Response of a nanoDot OSLD system in megavoltage photon beams

被引:9
作者
Hoshida, Kento [1 ]
Araki, Fujio [2 ]
Ohno, Takeshi [2 ]
Kobayashi, Ikuo [3 ]
机构
[1] Kumamoto Univ, Grad Sch Hlth Sci, 4-24-1 Kuhonji, Kumamoto 8620976, Japan
[2] Kumamoto Univ, Fac Life Sci, Dept Hlth Sci, 4-24-1 Kuhonji, Kumamoto 8620976, Japan
[3] Nagase Landauer Ltd, Block C22-1, Tsukuba, Ibaraki 3002686, Japan
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2019年 / 64卷
关键词
nanoDot optically stimulated luminescence dosimeter (OSLD); Megavoltage photon beams; Energy response; Monte Carlo simulation; HIGH-ENERGY PHOTON; STIMULATED LUMINESCENT DETECTOR; ANGULAR-DEPENDENCE; DOSIMETRY SYSTEMS; DOSE-RESPONSE; ELECTRON; AL2O3-C; AUDITS;
D O I
10.1016/j.ejmp.2019.06.014
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The aim of this study was to investigate the response of a nanoDot optically stimulated luminescence dosimeter (OSLD) system in megavoltage photon beams. Methods: The nanoDot response was compared with the ionization chamber measurements for 4-18-MV photons in a plastic water phantom. The response was also calculated by the Monte Carlo method. In addition, the perturbation correction factor, P-Q, in the nanoDot cavity was calculated according to the Burlin's cavity theory. The angular dependence of the nanoDot was evaluated using a spherical phantom. Results: The calculated and measured nanoDot responses at a 10-cm depth and 10 x 10-cm(2) field were in agreement within 1% for 4-18-MV. The response increased by 3% at a 20 x 20-cm(2) field for the lower energy of 4 MV; however, it was constant within +/- 1% for 6-18 MV. The response was in a range from 1.0 to 0.99 for mean photon energy of more than 1.0 MeV but it increased with less than the 1.0 MeV. P-Q for the nanoDot cavity was approximately constant at 0.96-0.97 for greater than and equal to 10 MV. The angular dependence decreased by 5% and 3% for 6 and 15 MV, respectively. Conclusions: The nanoDot was energy-independent in megavoltage photon beams.
引用
收藏
页码:74 / 80
页数:7
相关论文
共 27 条
  • [1] TLD and OSLD dosimetry systems for remote audits of radiotherapy external beam calibration
    Alvarez, P.
    Kry, S. F.
    Stingo, F.
    Followill, D.
    [J]. RADIATION MEASUREMENTS, 2017, 106 : 412 - 415
  • [2] [Anonymous], 2000, TECHNICAL REPORT SER
  • [3] The response of a radiophotoluminescent glass dosimeter in megavoltage photon and electron beams
    Araki, Fujio
    Ohno, Takeshi
    [J]. MEDICAL PHYSICS, 2014, 41 (12)
  • [4] Berger MJ, 1987, NBSIR 87 3597 XCOM P, P20899
  • [5] A GENERAL THEORY OF CAVITY IONISATION
    BURLIN, TE
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1966, 39 (466) : 727 - &
  • [6] Commissioning of optically stimulated luminescence dosimeters for use in radiotherapy
    Dunn, Leon
    Lye, Jessica
    Kenny, John
    Lehmann, Joerg
    Williams, Ivan
    Kron, Tomas
    [J]. RADIATION MEASUREMENTS, 2013, 51-52 : 31 - 39
  • [7] Angular dependence of dose sensitivity of nanoDot optically stimulated luminescent dosimeters in different radiation geometries
    Jursinic, Paul A.
    [J]. MEDICAL PHYSICS, 2015, 42 (10) : 5633 - 5641
  • [8] Kawrakow I., 2018, PIRS898 NAT RES COUN
  • [9] Angular dependence of the nanoDot OSL dosimeter
    Kerns, James R.
    Kry, Stephen F.
    Sahoo, Narayan
    Followill, David S.
    Ibbott, Geoffrey S.
    [J]. MEDICAL PHYSICS, 2011, 38 (07) : 3955 - 3962
  • [10] Dose response of commercially available optically stimulated luminescent detector, AL2O3:C for megavoltage photons and electrons
    Kim, Dong Wook
    Chung, Weon Kuu
    Shin, Dong Oh
    Yoon, Myonggeun
    Hwang, Ui-Jung
    Rah, Jeong-Eun
    Jeong, Hojin
    Lee, Sang Yeob
    Shin, Dongho
    Lee, Se Byeong
    Park, Sung Yong
    [J]. RADIATION PROTECTION DOSIMETRY, 2012, 149 (02) : 101 - 108