not apply to the Eclipse electron Monte Carlo code. However, at extended SSDs (105-110 cm), the results show good agreement (within 4%) only for higher energies (12, 16, and 20 MeV) for a field size of 3 cm.
引用
收藏
页码:75 / 85
页数:11
相关论文
共 14 条
  • [11] Sternick E. S., 1978, Practical Aspects of Electron Beam Treatment Planning, P52
  • [12] Evaluation of GafChromic EBT prototype B for external beam dose verification
    Todorovic, M.
    Fischer, M.
    Cremers, F.
    Thom, E.
    Schmidt, R.
    [J]. MEDICAL PHYSICS, 2006, 33 (05) : 1321 - 1328
  • [13] Evaluation of GAFCHROMIC® EBT film for CyberKnife® dosimetry
    Wilcox, Ellen E.
    Daskalov, George M.
    [J]. MEDICAL PHYSICS, 2007, 34 (06) : 1967 - 1974
  • [14] Characterization and use of EBT radiochromic film for IMRT dose verification
    Zeidan, Omar A.
    Stephenson, Stacy Ann L.
    Meeks, Sanford L.
    Wagner, Thomas H.
    Willoughby, Twyla R.
    Kupelian, Patrick A.
    Langen, Katja M.
    [J]. MEDICAL PHYSICS, 2006, 33 (11) : 4064 - 4072