Effect of early activation of cochlear implant on electrode impedance in pediatric population

被引:15
|
作者
Alhabib, Salman F. [1 ]
Abdelsamad, Yassin [2 ]
Yousef, Medhat [1 ,3 ]
Alzhrani, Farid [1 ]
Hagr, Abdulrahman [1 ]
机构
[1] King Saud Univ, Coll Med, King Abdullah Ear Specialist Ctr KAESC, Riyadh, Saudi Arabia
[2] MED El GmbH, Res Dept, Riyadh, Saudi Arabia
[3] Menoufia Univ Hosp, Fac Med, Audiol Unit, Menoufia, Egypt
关键词
One-day activation; Electrode impedance; Fibrosis; Minimal invasive cochlear implantation technique; SAFETY;
D O I
10.1016/j.ijporl.2020.110543
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
Objectives: To assess the evolution of electrode impedance after the early fitting of audio processors (activation after one-day) and classical fitting (activation after one-month) over an up-to-one year after cochlear implant (CI) surgery. Methods: A retrospective cohort study on Fifty-two CI recipients divided into two groups. The study group included 24 recipients (40 ears) who underwent early fitting, whereas the control group contained 28 recipients (40 ears) who underwent classical fitting. The electrode impedance was recorded during the surgery, switch-on session and at one, three, six, and twelve-months after the surgery. Results: In the study group, electrode impedance values obtained intraoperatively and at switch-on and one, three, six, and twelve-months were 4.89, 3.69, 6.52, 6.24, 6.05, and 5.81 K Omega, respectively, and only the switch-on and one-month values were significantly different (p < 0.0001). In the control group, electrode impedance values obtained intraoperatively and at switch-on and one, three, six, and twelve-months were 4.71, 7.19, 6.40, 6.05, and 5.73 K Omega, respectively. Thus, the electrode impedance value at switch-on was 52.65% (p < 0.001) greater than it intraoperatively. For both groups, the electrode impedance value at twelve-months was significantly higher than the respective intraoperative values (study group: 18.6% higher, P = 0.04; control group: 21.65% higher, P = 0.0001). Conclusion: Electrode impedance was significantly lower in the study group compared to the control group until one month after the surgery. However, the electrode impedance at twelve-months after the CI was similar in both groups.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] Cochlear Implantation: Long-Term Effect of Early Activation on Electrode Impedance
    Alahmadi, Asma
    Abdelsamad, Yassin
    Yousef, Medhat
    Almuhawas, Fida
    Hafez, Ahmed
    Alzhrani, Farid
    Hagr, Abdulrahman
    JOURNAL OF CLINICAL MEDICINE, 2024, 13 (11)
  • [2] The Effect of Intracochlear and Intratympanic Dexamethasone on Cochlear Implant Impedance
    Ardic, Fazil Necdet
    Aydemir, Gokce
    Tumkaya, Funda
    Altinoz, Ece
    Senol, Hande
    TURKISH ARCHIVES OF OTORHINOLARYNGOLOGY, 2023, 61 (03) : 103 - 108
  • [3] Cochlear Implant Electrode Impedance as Potential Biomarker for Residual Hearing
    Wimmer, Wilhelm
    Sclabas, Luca
    Caversaccio, Marco
    Weder, Stefan
    FRONTIERS IN NEUROLOGY, 2022, 13
  • [4] Comparison of electrode impedance measures between a dexamethasone-eluting and standard Cochlear™ Contour Advance® electrode in adult cochlear implant recipients
    Briggs, Robert
    O'Leary, Stephen
    Birman, Catherine
    Plant, Kerrie
    English, Ruth
    Dawson, Pamela
    Risi, Frank
    Gavrilis, Jason
    Needham, Karina
    Cowan, Robert
    HEARING RESEARCH, 2020, 390
  • [5] Detection of Translocation of Cochlear Implant Electrode Arrays by Intracochlear Impedance Measurements
    Dong, Yu
    Briaire, Jeroen J.
    Siebrecht, Michael
    Stronks, H. Christiaan
    Frijns, Johan H. M.
    EAR AND HEARING, 2021, 42 (05): : 1397 - 1404
  • [6] Evolution of impedance field telemetry after one day of activation in cochlear implant recipients
    Hu, Hao-Chun
    Chen, Joshua Kuang-Chao
    Tsai, Chia-Mi
    Chen, Hsing-Yi
    Tung, Tao-Hsin
    Li, Lieber Po-Hung
    PLOS ONE, 2017, 12 (03):
  • [7] Effect of age, electrode array, and time on cochlear implant impedances
    Velandia, Sandra
    Martinez, Diane
    Goncalves, Stefania
    Pena, Stefanie
    Bas, Esperanza
    Ein, Liliana
    Prentiss, Sandra
    Telischi, Fred
    Angeli, Simon
    Dinh, Christine T.
    COCHLEAR IMPLANTS INTERNATIONAL, 2020, 21 (06) : 344 - 352
  • [8] Effect of Surgical Technique on Electrode Impedance After Cochlear Implantation
    Jia, Huan
    Venail, Frederic
    Piron, Jean-Pierre
    Batrel, Charlene
    Pelliccia, Pierfrancesco
    Artieres, Francoise
    Uziel, Alain
    Mondain, Michel
    ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 2011, 120 (08): : 529 - 534
  • [9] A longitudinal study of electrical stimulation levels and electrode impedance in children using the Clarion cochlear implant
    Henkin, Y
    Kaplan-Neeman, R
    Kronenberg, J
    Migirov, L
    Hildesheimer, M
    Muchnik, C
    ACTA OTO-LARYNGOLOGICA, 2006, 126 (06) : 581 - 586
  • [10] Devellopment of a steerable cochlear implant electrode array
    Chen, B. K.
    Kha, H. N.
    Clark, G. M.
    3RD KUALA LUMPUR INTERNATIONAL CONFERENCE ON BIOMEDICAL ENGINEERING 2006, 2007, 15 : 607 - +