The iCare rebound tonometer: comparisons with Goldmann tonometry, and influence of central corneal thickness

被引:64
作者
Poostchi, Ali [1 ]
Mitchell, Robert [1 ]
Nicholas, Simon [1 ]
Purdie, Gordon [2 ]
Wells, Anthony [1 ]
机构
[1] Wellington Hosp, Dept Ophthalmol, Wellington, New Zealand
[2] Wellington Sch Med, Dept Publ Hlth, Wellington, New Zealand
关键词
Goldmann tonometry; iCare rebound tonometer; intraocular pressure; MEASURING INTRAOCULAR-PRESSURE; APPLANATION TONOMETRY; TONO-PEN; CLINICAL-EVALUATION; PORTABLE TONOMETERS; EYES; AGREEMENT; XL; GLAUCOMA;
D O I
10.1111/j.1442-9071.2009.02109.x
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
P>Purpose: To assess agreement between the iCare rebound tonometer and Goldmann tonometry and to assess the influence of central corneal thickness and the value of scleral rebound tonometer readings. Methods: Prospective single-centre cross-sectional study comparing iCare rebound tonometer (RT) intraocular pressure (IOP) readings taken from corneal and scleral locations to Goldmann applanation tonometer (GAT) readings in 100 subjects attending ophthalmology clinics. Results: There was a significant difference between RT and GAT, with RT tending to overestimate IOP. The mean difference between RT and GAT measurements was 3.36 mmHg. The mean difference between the log of RT and the log of GAT measurements was 0.2356, a ratio of 1.27 (P < 0.0001). A formula derived from a linear regression analysis suggested that a 10% increase in CCT increased the RT IOP reading by 9.9%. Scleral RT readings showed no relationship to GAT readings. Conclusions: The rebound tonometer cannot replace the Goldmann tonometer in the office setting given the wide limits of agreement between the two devices. Corneal rebound tonometer readings are influenced by CCT whereas scleral rebound tonometer readings are of no value.
引用
收藏
页码:687 / 691
页数:5
相关论文
共 37 条
[1]   Comparison of the ICare® rebound tonometer with the Goldmann applanation tonometer by experienced and inexperienced tonometrists [J].
Abraham, L. M. ;
Epasinghe, N. C. R. ;
Selva, D. ;
Casson, R. .
EYE, 2008, 22 (04) :503-506
[2]   Clinical comparison of the measurement of the IOP with the ocular blood flow tonometer, the Tonopen XL and the Goldmann applanation tonometer [J].
Bafa, M ;
Lambrinakis, I ;
Dayan, M ;
Birch, M .
ACTA OPHTHALMOLOGICA SCANDINAVICA, 2001, 79 (01) :15-18
[3]   Agreement between methods of measurement with multiple observations per individual [J].
Bland, J. Martin ;
Altman, Douglas G. .
JOURNAL OF BIOPHARMACEUTICAL STATISTICS, 2007, 17 (04) :571-582
[4]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[5]   LIMBAL PNEUMATONOMETRY [J].
BREITFELLER, JM ;
KROHN, DL .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1980, 89 (03) :344-352
[6]   The effect of corneal thickness on intraocular pressure measurement in patients with corneal pathology [J].
Browning, AC ;
Bhan, A ;
Rotchford, AP ;
Shah, S ;
Dua, HS .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2004, 88 (11) :1395-1399
[7]   Comparison of ICare tonometer with Goldmann applanation tonometer in glaucoma patients [J].
Brusini, P ;
Salvetat, ML ;
Zeppieri, M ;
Tosoni, C ;
Parisi, L .
JOURNAL OF GLAUCOMA, 2006, 15 (03) :213-217
[8]   Clinical evaluation of rebound tonometer [J].
Davies, LN ;
Bartlett, H ;
Mallen, EAH ;
Wolffsohn, JS .
ACTA OPHTHALMOLOGICA SCANDINAVICA, 2006, 84 (02) :206-209
[9]   Clinical evaluation of the dynamic rebound tonometer Icare® [J].
Detry-Morel, M. ;
Jamart, J. ;
Detry, M. B. ;
Pourjavan, S. ;
Charlier, L. ;
Dethinne, B. ;
Huge, L. ;
Ledoux, A. .
JOURNAL FRANCAIS D OPHTALMOLOGIE, 2006, 29 (10) :1119-1127
[10]  
EHLERS N, 1975, ACTA OPHTHALMOL, V53, P34