X-ray micro-tomography system for small-animal imaging with zoom-in imaging capability

被引:36
作者
Chun, IK [1 ]
Cho, MH [1 ]
Lee, SC [1 ]
Cho, MH [1 ]
Lee, SY [1 ]
机构
[1] Kyung Hee Univ, Dept Biomed Engn, Seoul, South Korea
关键词
D O I
10.1088/0031-9155/49/17/005
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Since a micro-tomography system capable of mum-resolution imaging cannot be used for whole-body imaging of a small laboratory animal without sacrificing its spatial resolution, it is desirable for a micro-tomography system to have local imaging capability. In this paper, we introduce an x-ray micro-tomography system capable of high-resolution imaging of a local region inside a small animal. By combining two kinds of projection data, one from a full field-of-view (FOV) scan of the whole body and the other from a limited FOV scan of the region of interest (ROI), we have obtained zoomed-in images of the ROI without any contrast anomalies commonly appearing in conventional local tomography. For experimental verification of the zoom-in imaging capability, we have integrated a micro-tomography system using a micro-focus x-ray source, a 1248 x 1248 flat-panel x-ray detector, and a precision scan mechanism. The mismatches between the two projection data caused by misalignments of the scan mechanism have been estimated with a calibration phantom, and the mismatch effects have been compensated in the image reconstruction procedure. Zoom-in imaging results of bony tissues with a spatial resolution of 10 1p mm(-1) suggest that zoom-in micro-tomography can be greatly used for high-resolution imaging of a local region in small-animal studies.
引用
收藏
页码:3889 / 3902
页数:14
相关论文
共 22 条
[1]  
Cho Z.H, 1993, FDN MED IMAGING, P71
[2]   LOCAL TOMOGRAPHY [J].
FARIDANI, A ;
RITMAN, EL ;
SMITH, KT .
SIAM JOURNAL ON APPLIED MATHEMATICS, 1992, 52 (02) :459-484
[3]   Local tomography .2. [J].
Faridani, A ;
Finch, DV ;
Ritman, EL ;
Smith, KT .
SIAM JOURNAL ON APPLIED MATHEMATICS, 1997, 57 (04) :1095-1127
[4]   A HIGH-RESOLUTION XRII-BASED QUANTITATIVE VOLUME CT SCANNER [J].
HOLDSWORTH, DW ;
DRANGOVA, M ;
FENSTER, A .
MEDICAL PHYSICS, 1993, 20 (02) :449-462
[5]   Cone-beam computed tomography with a flat-panel imager: Initial performance characterization [J].
Jaffray, DA ;
Siewerdsen, JH .
MEDICAL PHYSICS, 2000, 27 (06) :1311-1323
[6]  
JOGENSEN SM, 1998, AM J PHYSIOL, V275, pH1103
[7]  
KAK AC, 1988, PRINCIPLES COMPUTED, P86
[8]  
Kapadia R D, 1998, Technol Health Care, V6, P361
[9]   The effect of an osteolytic tumor on the three-dimensional trabecular bone morphology in an animal model [J].
Kurth, AA ;
Müller, R .
SKELETAL RADIOLOGY, 2001, 30 (02) :94-98
[10]   A flat-panel detector based micro-CT system: performance evaluation for small-animal imaging [J].
Lee, SC ;
Kim, HK ;
Chun, IK ;
Cho, MH ;
Lee, SY ;
Cho, MH .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (24) :4173-4185