A simple backprojection algorithm for 3D in vivo EPID dosimetry of IMRT treatments

被引:130
作者
Wendling, Markus [1 ]
McDermott, Leah N. [1 ]
Mans, Anton [1 ]
Sonke, Jan-Jakob [1 ]
van Herk, Marcel [1 ]
Mijnheer, Ben J. [1 ]
机构
[1] Netherlands Canc Inst, Dept Radiat Oncol, Antoni Van Leeuwenhoek Hosp, NL-1066 CX Amsterdam, Netherlands
关键词
cancer; computerised tomography; dosimetry; image reconstruction; intensity modulation; medical image processing; radiation therapy; tumours; 3-DIMENSIONAL DOSE RECONSTRUCTION; PORTAL IMAGING DEVICE; PRETREATMENT VERIFICATION; CONVOLUTION/SUPERPOSITION METHOD; TRANSIT DOSIMETRY; PHOTON SPECTRA; PROSTATE IMRT; MONTE-CARLO; IMAGES; RADIOTHERAPY;
D O I
10.1118/1.3148482
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Treatment plans are usually designed, optimized, and evaluated based on the total 3D dose distribution, motivating a total 3D dose verification. The purpose of this study was to develop a 2D transmission-dosimetry method using an electronic portal imaging device (EPID) into a simple 3D method that provides 3D dose information. In the new method, the dose is reconstructed within the patient volume in multiple planes parallel to the EPID for each gantry angle. By summing the 3D dose grids of all beams, the 3D dose distribution for the total treatment fraction is obtained. The algorithm uses patient contours from the planning CT scan but does not include tissue inhomogeneity corrections. The 3D EPID dosimetry method was tested for IMRT fractions of a prostate, a rectum, and a head-and-neck cancer patient. Planned and in vivo-measured dose distributions were within 2% at the dose prescription point. Within the 50% isodose surface of the prescribed dose, at least 97% of points were in agreement, evaluated with a 3D gamma method with criteria of 3% of the prescribed dose and 0.3 cm. Full 3D dose reconstruction on a 0.1x0.1x0.1 cm(3) grid and 3D gamma evaluation took less than 15 min for one fraction on a standard PC. The method allows in vivo determination of 3D dose-volume parameters that are common in clinical practice. The authors conclude that their EPID dosimetry method is an accurate and fast tool for in vivo dose verification of IMRT plans in 3D. Their approach is independent of the treatment planning system and provides a practical safety net for radiotherapy.
引用
收藏
页码:3310 / 3321
页数:12
相关论文
共 38 条
[11]   The long-term stability of amorphous silicon flat panel imaging devices for dosimetry purposes [J].
Louwe, RJW ;
McDermott, LN ;
Sonke, JJ ;
Tielenburg, R ;
Wendling, M ;
van Herk, MB ;
Mijnheer, BJ .
MEDICAL PHYSICS, 2004, 31 (11) :2989-2995
[12]   Three-dimensional dose reconstruction of breast cancer treatment using portal imaging [J].
Louwe, RJW ;
Damen, EMF ;
van Herk, M ;
Minken, AWH ;
Törzsök, O ;
Mijnheer, BJ .
MEDICAL PHYSICS, 2003, 30 (09) :2376-2389
[13]   Evaluation of the gamma dose distribution comparison method [J].
Low, DA ;
Dempsey, JF .
MEDICAL PHYSICS, 2003, 30 (09) :2455-2464
[14]   A technique for the quantitative evaluation of dose distributions [J].
Low, DA ;
Harms, WB ;
Mutic, S ;
Purdy, JA .
MEDICAL PHYSICS, 1998, 25 (05) :656-661
[15]  
MANS A, 2008, P 10 INT WORKS UNPUB, P118
[16]   A two-step algorithm for predicting portal dose images in arbitrary detectors [J].
McCurdy, BMC ;
Pistorius, S .
MEDICAL PHYSICS, 2000, 27 (09) :2109-2116
[17]   Clinical experience with EPID dosimetry for prostate IMRT pre-treatment dose verification [J].
McDermott, L. N. ;
Wendling, M. ;
van Asselen, B. ;
Stroom, J. ;
Sonke, J. -J. ;
van Herk, M. ;
Mijnheer, B. J. .
MEDICAL PHYSICS, 2006, 33 (10) :3921-3930
[18]   3D in vivo dose verification of entire hypo-fractionated IMRT treatments using an EPID and cone-beam CT [J].
McDermott, Leah N. ;
Wendling, Markus ;
Nijkamp, Jasper ;
Mans, Anton ;
Sonke, Jan-Jakob ;
Mijnheer, Ben J. ;
van Herk, Marcel .
RADIOTHERAPY AND ONCOLOGY, 2008, 86 (01) :35-42
[19]   Replacing pretreatment verification with in vivo EPID dosimetry for prostate IMRT [J].
McDermott, Leah N. ;
Wendling, Markus ;
Sonke, Jan-Jakob ;
van Herk, Marcel ;
Mijnheer, Ben J. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 67 (05) :1568-1577
[20]   Dose-response and ghosting effects of an amorphous silicon electronic portal imaging device [J].
McDermott, LN ;
Louwe, RJW ;
Sonke, JJ ;
van Herk, MB ;
Mijnheer, BJ .
MEDICAL PHYSICS, 2004, 31 (02) :285-295