Properties of the Stimulus Router System, a Novel Neural Prosthesis

被引:10
作者
Gan, Liu Shi [1 ,2 ]
Prochazka, Arthur [2 ]
机构
[1] Univ Alberta, Dept Biomed Engn, Edmonton, AB T6G 2E1, Canada
[2] Univ Alberta, Ctr Neurosci, Katz Grp 5005, Rexall Ctr Pharm & Hlth Res, Edmonton, AB T6G 2E1, Canada
基金
加拿大健康研究院;
关键词
Functional electrical stimulation; neural prostheses; passive implant; stimulus router system; STIMULATION; IMPLANT;
D O I
10.1109/TBME.2009.2031427
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Various types of neural prostheses (NPs) have been developed to restore motor function after neural injury. Surface NPs are noninvasive and inexpensive, but are often poorly selective, activating nontargeted muscles and cutaneous sensory nerves that can cause discomfort or pain. Implantable NPs are highly selective, but invasive and costly. The stimulus router system (SRS) is a novel NP consisting of fully implanted leads that "capture" and route some of the current flowing between a pair of surface electrodes to the vicinity of a target nerve. An SRS lead consists of a "pickup" terminal that is implanted subcutaneously under one of the surface electrodes and a "delivery" terminal that is secured on or near the target nerve. We have published a preliminary report on the basic properties of the SRS [L. S. Gan et al., "A new means of transcutaneous coupling for neural prostheses," IEEE Trans. Biomed. Eng., vol. 54, no. 3, pp. 509-517, Mar. 2007]. Here, we further characterize the SRS and identify aspects that maximize its performance as a motor NP. The surface current needed to activate nerves with an SRS, was found to depend on the proximity of the delivery terminal(s) to the nerve, electrode configurations, contact areas of the surface electrodes and implanted terminals, and the distance between the surface anode and the delivery terminal.
引用
收藏
页码:450 / 459
页数:10
相关论文
共 19 条
[1]   Deep brain stimulation [J].
Breit, S ;
Schulz, JB ;
Benabid, AL .
CELL AND TISSUE RESEARCH, 2004, 318 (01) :275-288
[2]   SACRAL ANTERIOR ROOT STIMULATORS FOR BLADDER CONTROL IN PARAPLEGIA - THE 1ST 50 CASES [J].
BRINDLEY, GS ;
POLKEY, CE ;
RUSHTON, DN ;
CARDOZO, L .
JOURNAL OF NEUROLOGY NEUROSURGERY AND PSYCHIATRY, 1986, 49 (10) :1104-1114
[3]   MULTIPLE ELECTRODE COCHLEAR IMPLANT [J].
CLARK, GM ;
TONG, YC ;
BLACK, R ;
FORSTER, IC ;
PATRICK, JF ;
DEWHURST, DJ .
JOURNAL OF LARYNGOLOGY AND OTOLOGY, 1977, 91 (11) :935-945
[4]  
GAN L, 2009, SOC NEUR CHIC IL
[5]   A new means of transcutaneous coupling for neural prostheses [J].
Gan, Liu Shi ;
Prochazka, Arthur ;
Bornes, Troy D. ;
Denington, Allen A. ;
Chan, K. Ming .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2007, 54 (03) :509-517
[6]  
GAN LS, 2009, THESIS U ALBERTA EDM
[7]   LONG-TERM VENTILATORY SUPPORT BY DIAPHRAGM PACING IN QUADRIPLEGIA [J].
GLENN, WWL ;
HOLCOMB, WG ;
SHAW, RK ;
HOGAN, JF ;
HOLSCHUH, KR .
ANNALS OF SURGERY, 1976, 183 (05) :566-577
[8]   The effect of stimulus pulse duration on selectivity of neural stimulation [J].
Grill, WM ;
Mortimer, JT .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1996, 43 (02) :161-166
[9]   ELECTRICAL-PROPERTIES OF IMPLANT ENCAPSULATION TISSUE [J].
GRILL, WM ;
MORTIMER, JT .
ANNALS OF BIOMEDICAL ENGINEERING, 1994, 22 (01) :23-33
[10]  
GRILL WM, 1995, P IEEE 17 ANN C ENG, P1089