The Amplatzer Duct Occluder II: A New Device for Percutaneous Ductus Arteriosus Closure

被引:24
作者
Saliba, Zakhia [1 ]
El-Rassi, Issam [1 ]
Abi-Warde, Marie-Therese [1 ]
Chehab, Ghassan [1 ]
Daou, Linda [1 ]
Khater, Daniele [1 ]
Gerbaka, Bernard [1 ]
Jebara, Victor [1 ]
机构
[1] St Joseph Univ, Hotel Dieu France Hosp, Beirut, Lebanon
关键词
TRANSCATHETER CLOSURE; PATENT; OCCLUSION; COILS;
D O I
10.1111/j.1540-8183.2009.00504.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Methods: Between June 2008 and March 2009, 20 consecutive patients were enrolled. In cases where different devices were applicable, we favored the use of the ADO II to maximize our experience with this device and prove its superiority. No coils were required in these 20 patients. Results: There were 15 females and 5 males (median age 2 years). ADO II group (n = 16): Immediate complete closure in 75% of the patients, rising to 93.7% at 24 hours. A residual shunt persisted at 3 months in one child. Aortic narrowing from device protrusion was noted in two type E ducts, without any significant gradient, however. ADO I group (n = 4): In two adolescents and in one adult patient, the duct was successfully closed. In a 2-year-old patient with a 6.6 mm type B duct, the ADO I totally obstructed the aortic flow and was retrieved before releasing. The child was sent for surgery. Conclusion: Even though we did not compare the ADO II to other devices, we feel that it has the capacity to substitute most of the coils, and some of the original ADO I indications. Arterial access was sufficient in most patients, but venous delivery is advised in small infants with large or long ducts, to avoid aortic protrusion and residual shunts. (J Interven Cardiol 2009;22:496-502).
引用
收藏
页码:496 / 502
页数:7
相关论文
共 13 条
  • [1] *AGA MED CORP, 2008, PROD SPEC COURT AGA
  • [2] The Amplatzer duct occluder: Experience in 209 patients
    Bilkis, AA
    Alwi, M
    Hasri, S
    Haifa, AL
    Geetha, K
    Rehman, MA
    Hasanah, I
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2001, 37 (01) : 258 - 261
  • [3] FORSEY J, 2009, CATHETER CARDIOVASC
  • [4] Comparison of Two Transcatheter Device Strategies for Occlusion of the Patent Ductus Arteriosus
    Gudausky, Todd M.
    Hirsch, Russel
    Khoury, Philip R.
    Beekman, Robert H.
    [J]. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2008, 72 (05) : 675 - 680
  • [5] Safety and efficacy of interventional occlusion of patent ductus arteriosus with detachable coils:: a multicentre experience
    Hofbeck, M
    Bartolomaeus, G
    Buheitel, G
    Esser, R
    Grävinghoff, L
    Hoffmann, W
    Kienast, W
    Michel-Behnke, I
    Scharabrine, EG
    Schranz, D
    Schmaltz, AA
    Shakhov, BE
    Singer, H
    Lindinger, A
    [J]. EUROPEAN JOURNAL OF PEDIATRICS, 2000, 159 (05) : 331 - 337
  • [6] ANGIOGRAPHIC CLASSIFICATION OF THE ISOLATED, PERSISTENTLY PATENT DUCTUS-ARTERIOSUS AND IMPLICATIONS FOR PERCUTANEOUS CATHETER OCCLUSION
    KRICHENKO, A
    BENSON, LN
    BURROWS, P
    MOES, CAF
    MCLAUGHLIN, P
    FREEDOM, RM
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1989, 63 (12) : 877 - 880
  • [7] Interventional treatment of patent ductus arteriosus in 2004
    Moore, JW
    Levi, DS
    Moore, SD
    Schneider, DJ
    Berdjis, F
    [J]. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2005, 64 (01) : 91 - 101
  • [8] Transcatheter Closure of Long Tubular Patent Arterial Ducts: The Amplatzer Duct Occluder II - A New and Valuable Tool
    Morgan, Gareth
    Tornetzki, Andrew J.
    Martin, Robin P.
    [J]. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2009, 73 (04) : 576 - 580
  • [9] Percutaneous closure of patent ductus arteriosus: complementary use of detachable Cook patent ductus arteriosus coils and Amplatzer duct occluders
    Podnar, T
    Gavora, P
    Masura, J
    [J]. EUROPEAN JOURNAL OF PEDIATRICS, 2000, 159 (04) : 293 - 296
  • [10] Saliba Z, 2000, ARCH MAL COEUR VAISS, V93, P533