A cross-sectional study to compare intraocular pressure measurement by sequential use of Goldman applanation tonometry, dynamic contour tonometry, ocular response analyzer, and Corvis ST

被引:14
作者
Tejwani, Sushma [1 ]
Dinakaran, Shoruba [1 ]
Joshi, Anuja [2 ]
Shetty, Rohit [3 ]
Roy, Abhijit Sinha [4 ]
机构
[1] Narayana Nethralaya, Dept Glaucoma, Bengaluru 560099, Karnataka, India
[2] Narayana Nethralaya, Dept Res, Bengaluru 560099, Karnataka, India
[3] Narayana Nethralaya, Dept Cornea & Refract Surg, Bengaluru 560099, Karnataka, India
[4] Narayana Nethralaya, Dept Imaging, Biomech & Math Modeling Solut Lab, Bengaluru 560099, Karnataka, India
关键词
Corneal biomechanics; corneal viscoelasticity; intraocular pressure; Scheimpflug noncontact tonometry; CORNEAL BIOMECHANICAL PROPERTIES; NONCONTACT TONOMETRY; GLAUCOMA; REPRODUCIBILITY; EYES;
D O I
10.4103/0301-4738.171956
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Objective: To study the correlation and effect of sequential measurement of intraocular pressure (IOP) with Goldmann applanation tonometer (GAT), ocular response analyzer (ORA), dynamic contour tonometer (DCT), and Corvis ST. Setting and Design: Observational cross-sectional series from the comprehensive clinic of a tertiary eye care center seen during December 2012. Methods: One hundred and twenty-five study eyes of 125 patients with normal IOP and biomechanical properties underwent IOP measurement on GAT, DCT, ORA, and Corvis ST; in four different sequences. Patients with high refractive errors, recent surgeries, glaucoma, and corneal disorders were excluded so as to rule out patients with evident altered corneal biomechanics. Statistical Analysis: Linear regression and Bland-Altman using MedCalc software. Results: Multivariate analysis of variance with repeated measures showed no influence of sequence of device use on IOP (P = 0.85). Linear regression r(2) between GAT and Corvis ST, Corvis ST and Goldmann-correlated IOP (IOPg), and DCT and Corvis ST were 0.37 (P = 0.675), 0.63 (P = 0.607), and 0.19 (P = 0.708), respectively. The Bland-Altman agreement of Corvis ST with GAT, corneal compensated IOP, and IOPg was 2 mmHg (-5.0 to + 10.3), -0.5 mmHg (-8.1 to 7.1), and 0.5 mmHg (-6.2 to 7.1), respectively. Intraclass correlation coefficient for repeatability ranged from 0.81 to 0.96. Conclusions: Correlation between Corvis ST and ORA was found to be good and not so with GAT. However, agreement between the devices was statistically insignificant, and no influence of sequence was observed.
引用
收藏
页码:815 / 820
页数:6
相关论文
共 25 条
[1]   Corneal biomechanical properties in eyes with no previous surgery, with previous penetrating keratoplasty and with deep anterior lamellar keratoplasty [J].
Acar, Banu Torun ;
Akdemir, Mehmet Orcun ;
Acar, Suphi .
JAPANESE JOURNAL OF OPHTHALMOLOGY, 2013, 57 (01) :85-89
[2]   Spectral Analysis of Intraocular Pressure Pulse Wave in Open Angle Glaucomas and Healthy Eyes [J].
Bozic, Marija ;
Dukic, Miroslav L. ;
Stojkovic, Milenko .
CURRENT EYE RESEARCH, 2012, 37 (11) :1019-1024
[3]   Scheimpflug-Based Tomography and Biomechanical Assessment in Pressure-Induced Stromal Keratopathy [J].
Faria-Correia, Fernando ;
Ramos, Isaac ;
Valbon, Bruno ;
Luz, Allan ;
Roberts, Cynthia J. ;
Ambrosio, Renato, Jr. .
JOURNAL OF REFRACTIVE SURGERY, 2013, 29 (05) :356-358
[4]   Clinical and Corneal Biomechanical Changes After Collagen Cross-Linking With Riboflavin and UV Irradiation in Patients With Progressive Keratoconus: Results After 2 Years of Follow-up [J].
Goldich, Yakov ;
Marcovich, Arie L. ;
Barkana, Yaniv ;
Mandel, Yossi ;
Hirsh, Ami ;
Morad, Yair ;
Avni, Isaac ;
Zadok, David .
CORNEA, 2012, 31 (06) :609-614
[5]   Is the PASCAL®-Tonometer suitable for measuring intraocular pressure in clinical routine?: Long- and short-term reproducibility of dynamic contour tonometry [J].
Herdener, S. ;
Hafizovic, D. ;
Pache, M. ;
Lautebach, S. ;
Funk, J. .
EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2008, 18 (01) :39-43
[6]   Corneal Deformation Measurement Using Scheimpflug Noncontact Tonometry [J].
Hon, Ying ;
Lam, Andrew K. C. .
OPTOMETRY AND VISION SCIENCE, 2013, 90 (01) :E1-E8
[7]   A New Tonometer-The Corvis ST Tonometer: Clinical Comparison with Noncontact and Goldmann Applanation Tonometers [J].
Hong, Jiaxu ;
Xu, Jianjiang ;
Wei, Anji ;
Deng, Sophie X. ;
Cui, Xinhan ;
Yu, Xiaobo ;
Sun, Xinghuai .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2013, 54 (01) :659-665
[8]   Relationship Between Corneal Biomechanical Properties, Central Corneal Thickness, and Intraocular Pressure Across the Spectrum of Glaucoma [J].
Kaushik, Sushmita ;
Pandav, Surinder Singh ;
Banger, Anupam ;
Aggarwal, Kanika ;
Gupta, Amod .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2012, 153 (05) :840-849
[9]  
Kirwan C, 2007, AM J OPHTHALMOL, V144, P642
[10]   Influence of corneal biomechanical properties on intraocular pressure measurement - Quantitative analysis [J].
Liu, J ;
Roberts, CJ .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (01) :146-155