The iCare rebound tonometer: comparisons with Goldmann tonometry, and influence of central corneal thickness

被引:62
|
作者
Poostchi, Ali [1 ]
Mitchell, Robert [1 ]
Nicholas, Simon [1 ]
Purdie, Gordon [2 ]
Wells, Anthony [1 ]
机构
[1] Wellington Hosp, Dept Ophthalmol, Wellington, New Zealand
[2] Wellington Sch Med, Dept Publ Hlth, Wellington, New Zealand
关键词
Goldmann tonometry; iCare rebound tonometer; intraocular pressure; MEASURING INTRAOCULAR-PRESSURE; APPLANATION TONOMETRY; TONO-PEN; CLINICAL-EVALUATION; PORTABLE TONOMETERS; EYES; AGREEMENT; XL; GLAUCOMA;
D O I
10.1111/j.1442-9071.2009.02109.x
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
P>Purpose: To assess agreement between the iCare rebound tonometer and Goldmann tonometry and to assess the influence of central corneal thickness and the value of scleral rebound tonometer readings. Methods: Prospective single-centre cross-sectional study comparing iCare rebound tonometer (RT) intraocular pressure (IOP) readings taken from corneal and scleral locations to Goldmann applanation tonometer (GAT) readings in 100 subjects attending ophthalmology clinics. Results: There was a significant difference between RT and GAT, with RT tending to overestimate IOP. The mean difference between RT and GAT measurements was 3.36 mmHg. The mean difference between the log of RT and the log of GAT measurements was 0.2356, a ratio of 1.27 (P < 0.0001). A formula derived from a linear regression analysis suggested that a 10% increase in CCT increased the RT IOP reading by 9.9%. Scleral RT readings showed no relationship to GAT readings. Conclusions: The rebound tonometer cannot replace the Goldmann tonometer in the office setting given the wide limits of agreement between the two devices. Corneal rebound tonometer readings are influenced by CCT whereas scleral rebound tonometer readings are of no value.
引用
收藏
页码:687 / 691
页数:5
相关论文
共 50 条
  • [1] Comparability of Icare Pro Rebound Tonometer with Goldmann Applanation and Noncontact Tonometer in a Wide Range of Intraocular Pressure and Central Corneal Thickness
    Tamcelik, Nevbahar
    Atalay, Eray
    Cicik, Erdogan
    Ozkok, Ahmet
    OPHTHALMIC RESEARCH, 2015, 54 (01) : 18 - 25
  • [2] Relationship of Central Corneal Thickness and Intraocular Pressure by iCare Rebound Tonometer
    Rao, Aparna
    Kumar, Mukesh
    Prakash, B.
    Varshney, Gopal
    JOURNAL OF GLAUCOMA, 2014, 23 (06) : 380 - 384
  • [3] Comparison of the iCare rebound tonometer and the Goldmann applanation tonometer
    Gao, Feng
    Liu, Hu
    Zhao, Qing
    Pan, Yingzhe
    EXPERIMENTAL AND THERAPEUTIC MEDICINE, 2017, 13 (05) : 1912 - 1916
  • [4] Precision and accuracy of the ICare tonometer - Peripheral and central IOP measurements by rebound tonometry
    Muttuvelu, Danson V.
    Baggesen, Kirsten
    Ehlers, Niels
    ACTA OPHTHALMOLOGICA, 2012, 90 (04) : 322 - 326
  • [5] Evaluation of a New Rebound Self-tonometer, Icare HOME: Comparison With Goldmann Applanation Tonometer
    Takagi, Daisuke
    Sawada, Akira
    Yamamoto, Tetsuya
    JOURNAL OF GLAUCOMA, 2017, 26 (07) : 613 - 618
  • [6] The influence of refractive errors on IOP measurement by rebound tonometry (ICare) and Goldmann applanation tonometry
    Avitabile, Teresio
    Longo, Antonio
    Rocca, Daniela
    Amato, Roberta
    Gagliano, Caterina
    Castaing, Marine
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2010, 248 (04) : 585 - 591
  • [7] Clinical Evaluation of the New Rebound Tonometers Icare PRO and Icare ONE Compared With the Goldmann Tonometer
    Moreno-Montanes, Javier
    Martinez-de-la-Casa, Jose M.
    Sabater, Alfonso L.
    Morales-Fernandez, Laura
    Saenz, Cristina
    Garcia-Feijoo, Julian
    JOURNAL OF GLAUCOMA, 2015, 24 (07) : 527 - 532
  • [8] Correlation Between Goldmann Applanation Tonometry and Rebound Tonometry in Relation to Central Corneal Thickness
    Brandtner, H.
    Hitzl, W.
    Strohmaier, C.
    Abri, S.
    Wintersteller, C.
    Hohensinn, M.
    Ruckhofer, J.
    Stoiber, J.
    Grabner, G.
    Reitsamer, H.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (13)
  • [9] The influence of corneal density and thickness on tonometry measurement with goldmann applanation, non-contact and iCare tonometry methods
    Ahmed Lubbad
    Irene Oluwatoba-Popoola
    Melanie Haar
    Carsten Framme
    Anna Bajor
    International Ophthalmology, 2022, 42 : 2167 - 2174
  • [10] The influence of corneal density and thickness on tonometry measurement with goldmann applanation, non-contact and iCare tonometry methods
    Lubbad, Ahmed
    Oluwatoba-Popoola, Irene
    Haar, Melanie
    Framme, Carsten
    Bajor, Anna
    INTERNATIONAL OPHTHALMOLOGY, 2022, 42 (07) : 2167 - 2174