Two dimensional non-scanning transform-free spatial-domain optical coherence tomography

被引:1
作者
Lin, Yu-Kai [1 ]
Chang, Chun-Wei [1 ]
Hsu, I-Jen [1 ]
机构
[1] Chung Yuan Christian Univ, Dept Phys, Taoyuan 32023, Taiwan
关键词
SINGLE-SHOT; INCOHERENT ILLUMINATION; RESOLUTION IMPROVEMENT; SWEPT SOURCE; SENSITIVITY; SUPPRESSION; ALGORITHM; SYSTEM;
D O I
10.1364/BOE.10.005789
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A novel optical coherence tomography system that can perform scanless two dimensional imaging without Fourier transform is proposed and demonstrated. In the system, a convex cylindrical mirror generates an extended spatial distribution of optical delay in the reference arm and a cylindrical lens is used to form a focused line beam in the sample arm. A charge-coupled device camera captures the two dimensional tomographic image of a sample in a snap-shot manner. Due to its simple configuration and operation, the system is suitable for developing a compact device for tomographic imaging and measurement. (C) 2019 Optical Society of America under the terms of the OSA Open Access Publishing Agreement
引用
收藏
页码:5789 / 5800
页数:12
相关论文
共 32 条
[1]   Methods to assess sensitivity of optical coherence tomography systems [J].
Agrawal, Anant ;
Pfefer, T. Joshua ;
Woolliams, Peter D. ;
Tomlins, Peter H. ;
Nehmetallah, George .
BIOMEDICAL OPTICS EXPRESS, 2017, 8 (02) :902-917
[2]   A systematic approach to the design of distributed wearable systems [J].
Anliker, U ;
Beutel, J ;
Dyer, M ;
Enzler, R ;
Lukowicz, P ;
Thiele, L ;
Tröster, G .
IEEE TRANSACTIONS ON COMPUTERS, 2004, 53 (08) :1017-1033
[3]   Full-field optical coherence microscopy [J].
Beaurepaire, E ;
Boccara, AC ;
Lebec, M ;
Blanchot, L ;
Saint-Jalmes, H .
OPTICS LETTERS, 1998, 23 (04) :244-246
[4]   Single shot, time-resolved measurement of the coherence properties of OCT swept source lasers [J].
Butler, T. ;
Slepneva, S. ;
O'Shaughnessy, B. ;
Kelleher, B. ;
Goulding, D. ;
Hegarty, S. P. ;
Lyu, H. -C. ;
Karnowski, K. ;
Wojtkowski, M. ;
Huyet, G. .
OPTICS LETTERS, 2015, 40 (10) :2277-2280
[5]   Leaf anatomy of Ficus subsection Urostigma (Moraceae) [J].
Chantarasuwan, Bhanumas ;
Baas, Pieter ;
Van Heuven, Bertie-Joan ;
Baider, Claudia ;
Van Welzen, Peter C. .
BOTANICAL JOURNAL OF THE LINNEAN SOCIETY, 2014, 175 (02) :259-281
[6]   Optical coherence tomography using a frequency-tunable optical source [J].
Chinn, SR ;
Swanson, EA ;
Fujimoto, JG .
OPTICS LETTERS, 1997, 22 (05) :340-342
[7]   Quantitative assessment of touch-screen panel by nondestructive inspection with three-dimensional real-time display optical coherence tomography [J].
Cho, Nam Hyun ;
Park, Kibeom ;
Kim, Jae-Young ;
Jung, Yeongri ;
Kim, Jeehyun .
OPTICS AND LASERS IN ENGINEERING, 2015, 68 :50-57
[8]   Sensitivity advantage of swept source and Fourier domain optical coherence tomography [J].
Choma, MA ;
Sarunic, MV ;
Yang, CH ;
Izatt, JA .
OPTICS EXPRESS, 2003, 11 (18) :2183-2189
[9]   Crosstalk rejection in parallel optical coherence tomography using spatially incoherent illumination with partially coherent sources [J].
Dhalla, Al-Hafeez ;
Migacz, Justin V. ;
Izatt, Joseph A. .
OPTICS LETTERS, 2010, 35 (13) :2305-2307
[10]   Optical coherence tomography today: speed, contrast, and multimodality [J].
Drexler, Wolfgang ;
Liu, Mengyang ;
Kumar, Abhishek ;
Kamali, Tschackad ;
Unterhuber, Angelika ;
Leitgeb, Rainer A. .
JOURNAL OF BIOMEDICAL OPTICS, 2014, 19 (07)