Effect of early activation of cochlear implant on electrode impedance in pediatric population

被引:17
作者
Alhabib, Salman F. [1 ]
Abdelsamad, Yassin [2 ]
Yousef, Medhat [1 ,3 ]
Alzhrani, Farid [1 ]
Hagr, Abdulrahman [1 ]
机构
[1] King Saud Univ, Coll Med, King Abdullah Ear Specialist Ctr KAESC, Riyadh, Saudi Arabia
[2] MED El GmbH, Res Dept, Riyadh, Saudi Arabia
[3] Menoufia Univ Hosp, Fac Med, Audiol Unit, Menoufia, Egypt
关键词
One-day activation; Electrode impedance; Fibrosis; Minimal invasive cochlear implantation technique; SAFETY;
D O I
10.1016/j.ijporl.2020.110543
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
Objectives: To assess the evolution of electrode impedance after the early fitting of audio processors (activation after one-day) and classical fitting (activation after one-month) over an up-to-one year after cochlear implant (CI) surgery. Methods: A retrospective cohort study on Fifty-two CI recipients divided into two groups. The study group included 24 recipients (40 ears) who underwent early fitting, whereas the control group contained 28 recipients (40 ears) who underwent classical fitting. The electrode impedance was recorded during the surgery, switch-on session and at one, three, six, and twelve-months after the surgery. Results: In the study group, electrode impedance values obtained intraoperatively and at switch-on and one, three, six, and twelve-months were 4.89, 3.69, 6.52, 6.24, 6.05, and 5.81 K Omega, respectively, and only the switch-on and one-month values were significantly different (p < 0.0001). In the control group, electrode impedance values obtained intraoperatively and at switch-on and one, three, six, and twelve-months were 4.71, 7.19, 6.40, 6.05, and 5.73 K Omega, respectively. Thus, the electrode impedance value at switch-on was 52.65% (p < 0.001) greater than it intraoperatively. For both groups, the electrode impedance value at twelve-months was significantly higher than the respective intraoperative values (study group: 18.6% higher, P = 0.04; control group: 21.65% higher, P = 0.0001). Conclusion: Electrode impedance was significantly lower in the study group compared to the control group until one month after the surgery. However, the electrode impedance at twelve-months after the CI was similar in both groups.
引用
收藏
页数:6
相关论文
共 27 条
[1]   Cochlear Implant Device Activation and Programming: 5 Days Postimplantation [J].
Alsabellha, Rabea M. ;
Hagr, Abdulrahman ;
Al-momani, Murad O. ;
Garadat, Soha N. .
OTOLOGY & NEUROTOLOGY, 2014, 35 (04) :E130-E134
[2]  
Busby P A, 2002, Cochlear Implants Int, V3, P87, DOI 10.1179/cim.2002.3.2.87
[3]   Electrode Impedance Fluctuations as a Biomarker for Inner Ear Pathology After Cochlear Implantation [J].
Choi, June ;
Payne, Matthew R. ;
Campbell, Luke J. ;
Bester, Christo W. ;
Newbold, Carrie ;
Eastwood, Hayden ;
O'Leary, Stephen J. .
OTOLOGY & NEUROTOLOGY, 2017, 38 (10) :1433-1439
[4]  
Clark G M, 1995, Ann Otol Rhinol Laryngol Suppl, V166, P40
[5]   Electrical impedance spectroscopy study of biological tissues [J].
Dean, D. A. ;
Ramanathan, T. ;
Machado, D. ;
Sundararajan, R. .
JOURNAL OF ELECTROSTATICS, 2008, 66 (3-4) :165-177
[6]   LONG-TERM MEASURES OF ELECTRODE IMPEDANCE AND AUDITORY-THRESHOLDS FOR THE INERAID COCHLEAR IMPLANT [J].
DORMAN, MF ;
SMITH, LM ;
DANKOWSKI, K ;
MCCANDLESS, G ;
PARKIN, JL .
JOURNAL OF SPEECH AND HEARING RESEARCH, 1992, 35 (05) :1126-1130
[7]   Electrode impedance changes over time in MED El cochlear implant children recipients: Relation to stimulation levels and behavioral measures [J].
Fayed, Eman Asaad ;
Zaghloul, Hesham Saad ;
Morgan, Ashraf Elsayed .
COCHLEAR IMPLANTS INTERNATIONAL, 2020, 21 (04) :192-197
[8]   Feasibility of one-day activation in cochlear implant recipients [J].
Hagr, Abdulrahman ;
Garadat, Soha N. ;
Al-Momani, Murad ;
Alsabellha, Rabea M. ;
Almuhawas, Fida A. .
INTERNATIONAL JOURNAL OF AUDIOLOGY, 2015, 54 (05) :323-328
[9]   Financial Benefits of the Early Fitting of a Cochlear Implant Speech Processor: Assessment of the Direct Cost [J].
Hajr, Eman A. ;
Almuhawas, Fida A. .
CUREUS JOURNAL OF MEDICAL SCIENCE, 2019, 11 (09)
[10]   Histopathology of the peripheral vestibular system after cochlear implantation in the human [J].
Handzel, O ;
Burgess, BJ ;
Nadol, JB .
OTOLOGY & NEUROTOLOGY, 2006, 27 (01) :57-64