Biomechanical Comparison of Polyaxial-Type Locking Plates and a Fixed-Angle Locking Plate for Internal Fixation of Distal Femur Fractures

被引:68
|
作者
Otto, Randall J. [1 ]
Moed, Berton R. [1 ]
Bledsoe, J. Gary [2 ]
机构
[1] St Louis Univ, Sch Med, Dept Orthopaed Surg, St Louis, MO 63110 USA
[2] St Louis Univ, Pk Coll Engn Aviat & Technol, Dept Biomed Engn, St Louis, MO 63110 USA
关键词
polyaxial; distal femur; locking plate; fracture fixation; biomechanics; INVASIVE STABILIZATION SYSTEM; BLADE PLATE; SCREW;
D O I
10.1097/BOT.0b013e3181a567c8
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objectives: To test the stability to axial loading of 2 new polyaxial locking screw-plate designs and analyze different angles of screw insertion. The noncontact bridging (NCB) polyaxial locking plate (Zimmer) and the POLYAX plate (DePuy) were compared with a fixed-angle less invasive stabilization system (LISS; Synthes). Methods: Twenty-five synthetic femurs were divided into 5 groups and assigned fixation with the LISS plate (group I), POLYAX plate (groups IIA and IIB), or NCB plate (groups IIIA and IIIB). The polyaxial constructs were divided into parallel and crossed distal condylar screw configurations. Each construct was tested under axial loading and stressed to failure at a displacement rate of 5 mm/min with a preload of 100 N. Outcome measurements included stiffness, load to failure, peak force, and mode of failure, Results: All LISS and POLYAX constructs failed by plastic deformation of the plate, whereas 9 of 10 NCB constructs failed by an intraarticular lateral condyle fracture. No failures occurred at the screwplate interface in either polyaxial constructs. Load to failure of the LISS was 33% greater than the parallel POLYAX (P < 0.01) and 24% greater than the crossed POLYAX (P < 0.01). Load to failure of NCB (parallel and crossed) were 24% greater than the parallel POLYAX (P < 0.01 and P < 0.01, respectively) and 1501,, greater than the crossed POLYAX (P < 0.01 and P = 0.02, respectively). The POLYAX also had significantly lower stiffness and peak force compared with the LISS and NCB. There was no difference between the LISS and NCB with regard to stiffness, load to failure, and peak force. Parallel and crossed polyaxial constructs showed no difference in stiffness or failure loads. Conclusions: There were no failures of either polyaxial screw-plate interface despite large forces and screw angle did not affect the overall strength of these constructs, supporting the biomechanical soundness of both polyaxial device designs under axial loading. However, the POLYAX supported smaller loads compared with the LISS and NCB while under axial loading. In addition, the mode of failure of the NCB plate, creating an intra-articular fracture propagating from the distal posterior screw hole, may be of some concern. Additional testing is needed to determine the clinical importance of the demonstrated differences among these plate designs.
引用
收藏
页码:645 / 652
页数:8
相关论文
共 50 条
  • [31] Comparison Between Locking and Non-Locking Plates for Fixation of Metacarpal Fractures in an Animal Model
    Ochman, Sabine
    Doht, Stephanie
    Paletta, Juergen
    Langer, Martin
    Raschke, Michael J.
    Meffert, Rainer H.
    JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2010, 35A (04): : 597 - 603
  • [32] Clinical outcome of locking plate fixation in distal tibia fractures
    Mohebi, Ahmad Shakib
    Nadeem, Rana Dilawaiz
    Sah, Ranjit Kumar
    Mohy-Ud-Din, Iftikhar
    Akhtar, Muhammad
    JOURNAL OF THE PAKISTAN MEDICAL ASSOCIATION, 2023, 73 (05) : 1083 - 1086
  • [33] Volar locking plate fixation of unstable distal radius fractures
    Kilic, Ayhan
    Kabukcuoglu, Yavuz
    Ozkaya, Ufuk
    Gul, Murat
    Sokucu, Sami
    Ozdogan, Umit
    ACTA ORTHOPAEDICA ET TRAUMATOLOGICA TURCICA, 2009, 43 (04) : 303 - 308
  • [34] Early Mechanical Failures of the Synthes Variable Angle Locking Distal Femur Plate
    Tank, Jason C.
    Schneider, Prism S.
    Davis, Elizabeth
    Galpin, Matthew
    Prasarn, Mark L.
    Choo, Andrew M.
    Munz, John W.
    Achor, Timothy S.
    Kellam, James F.
    Gary, Joshua L.
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2016, 30 (01) : E7 - E11
  • [35] Early Clinical Outcomes of Polyaxial Locking Cap Plate Fixation for Fixation of Displaced Olecranon Fractures
    Gutman, Michael Jason
    Patel, Manan Sunil
    Connelly, James William
    Paxton, Edward Scott
    Namdari, Surena
    Horneff, John Gabriel
    ARCHIVES OF BONE AND JOINT SURGERY-ABJS, 2025, 13 (01): : 23 - 29
  • [36] Distal radius fractures: Minimally invasive plate osteosynthesis with dorsal bicolumnar locking plates fixation
    Alvin Chao-Yu Chen
    Ying-Chao Chou
    Chun-Ying Cheng
    Indian Journal of Orthopaedics, 2017, 51 : 93 - 98
  • [37] Locking Plate Fixation of Periprosthetic Femur Fractures with and without Cerclage Wires
    Ebraheim, Nabil A.
    Sochacki, Kyle R.
    Liu, Xiaochen
    Hirschfeld, Adam G.
    Liu, Jiayong
    ORTHOPAEDIC SURGERY, 2013, 5 (03) : 183 - 187
  • [38] Biomechanical Properties of Fixed-Angle Volar Distal Radius Plates Under Dynamic Loading
    Dahl, William J.
    Nassab, Paul F.
    Burgess, Kraig M.
    Postak, Paul D.
    Evans, Peter J.
    Seitz, William H.
    Greenwald, A. Seth
    Lawton, Jeffrey N.
    JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2012, 37A (07): : 1381 - 1387
  • [39] Retrograde locking nail fixation of distal femoal fractures with the distal femur nail (DFN)
    Grass, R
    Biewener, A
    Rammelt, S
    Zwipp, H
    UNFALLCHIRURG, 2002, 105 (04): : 298 - 314
  • [40] Number and Locations of Screw Fixation for Volar Fixed-Angle Plating of Distal Radius Fractures: Biomechanical Study
    Mehling, Isabella
    Mueller, Lars P.
    Delinsky, Katharina
    Mehler, Dorothea
    Burkhart, Klaus J.
    Rommens, Pol M.
    JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2010, 35A (06): : 885 - 891