Flexible Distributed Pressure Sensing Strip for a Urethral Catheter

被引:10
作者
Ahmadi, Mahdi [1 ]
Rajamani, Rajesh [1 ]
Timm, Gerald [2 ]
Sezen, A. Serdar [1 ]
机构
[1] Univ Minnesota, Dept Mech Engn, St Paul, MN 55455 USA
[2] Univ Minnesota, Dept Urol, St Paul, MN 55455 USA
基金
美国国家卫生研究院;
关键词
Instrumented catheter; urethral sensors; in vivo sensors; catheter pressure sensors; in vivo force sensors; parasitic capacitance; capacitive sensors; URINARY-INCONTINENCE; PREVALENCE; EXERCISE; OLDER;
D O I
10.1109/JMEMS.2015.2444992
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A multi-sensor flexible strip is developed for a urethral catheter to measure distributed pressure in a human urethra. The developed sensor strip has important clinical applications in urodynamic testing for analyzing the causes of urinary incontinence in patients. There are two major challenges in the development of the sensor. First, a highly sensitive sensor strip that is flexible enough for urethral insertion into a human body is required. Second, the sensor has to work reliably in a liquid in vivo environment in the human body. Capacitive force sensors are designed and micro-fabricated using polyimide/polydimethylsiloxane (PDMS) substrates and copper electrodes. To remove the parasitic influence of urethral tissues, which create fringe capacitance that can lead to significant errors, a reference fringe capacitance measurement sensor is incorporated on the strip. The sensing strip is embedded on a catheter and experimental in vitro evaluation is presented using a bench-top pressure chamber. The sensors on the strip are able to provide the required sensitivity and the range. Preliminary experimental results also show promise that by using measurements from the reference parasitic sensor on the strip, the influence of parasitics from human tissue on the pressure measurements can be removed. [2014-0316]
引用
收藏
页码:1840 / 1847
页数:8
相关论文
共 25 条
[1]  
Abrams P., 1990, INT UROGYNEC J, V1, P45, DOI 10.1007/BF00373608
[2]  
Abrams P., 1997, Urodynamics
[3]  
Bo K, 2004, SPORTS MED, V34, P451
[4]   Implantable micromechanical parylene-based pressure sensors for unpowered intraocular pressure sensing [J].
Chen, Po-Jui ;
Rodger, Damien C. ;
Agrawal, Rajat ;
Saati, Saloomeh ;
Meng, Ellis ;
Varma, Rohit ;
Humayun, Mark S. ;
Tai, Yu-Chong .
JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 2007, 17 (10) :1931-1938
[5]   Prevalence of urinary incontinence among Spanish older people living at home [J].
Damián, J ;
Martín-Moreno, JM ;
Lobo, F ;
Bonache, J ;
Cerviño, J ;
Redondo-Márquez, L ;
Martínez-Agulló, E .
EUROPEAN UROLOGY, 1998, 34 (04) :333-338
[6]  
DIOKNO AC, 1986, J UROLOGY, V136, P1022
[7]   URINARY-INCONTINENCE - PREVALENCE AND NEEDS [J].
FENELEY, RCL ;
SHEPHERD, AM ;
POWELL, PH ;
BLANNIN, J .
BRITISH JOURNAL OF UROLOGY, 1979, 51 (06) :493-496
[8]  
Giovanni M.D., 1982, Flat and corrugated diaphgram design handbook
[9]   AMBULATORY MONITORING OF BLADDER AND DETRUSOR PRESSURE DURING NATURAL FILLING [J].
GRIFFITHS, CJ ;
ASSI, MS ;
STYLES, RA ;
RAMSDEN, PD ;
NEAL, DE .
JOURNAL OF UROLOGY, 1989, 142 (03) :780-784
[10]   PREVALENCE AND INCIDENCE OF URINARY-INCONTINENCE IN COMMUNITY-DWELLING POPULATIONS [J].
HERZOG, AR ;
FULTZ, NH .
JOURNAL OF THE AMERICAN GERIATRICS SOCIETY, 1990, 38 (03) :273-281