Evaluation of the eclipse electron Monte Carlo dose calculation for small fields

被引:16
作者
Xu, Zhigang [1 ]
Walsh, Sarah E. [1 ]
Telivala, Tejas P. [1 ]
Meek, Allen G. [1 ]
Yang, Guozhen [2 ]
机构
[1] SUNY Stony Brook, Med Ctr, Dept Radiat Oncol, Stony Brook, NY 11794 USA
[2] Mem Sloan Kettering Canc Ctr, Commack, NY USA
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2009年 / 10卷 / 03期
关键词
electron beam; Monte Carlo; small fields; MU calculation; extended SSD; RADIOCHROMIC FILM;
D O I
10.1120/jacmp.v10i3.2834
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Varian Medical Systems (Palo Alto, CA) has implemented the Monte Carlo electron dose calculation algorithm (eMC) in the Eclipse treatment planning system. Previous algorithms for electron treatment planning were limited in their calculation ability for small field depth doses and monitor units. An old rule of thumb to approximate the limiting cutout size for an electron field was determined by the lateral scatter equilibrium and approximated by E (MeV)/2.5 in centimeters of water. In this study, we compared eMC calculations and measurements of depth doses, isodose distributions, and monitor units for several different energy and small field cutout size combinations at different SSDs. Measurements were made using EBT film (International Specialty Products, Wayne, NJ) and a PinPoint ion chamber (PTW-New York Corp., Hicksville, NY). Our results indicate that the eMC algorithm can accurately predict depth doses, isodose distributions, and monitor units (within 2.5%) for field sizes as small as 3.0 cm diameter for energies in the 6 to 20 MeV range at 100 cm SSD. Therefore, the previous energy dependent rule of thumb does not apply to the Eclipse electron Monte Carlo code. However, at extended SSDs (105-110 cm), the results show good agreement (within 4%) only for higher energies (12, 16, and 20 MeV) for a field size of 3 cm.
引用
收藏
页码:75 / 85
页数:11
相关论文
共 14 条
  • [1] Precise radiochromic film dosimetry using a flat-bed document scanner
    Devic, S
    Seuntjens, J
    Sham, E
    Podgorsak, EB
    Schmidtlein, CR
    Kirov, AS
    Soares, CG
    [J]. MEDICAL PHYSICS, 2005, 32 (07) : 2245 - 2253
  • [2] First macro Monte Carlo based commercial dose calculation module for electron beam treatment planning - new issues for clinical consideration
    Ding, George X.
    Duggan, Dennis M.
    Coffey, Charles W.
    Shokrani, Parvaneh
    Cygler, Joanna E.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (11) : 2781 - 2799
  • [3] Hogstrom K.R., 1992, AAPM MONOGRAPH, V19, P900
  • [4] COLLIMATION OF HIGH-ENERGY ELECTRON-BEAMS
    LAX, I
    BRAHME, A
    [J]. ACTA RADIOLOGICA ONCOLOGY, 1980, 19 (03): : 199 - 207
  • [5] Evaluation of the gamma dose distribution comparison method
    Low, DA
    Dempsey, JF
    [J]. MEDICAL PHYSICS, 2003, 30 (09) : 2455 - 2464
  • [6] A technique for the quantitative evaluation of dose distributions
    Low, DA
    Harms, WB
    Mutic, S
    Purdy, JA
    [J]. MEDICAL PHYSICS, 1998, 25 (05) : 656 - 661
  • [7] MMC - A HIGH-PERFORMANCE MONTE-CARLO CODE FOR ELECTRON-BEAM TREATMENT PLANNING
    NEUENSCHWANDER, H
    MACKIE, TR
    RECKWERDT, PJ
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (04) : 543 - 574
  • [8] The use of improved radiochromic film for in vivo quality assurance of high dose rate brachytherapy
    Pai, S
    Reinstein, LE
    Gluckman, G
    Xu, ZG
    Weiss, T
    [J]. MEDICAL PHYSICS, 1998, 25 (07) : 1217 - 1221
  • [9] Comprehensive evaluation of a commercial macro Monte Carlo electron dose calculation implementation using a standard verification data set
    Popple, Richard A.
    Weinberg, Rebecca
    Antolak, John A.
    Ye, Sung-Joon
    Pareek, Prem N.
    Duan, Jun
    Shen, Sui
    Brezovich, Ivan A.
    [J]. MEDICAL PHYSICS, 2006, 33 (06) : 1540 - 1551
  • [10] Optical density dependence on postirradiation temperature and time for MD-55-2 type radiochromic film
    Reinstein, LE
    Gluckman, GR
    [J]. MEDICAL PHYSICS, 1999, 26 (03) : 478 - 484