Establishment of diagnostic reference levels in cardiac computed tomography

被引:11
作者
Rawashdeh, Mohammad [1 ]
Saade, Charbel [2 ]
Zaitoun, Maha [1 ]
Abdelrahman, Mostafa [1 ]
Brennan, Patrick [3 ,4 ]
Alewaidat, Haytham [1 ]
McEntee, Mark F. [5 ]
机构
[1] Jordan Univ Sci & Technol, Fac Appl Med Sci, Irbid, Jordan
[2] Univ Beirut, Dept Diagnost Radiol, Med Ctr, Beirut, Lebanon
[3] Univ Sydney, Fac Hlth Sci, Med Image Optimizat & Percept Grp MIOPeG, Sydney, NSW, Australia
[4] Univ Sydney, Brain & Mind Ctr, Sydney, NSW, Australia
[5] Univ Coll Cork, Discipline Diagnost Radiog, Brookfield Hlth Sci, UG 12 Aras Watson,Coll Rd, Cork, Ireland
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2019年 / 20卷 / 10期
关键词
CCT; CTDIvol; DLP; DRL; Jordan; DOSE REDUCTION; CT ANGIOGRAPHY;
D O I
10.1002/acm2.12711
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim of this study was to determine diagnostic reference levels (DRLs) for cardiac computed tomography (CCT) in Jordan. Volume computed tomography dose index (CTDIvol) and dose-length product (DLP) were collected from 228 CCTs performed at seven Jordanian hospitals specialized in cardiac CT. DRLs for cardiac CT were defined at the 75th percentile of CTDIvol and DLP. CTDIvol and DLP were collected from 30 successive cardiac CT in each center except for one center (18 scans). The 75th percentile of the CTDIvol and the DLP of the centers calculated from mixed retrospective and prospective gated modes were 47.74 milligray (mGy) and 1035 mGy/cm, respectively. This study demonstrated wide dose variations among the surveyed hospitals for cardiac CT scans; there was a 5.1-fold difference between the highest and lowest median DLP with a range of 223.2-1146.7 mGy/cm. Differences were associated with variations in the mAs and kVp. This study confirmed large variability in CTDIvol and DLP for cardiac CT scans; variation was associated with acquisition protocols and highlights the need for dose optimization. DRLs are proposed for CCT; there remains substantial potential for optimization of cardiac CT examinations for adults in Jordan.
引用
收藏
页码:181 / 186
页数:6
相关论文
共 31 条
  • [1] Assessment of Radiation Dose and Image Quality of Multidetector Computed Tomography
    Al Ewaidat, Haytham
    Zheng, Xiaoming
    Khader, Yousef
    Abdelrahman, Mostafa
    Alhasan, Mostafa Khaled Mustafa
    Rawashdeh, Mohammad Ahmmad
    Al Mousa, Dana Samir
    Alawneh, Khaled Zayed Ali
    [J]. IRANIAN JOURNAL OF RADIOLOGY, 2018, 15 (03)
  • [2] ESTABLISHING DIAGNOSTIC REFERENCE LEVELS FOR CARDIAC COMPUTED TOMOGRAPHY ANGIOGRAPHY IN SAUDI ARABIA
    Alhailiy, Ali B.
    Kench, Peter L.
    McEntee, Mark F.
    Brennan, Patrick C.
    Ryan, Elaine A.
    [J]. RADIATION PROTECTION DOSIMETRY, 2018, 181 (02) : 129 - 134
  • [3] [Anonymous], 2007, 2007 RECOMMENDATIONS
  • [4] Clarke R. H., 1993, Documents of the NRPB, V4, P1
  • [5] Diagnostic reference level of computed tomography (CT) in japan
    Fukushima, Yasuhiro
    Tsushima, Yoshito
    Takei, Hiroyuki
    Taketomi-Takahashi, Ayako
    Otake, Hidenori
    Endo, Keigo
    [J]. RADIATION PROTECTION DOSIMETRY, 2012, 151 (01) : 51 - 57
  • [6] Radiation Dose and Safety in Cardiac Computed Tomography
    Gerber, Thomas C.
    Kantor, Birgit
    McCollough, Cynthia H.
    [J]. CARDIOLOGY CLINICS, 2009, 27 (04) : 665 - +
  • [7] Hart D., 2002, Radiation exposure of the UK population from medical and dental x-ray examinations
  • [8] Estimated Radiation Dose Associated With Cardiac CT Angiography
    Hausleiter, Joerg
    Meyer, Tanja
    Hermann, Franziska
    Hadamitzky, Martin
    Krebs, Markus
    Gerber, Thomas C.
    McCollough, Cynthia
    Martinoff, Stefan
    Kastrati, Adnan
    Schoemig, Albert
    Achenbach, Stephan
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2009, 301 (05): : 500 - 507
  • [9] Prospective versus retrospective ECG-gated 64-detector coronary CT angiography: Assessment of image quality, stenosis, and radiation dose
    Hirai, Nobuhiko
    Horiguchi, Jun
    Fujioka, Chikako
    Kiguchi, Masao
    Yamamoto, Hideya
    Matsuura, Noriaki
    Kitagawa, Toshiro
    Teragawa, Hiroki
    Kohno, Nobuoki
    Ito, Katsuhide
    [J]. RADIOLOGY, 2008, 248 (02) : 424 - 430
  • [10] IPSM N CoR Institute of Physical Sciences in Medicine National Radiological Protection Board and College of Radiographers, 1992, NAT PROT PAT DOS MEA