System-specific evaluation of the dual flip angle MRI technique for quantitative T1 measurement

被引:0
作者
Yee, Seonghwan [1 ]
Fadell, Michael [1 ,2 ]
机构
[1] Univ Colorado, Sch Med, Dept Radiol, Anschutz Med Campus, Aurora, CO 80045 USA
[2] Childrens Hosp Colorado, 13123 East 16th Ave, Aurora, CO 80045 USA
关键词
calibration; dual flip angle; T-1; mapping; variable flip angle; GADOLINIUM-ENHANCED MRI; CARTILAGE; ACCURACY; QUANTIFICATION; ACQUISITION; PRECISION; FAILURE; B-1;
D O I
10.1002/mp.14864
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To investigate if the accuracy of the dual flip angle (DFA) technique for T-1 measurement is affected by the system-specific RF excitation performance. Methods: A T-1 phantom, made of 12 vials of unique T-1 value ranging approximately from 200 ms to 2000 ms, was built and tested on seven different clinical scanners. For each experiment, the reference T-1 of each vial was obtained by the inversion recovery-based technique, and the DFA technique was applied repeatedly with several flip angle (FA) pairs conventionally proposed as optimal. The accuracy of the DFA technique for each FA pair was then evaluated by comparing the measured T-1 values for the vials to the references. Any variation of the accuracy was then evaluated across different FA pairs, and across different MRI systems. To improve accuracy with a selected FA pair, the signal ratio (SR) curve, obtained from the phantom, was utilized in a calibration strategy of the DFA technique. Results: When combined for all the vials, the average ratio of the measured T-1 to the reference generally increased as the FA pair window gradually slid from the smaller to the larger FA values. Furthermore, among several optimal FA pairs, the pair of the best accuracy varied slightly by the MRI system. The accuracy for any FA pair could be improved when the calibration strategy was utilized. Conclusions: The RF excitation performance may vary by the specific FA pair and by the specific MRI system, influencing the accuracy of the DFA technique. The system-specific evaluation, and, if needed, its calibration, would help improve the accuracy of the DFA technique. (C) 2021 American Association of Physicists in Medicine.
引用
收藏
页码:2790 / 2799
页数:10
相关论文
共 25 条
[1]   Preoperative planning for redirective, periacetabular osteotomies [J].
Albers, Christoph E. ;
Rogers, Piet ;
Wambeek, Nicholas ;
Ahmad, Sufian S. ;
Yates, Piers J. ;
Prosser, Gareth H. .
JOURNAL OF HIP PRESERVATION SURGERY, 2017, 4 (04) :276-288
[2]   T1 mapping with the variable flip angle technique: A simple correction for insufficient spoiling of transverse magnetization [J].
Baudrexel, Simon ;
Noeth, Ulrike ;
Schuere, Jan-Ruediger ;
Deichmann, Ralf .
MAGNETIC RESONANCE IN MEDICINE, 2018, 79 (06) :3082-3092
[3]   Delayed gadolinium-enhanced magnetic resonance imaging of cartilage to predict early failure of Bernese periacetabular osteotomy for hip dysplasia [J].
Cunningham, Torin ;
Jessel, Rebecca ;
Zurakowski, David ;
Millis, Michael B. ;
Kim, Young-Jo .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2006, 88A (07) :1540-1548
[4]   Determining and optimizing the precision of quantitative measurements of perfusion from dynamic contrast enhanced MRI [J].
Dale, BM ;
Jesberger, JA ;
Lewin, JS ;
Hillenbrand, CM ;
Duerk, JL .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2003, 18 (05) :575-584
[5]   Rapid combined T1 and T2 mapping using gradient recalled acquisition in the steady state [J].
Deoni, SCL ;
Rutt, BK ;
Peters, TM .
MAGNETIC RESONANCE IN MEDICINE, 2003, 49 (03) :515-526
[6]   Is the Damage of Cartilage a Global or Localized Phenomenon in Hip Dysplasia, Measured by dGEMRIC? [J].
Hingsammer, Andreas ;
Chan, Jenny ;
Kalish, Leslie A. ;
Mamisch, Tallal C. ;
Kim, Young-Jo .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2013, 471 (01) :301-307
[7]  
Keenan K.E., 2016, Proc. Intl. Soc. Mag. Res. Med, P3290
[8]   T1-mapping in the heart: accuracy and precision [J].
Kellman, Peter ;
Hansen, Michael S. .
JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE, 2014, 16
[9]   Models and methods for analyzing DCE-MRI: A review [J].
Khalifa, Fahmi ;
Soliman, Ahmed ;
El-Baz, Ayman ;
Abou El-Ghar, Mohamed ;
El-Diasty, Tarek ;
Gimel'farb, Georgy ;
Ouseph, Rosemary ;
Dwyer, Amy C. .
MEDICAL PHYSICS, 2014, 41 (12)
[10]   Anterior Delayed Gadolinium-enhanced MRI of Cartilage Values Predict Joint Failure After Periacetabular Osteotomy [J].
Kim, Sang Do ;
Jessel, Rebecca ;
Zurakowski, David ;
Millis, Michael B. ;
Kim, Young-Jo .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2012, 470 (12) :3332-3341