Comparative biomechanical study of three types of osteosynthesis used on supra and intercondylar fractures of the distal end of the humerus in adult

被引:0
|
作者
Fornasieri, C [1 ]
Staub, C [1 ]
Tourne, Y [1 ]
Rumelhart, C [1 ]
Saragaglia, D [1 ]
机构
[1] CHU GRENOBLE,HOP SUD,SERV CHIRURG ORTHOPED & TRAUMATOL SPORT,F-38130 ECHIROLLES,FRANCE
来源
REVUE DE CHIRURGIE ORTHOPEDIQUE ET REPARATRICE DE L APPAREIL MOTEUR | 1997年 / 83卷 / 03期
关键词
D O I
暂无
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Purpose of the study Supra and intercondylar fractures are the most common fractures of the distal end of the humerus in adult. An osteosynthesis consisting of a plate is the treatment of choice. But location and type of plate always remain open for debate. The authors present the results of an in vitro biomechanical study, which compared the stiffness of three types of osteosynthesis commonly used in these fractures. Material and methods The devices were on one hand the premolded lateral plate of Lecestre and Dupont (Howmedica(R)) used alone or in conjunction with a medial 1/3 tubular plate of the AO group, and on the other hand a posterior plate, of which we are developping a new model, the Lambda plate (Protek(R)) ''Y''-shaped, monoblock, flat and molded on the humerus during operation. The study compared these different methods of fixation on fresh human humeri. In a first part, the posterior plate was compared to the single lateral one; in a second part, the, posterior plate was compared to the coupled lateral and medial plates. The three plates were made of identical material. Both studies used eight pairs of bones with supra and intercondylar fractures realized by sawing. Each bane of a pair was fixed with one of the two types of device. Each humerus underwent different loading forces; sagittal bending (anterior and posterior) and torsion. The displacements were recorded using a calibrated measuring device coupled to two displacement sensors. The stiffness was calculated on force/displacement curves. Results In anterior bending, the Lambda posterior fixation was significantly stiffer than the Single lateral one (p < 0.05) (239 +/- 109 versus 129 +/- 65 N/mm), and was not significantly different of the bilateral fixation (229 +/- 93 versus 224 +/- 108 N/mm). In posterior bending, the mean stiffness of the Lambda fixation was not significantly different to that of the lateral fixation (91 +/- 27 versus 91 +/- 52 N/mm), and less than that of the bilateral one (130 +/- 39 Versus 170 +/- 70 N/mm), but not significantly. In torsion the mean stiffness of the Lambda fixation was superior to that of the lateral one (146.75 +/- 50.66 versus 119.75 +/- 58.8 Nm/rad), and bilateral one (233.31 +/- 107.47 versus 212.31 +/- 113.55 Nm/rad), but again not significantly. Discussion The ideal osteosynthesis for the fractures of the humeral distal doesn't exist, because the bane undergoes antero-posterior and posteroanterior cyclical forces during elbow flexion. Therefore the best device should be placed on both sides of the bone, but anatomical reasons make this location impossible. According to our study, we think the ''less worst'' device is the posterior one using the Lambda plate. Its symetrical design allows a best loading distribution on the two columns. Its thickness compensates for its posterior location and the short interval between two holes allows to put many screws (4 to 6) into the epiphysis. All supra and intercondylar fractures, comminuted or not, can be treated with this material. The single lateral device isn't stiff enough. Its stiffness mainly depends on the orientation of the oblique screw in the medial column. But the design of this column doesn't always allow for an optimal location of the screw, which is the reason of several failures. The bilateral device gives a stiffness, comparable to the posterior one, but doesn't allow as many screws as the Lambda plate in the epiphysis. Therefore, it can't be used in very distal fractures. Conclusion In spite of the progress of the material and the accuracy of the indications, the treatment of supra and intercondylar fractures of the distal end of the humerus is always a difficult problem. The devices have to be as stable as possible to allow an early motion. The best one should be placed on both sides of the distal humerus, because of the sagittal cyclic forces it undergoes, but this location is anatomically impossible. Our study concludes that the device using the Lambda plate is stiffer than the single lateral one and equivalent to the bilateral one. This plate isn't a new concept, but its innovation lies in its physical specifications (thickness, short interval between two holes), which allow for all supra and intercondylar fractures very stiff devices.
引用
收藏
页码:237 / 242
页数:6
相关论文
共 50 条
  • [31] Length of Plates and Number of Screws for the Fixation of Distal Humerus Fractures: A Finite Element Biomechanical Study
    Zarifian, Ahmadreza
    Fough, Ali Akbarinezhad
    Eygendaal, Denise
    Rivlin, Michael
    Shaegh, Seyed Ali Mousavi
    Kachooei, Amir R.
    JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2022, 47 (07): : 690.e1 - 690.e11
  • [32] Retrograde nailing of humeral shaft fractures: A biomechanical study of its effects on the strength of the distal humerus
    Strothman, D
    Templeman, DC
    Varecka, T
    Bechtold, J
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2000, 14 (02) : 101 - 104
  • [33] Angle-fixed plate fixation or double-plate osteosynthesis in fractures of the proximal humerus: a biomechanical study
    Hessmann, Martin H.
    Korner, Jan
    Hofmann, Alexander
    Sternstein, Werner
    Rommens, Pol M.
    BIOMEDIZINISCHE TECHNIK, 2008, 53 (03): : 130 - 137
  • [34] Plate selection for fixation of extra-articular distal humerus fractures: A biomechanical comparison of three different implants
    Mutlu, Serhat
    Mutlu, Harun
    Erdil, Mehmet
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2015, 46 (08): : 1696 - 1697
  • [35] Plate selection for fixation of extra-articular distal humerus fractures: A biomechanical comparison of three different implants
    Scolaro, John A.
    Hsu, Jason E.
    Svach, David J.
    Mehta, Samir
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2014, 45 (12): : 2040 - 2044
  • [36] THE MECHANICAL-PROPERTIES OF DEVICES USED FOR THE FIXATION OF FEMORAL-NECK FRACTURES - COMPARATIVE-STUDY OF DIFFERENT TYPES OF OSTEOSYNTHESIS
    KEMPF, I
    JAEGER, JH
    FREUND, J
    RENAULT, D
    BITAR, S
    KONSBRUCK, R
    BUTEL, J
    FAURE, C
    BONNEL, F
    REVUE DE CHIRURGIE ORTHOPEDIQUE ET REPARATRICE DE L APPAREIL MOTEUR, 1981, 67 (01): : 59 - 69
  • [37] Comparative study of three models of extra-articular distal humerus fracture osteosynthesis using the finite element method on an osteoporotic computational model
    Sabalic, Srecko
    Kodvanj, Janos
    Pavic, Arsen
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2013, 44 : S56 - S61
  • [38] Minimally invasive approach in radio distal end fractures, three different types of incisions
    Prieto, Fernando Xavier Romero
    Garcia, Carlos Raul Reyes
    JOURNAL OF SURGICAL CASE REPORTS, 2021, (05):
  • [39] A comparative biomechanical study of the Distal Tibia Nail against compression plating for the osteosynthesis of supramalleolar corrective osteotomies
    Greenfield, Julia
    Appelmann, Philipp
    Lafon, Yoann
    Bruyere-Garnier, Karine
    Rommens, Pol Maria
    Kuhn, Sebastian
    SCIENTIFIC REPORTS, 2021, 11 (01)
  • [40] A comparative biomechanical study of the Distal Tibia Nail against compression plating for the osteosynthesis of supramalleolar corrective osteotomies
    Julia Greenfield
    Philipp Appelmann
    Yoann Lafon
    Karine Bruyère-Garnier
    Pol Maria Rommens
    Sebastian Kuhn
    Scientific Reports, 11