The iCare rebound tonometer: comparisons with Goldmann tonometry, and influence of central corneal thickness

被引:64
作者
Poostchi, Ali [1 ]
Mitchell, Robert [1 ]
Nicholas, Simon [1 ]
Purdie, Gordon [2 ]
Wells, Anthony [1 ]
机构
[1] Wellington Hosp, Dept Ophthalmol, Wellington, New Zealand
[2] Wellington Sch Med, Dept Publ Hlth, Wellington, New Zealand
关键词
Goldmann tonometry; iCare rebound tonometer; intraocular pressure; MEASURING INTRAOCULAR-PRESSURE; APPLANATION TONOMETRY; TONO-PEN; CLINICAL-EVALUATION; PORTABLE TONOMETERS; EYES; AGREEMENT; XL; GLAUCOMA;
D O I
10.1111/j.1442-9071.2009.02109.x
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
P>Purpose: To assess agreement between the iCare rebound tonometer and Goldmann tonometry and to assess the influence of central corneal thickness and the value of scleral rebound tonometer readings. Methods: Prospective single-centre cross-sectional study comparing iCare rebound tonometer (RT) intraocular pressure (IOP) readings taken from corneal and scleral locations to Goldmann applanation tonometer (GAT) readings in 100 subjects attending ophthalmology clinics. Results: There was a significant difference between RT and GAT, with RT tending to overestimate IOP. The mean difference between RT and GAT measurements was 3.36 mmHg. The mean difference between the log of RT and the log of GAT measurements was 0.2356, a ratio of 1.27 (P < 0.0001). A formula derived from a linear regression analysis suggested that a 10% increase in CCT increased the RT IOP reading by 9.9%. Scleral RT readings showed no relationship to GAT readings. Conclusions: The rebound tonometer cannot replace the Goldmann tonometer in the office setting given the wide limits of agreement between the two devices. Corneal rebound tonometer readings are influenced by CCT whereas scleral rebound tonometer readings are of no value.
引用
收藏
页码:687 / 691
页数:5
相关论文
共 37 条
[11]   Correlation of intraocular pressure and central corneal thickness in normal myopic eyes and after laser in situ keratomileusis [J].
Emara, B ;
Probst, LE ;
Tingey, DP ;
Kennedy, DW ;
Willms, LJ ;
Machat, J .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1998, 24 (10) :1320-1325
[12]   Comparison of the ICare® rebound tonometer with the Goldmann tonometer in a normal population [J].
Fernandes, P ;
Díaz-Rey, JA ;
Queirós, A ;
Gonzalez-Meijome, JM ;
Jorge, J .
OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2005, 25 (05) :436-440
[13]   Accuracy of the new ICare rebound tonometer vs. other portable tonometers in healthy eyes [J].
García-Resúa, C ;
González-Meijome, JM ;
Gilino, J ;
Yebra-Pimentel, E .
OPTOMETRY AND VISION SCIENCE, 2006, 83 (02) :102-107
[14]   Central corneal thickness in normal, glaucomatous, and ocular hypertensive eyes [J].
Herndon, LW ;
Choudhri, SA ;
Cox, T ;
Damji, KF ;
Shields, MB ;
Allingham, RR .
ARCHIVES OF OPHTHALMOLOGY, 1997, 115 (09) :1137-1141
[15]   Comparison of the Tono-Pen and Goldmann tonometer for measuring intraocular pressure in patients with glaucoma [J].
Horowitz, GS ;
Byles, J ;
Lee, JL ;
D'Este, C .
CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2004, 32 (06) :584-589
[16]   New TonoPen XL: comparison with the Goldmann tonometer [J].
Iester, M ;
Mermoud, A ;
Achache, F ;
Roy, S .
EYE, 2001, 15 (1) :52-58
[17]   Comparison of rebound tonometry with Goldmann applanation tonometry and correlation with central corneal thickness [J].
Iliev, M. E. ;
Goldblum, D. ;
Katsoulis, K. ;
Amstutz, C. ;
Frueh, B. .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2006, 90 (07) :833-835
[18]   Standardizing the measurement of intraocular pressure for clinical research - Guidelines from the eye care technology forum [J].
Kass, MA .
OPHTHALMOLOGY, 1996, 103 (01) :183-185
[19]   Effect of central corneal thickness, corneal curvature, and axial length on applanation tonometry [J].
Kohlhaas, M ;
Boehm, AG ;
Spoerl, E ;
Pürsten, A ;
Grein, HJ ;
Pillunat, LE .
ARCHIVES OF OPHTHALMOLOGY, 2006, 124 (04) :471-476
[20]   Comparison of scleral and corneal Tono-Pen readings [J].
Kolin, T ;
Wedemeyer, LL ;
Kolin, E ;
Braun, Y .
JOURNAL OF AAPOS, 2003, 7 (04) :291-292