摘要
Modulated electron radiation therapy (MERT) is able to deliver conformal dose to shallow tumors while significantly reducing dose to distal structures and surrounding tissues. An electron specific multileaf collimator (eMLC) has been proposed and constructed as an effective means of delivering electron beams for MERT. The aim of this work is to apply the Monte Carlo method to investigate the fluence profiles collimated by the eMLC in order to achieve accurate beam delivery for MERT. In this work, the EGS4/BEAM code was used to simulate the eMLC collimated electron beams of 6-20 MeV generated from a Varian Clinac 2100C linear accelerator. An attempt was made to describe the fluence profiles with an analytic Sigmoid function. The function parameters were determined by the fittings of the Monte Carlo simulated fluence profiles. How the function parameters depend on the eMLC aperture size, the off-axis location, and the electron beam energy has been investigated. It has been found that the eMLC collimated fluence profiles are dependent on beam energy, while almost independent of leaf location or dimension of MLC aperture. There is little difference in the fluence profiles collimated by the leaf side and the leaf end for the straight-edged leaves. It is possible that these energy-dependent Sigmoid functions can serve as operators to account for the energy dependence of the eMLC collimated fluence profiles. These operators can be incorporated into the inverse planning algorithm to derive desired dose distributions using a set of electron beams of variable energy and field size suitable for delivery by the eMLC. (C) 2002 American Association of Physicists in Medicine.
引用
收藏
页码:2472 / 2483
页数:12
相关论文
共 9 条
  • [1] Monte Carlo Investigation of the Mobetron to Perform Modulated Electron Beam Therapy
    Emam, I.
    Eldib, A.
    Hosini, M.
    AlSaeed, E.
    Ma, C.
    MEDICAL PHYSICS, 2014, 41 (06) : 110 - 110
  • [2] Monte Carlo simulation of small electron fields collimated by the integrated photon MLC
    Mihaljevic, Josip
    Soukup, Martin
    Dohm, Oliver
    Alber, Markus
    PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (03) : 829 - 843
  • [3] Auto-commissioning of a Monte Carlo electron beam model with application to photon MLC shaped electron fields
    Fix, M. K.
    Frei, D.
    Mueller, S.
    Guyer, G.
    Loebner, H. A.
    Volken, W.
    Manser, P.
    PHYSICS IN MEDICINE AND BIOLOGY, 2023, 68 (04)
  • [4] A MONTE-CARLO INVESTIGATION OF ELECTRON-BEAM APPLICATOR SCATTER
    EBERT, MA
    HOBAN, PW
    MEDICAL PHYSICS, 1995, 22 (09) : 1431 - 1435
  • [5] Electron modulated arc therapy (EMAT) using photon MLC for postmastectomy chest wall treatment I: Monte Carlo-based dosimetric characterizations
    Ma, Chaoqiong
    Parsons, David
    Chen, Mingli
    Jiang, Steve
    Hou, Qing
    Gu, Xuejun
    Lu, Weiguo
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2019, 67 : 1 - 8
  • [6] Considerations and limitations of fast Monte Carlo electron transport in radiation therapy based on precalculated data
    Jabbari, Keyvan
    Keall, Paul
    Seuntjens, Jan
    MEDICAL PHYSICS, 2009, 36 (02) : 530 - 540
  • [7] The use of intensity-modulated radiation therapy photon beams for improving the dose uniformity of electron beams shaped with MLC
    Mosalaei, Homeira
    Karnas, Scott
    Shah, Sheel
    Van Doodewaard, Sharon
    Foster, Tim
    Chen, Jeff
    MEDICAL DOSIMETRY, 2012, 37 (01) : 76 - 83
  • [8] Electron beam quality kQ,Q0 factors for various ionization chambers: a Monte Carlo investigation with PENELOPE
    Erazo, F.
    Brualla, L.
    Lallena, A. M.
    PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (21) : 6673 - 6691
  • [9] Characterization of Radiation Dose Associated with Dark Currents During Beam Hold for Respiratory-Gated Electron Therapy
    Hessler, J.
    Gupta, N.
    Rong, Y.
    Weldon, M.
    MEDICAL PHYSICS, 2014, 41 (06) : 237 - 237