A simple backprojection algorithm for 3D in vivo EPID dosimetry of IMRT treatments

被引:130
作者
Wendling, Markus [1 ]
McDermott, Leah N. [1 ]
Mans, Anton [1 ]
Sonke, Jan-Jakob [1 ]
van Herk, Marcel [1 ]
Mijnheer, Ben J. [1 ]
机构
[1] Netherlands Canc Inst, Dept Radiat Oncol, Antoni Van Leeuwenhoek Hosp, NL-1066 CX Amsterdam, Netherlands
关键词
cancer; computerised tomography; dosimetry; image reconstruction; intensity modulation; medical image processing; radiation therapy; tumours; 3-DIMENSIONAL DOSE RECONSTRUCTION; PORTAL IMAGING DEVICE; PRETREATMENT VERIFICATION; CONVOLUTION/SUPERPOSITION METHOD; TRANSIT DOSIMETRY; PHOTON SPECTRA; PROSTATE IMRT; MONTE-CARLO; IMAGES; RADIOTHERAPY;
D O I
10.1118/1.3148482
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Treatment plans are usually designed, optimized, and evaluated based on the total 3D dose distribution, motivating a total 3D dose verification. The purpose of this study was to develop a 2D transmission-dosimetry method using an electronic portal imaging device (EPID) into a simple 3D method that provides 3D dose information. In the new method, the dose is reconstructed within the patient volume in multiple planes parallel to the EPID for each gantry angle. By summing the 3D dose grids of all beams, the 3D dose distribution for the total treatment fraction is obtained. The algorithm uses patient contours from the planning CT scan but does not include tissue inhomogeneity corrections. The 3D EPID dosimetry method was tested for IMRT fractions of a prostate, a rectum, and a head-and-neck cancer patient. Planned and in vivo-measured dose distributions were within 2% at the dose prescription point. Within the 50% isodose surface of the prescribed dose, at least 97% of points were in agreement, evaluated with a 3D gamma method with criteria of 3% of the prescribed dose and 0.3 cm. Full 3D dose reconstruction on a 0.1x0.1x0.1 cm(3) grid and 3D gamma evaluation took less than 15 min for one fraction on a standard PC. The method allows in vivo determination of 3D dose-volume parameters that are common in clinical practice. The authors conclude that their EPID dosimetry method is an accurate and fast tool for in vivo dose verification of IMRT plans in 3D. Their approach is independent of the treatment planning system and provides a practical safety net for radiotherapy.
引用
收藏
页码:3310 / 3321
页数:12
相关论文
共 38 条
[1]   Three-dimensional portal image-based dose reconstruction in a virtual phantom for rapid evaluation of IMRT plans [J].
Ansbacher, W. .
MEDICAL PHYSICS, 2006, 33 (09) :3369-3382
[2]   RADIATION-FIELD EDGE-DETECTION IN PORTAL IMAGES [J].
BIJHOLD, J ;
GILHUIJS, KGA ;
VANHERK, M ;
MEERTENS, H .
PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (12) :1705-1710
[3]  
BJAUMLRNGARD BE, 1994, MED PHYS, V21, P1069, DOI DOI 10.1118/1.597349
[4]   New method to obtain the midplane dose using portal in vivo dosimetry [J].
Boellaard, R ;
Essers, M ;
van Herk, M ;
Mijnheer, BJ .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 41 (02) :465-474
[5]  
Catmull E., 1974, Computer Aided Geometric Design, P317, DOI [10.1016/B978-0-12-079050-0.50020-5, DOI 10.1016/B978-0-12-079050-0.50020-5]
[6]   In vivo dosimetry during external photon beam radiotherapy [J].
Essers, M ;
Mijnheer, BJ .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1999, 43 (02) :245-259
[7]   The application of transit dosimetry to precision radiotherapy [J].
Hansen, VN ;
Evans, PM ;
Swindell, W .
MEDICAL PHYSICS, 1996, 23 (05) :713-721
[8]   X-RAY SCATTER IN MEGAVOLTAGE TRANSMISSION RADIOGRAPHY - PHYSICAL CHARACTERISTICS AND INFLUENCE ON IMAGE QUALITY [J].
JAFFRAY, DA ;
BATTISTA, JJ ;
FENSTER, A ;
MUNRO, P .
MEDICAL PHYSICS, 1994, 21 (01) :45-60
[9]   Analytical considerations of beam hardening in medical accelerator photon spectra [J].
Kleinschmidt, C .
MEDICAL PHYSICS, 1999, 26 (09) :1995-1999
[10]   Three-dimensional heart dose reconstruction to estimate normal tissue complication probability after breast irradiation using portal dosimetry [J].
Louwe, R. J. W. ;
Wendling, M. ;
van Herk, M. B. ;
Mijnheer, B. J. .
MEDICAL PHYSICS, 2007, 34 (04) :1354-1363