Evaluation of an electrical impedance tomography-based global inhomogeneity index for pulmonary ventilation distribution

被引:220
|
作者
Zhao, Zhanqi [1 ,2 ,3 ]
Moeller, Knut [3 ]
Steinmann, Daniel [2 ]
Frerichs, Inez [4 ]
Guttmann, Josef [2 ]
机构
[1] Anaesthesiol Univ Klin Freiburg, Sekt Expt Anaesthesiol, D-79106 Freiburg, Germany
[2] Univ Med Ctr, Dept Anesthesiol & Crit Care Med, Sect Expt Anesthesiol, Freiburg, Germany
[3] Furtwangen Univ, Dept Biomed Engn, Villingen Schwenningen, Germany
[4] Univ Med Ctr Schleswig Holstein, Dept Anesthesiol & Intens Care Med, Kiel, Germany
关键词
Inhomogeneity index; Electrical impedance tomography; Pulmonary ventilation distribution; Ventilation therapy; END-EXPIRATORY PRESSURE; TIDAL VOLUMES; LUNG-TISSUE; RECRUITMENT; AIR; DERECRUITMENT; CURVES; MODEL;
D O I
10.1007/s00134-009-1589-y
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
To evaluate the plausibility, stability, and interindividual comparability of the global inhomogeneity index (GI) based on electrical impedance tomography (EIT). The lung area in an EIT image was identified by using the lung area estimation method, which mirrors the lung regions in the functional EIT image and subsequently subtracts the cardiac-related areas. The tidal EIT image, showing the difference in impedances between end-inspiration and end-expiration, was calculated and the variations in its pixel values within the predefined lung area were then used as an indicator of inhomogeneous ventilation (the GI index). Fifty patients were investigated including 40 patients tracheally intubated with double-lumen tubes (test group) and 10 patients under anesthesia without pulmonary disease (control group). Positive end-expiratory pressure (PEEP) of 5 mbar was applied in the test group during both two-lung ventilation (TLV) and subsequent one-lung ventilation (OLV). The patients of the control group were ventilated without PEEP. EIT data were recorded in both groups. A significantly lower GI value was found in the control group (0.40 +/- A 0.05, P = 0.025 vs. TLV 0.74 +/- A 0.47 and P < 0.002 vs. OLV 1.51 +/- A 1.45). A significant difference was also found in the test group between TLV and OLV (P < 0.002). If GI was calculated only in the ventilated lung area during OLV (0.71 +/- A 0.32), it did not significantly differ from the test group during TLV. The GI index quantifies the gas distribution in the lung with a single number and reveals good interpatient comparability.
引用
收藏
页码:1900 / 1906
页数:7
相关论文
共 50 条
  • [41] Electrical impedance tomography: the holy grail of ventilation and perfusion monitoring?
    Leonhardt, Steffen
    Lachmann, Burkhard
    INTENSIVE CARE MEDICINE, 2012, 38 (12) : 1917 - 1929
  • [42] Electrical impedance tomography guided ventilation therapy
    Putensen, Christian
    Wrigge, Hermann
    Zinserling, Joerg
    CURRENT OPINION IN CRITICAL CARE, 2007, 13 (03) : 344 - 350
  • [43] Regional ventilation in cystic fibrosis measured by electrical impedance tomography
    Zhao, Zhanqi
    Fischer, Rainald
    Frerichs, Inez
    Mueller-Lisse, Ullrich
    Moeller, Knut
    JOURNAL OF CYSTIC FIBROSIS, 2012, 11 (05) : 412 - 418
  • [44] Impact of PEEP during laparoscopic surgery on early postoperative ventilation distribution visualized by electrical impedance tomography
    Karsten, J.
    Heinze, H.
    Meier, T.
    MINERVA ANESTESIOLOGICA, 2014, 80 (02) : 158 - 166
  • [45] Electrical Impedance Tomography for Cardio-Pulmonary Monitoring
    Putensen, Christian
    Hentze, Benjamin
    Muenster, Stefan
    Muders, Thomas
    JOURNAL OF CLINICAL MEDICINE, 2019, 8 (08)
  • [46] Assessment of Regional Ventilation Distribution: Comparison of Vibration Response Imaging (VRI) with Electrical Impedance Tomography (EIT)
    Shi, Chang
    Boehme, Stefan
    Bentley, Alexander H.
    Hartmann, Erik K.
    Klein, Klaus U.
    Bodenstein, Marc
    Baumgardner, James E.
    David, Matthias
    Ullrich, Roman
    Markstaller, Klaus
    PLOS ONE, 2014, 9 (01):
  • [47] Design and Implementation of a System of Pulmonary Electrical Impedance Tomography
    Tolabin, D. E.
    Gomez Buxo, M. D.
    Rossi, R.
    Molina, G. R.
    2016 IEEE BIENNIAL CONGRESS OF ARGENTINA (ARGENCON), 2016,
  • [48] The time taken for the regional distribution of ventilation to stabilise: an investigation using electrical impedance tomography
    Caruana, L. R.
    Paratz, J. D.
    Chang, A.
    Barnett, A. G.
    Fraser, J. F.
    ANAESTHESIA AND INTENSIVE CARE, 2015, 43 (01) : 88 - 91
  • [49] Distribution of ventilation in pregnant Shetland ponies Shetland ponies measured by Electrical Impedance Tomography
    Schramel, Johannes
    Nagel, Christina
    Auer, Ulrike
    Palm, Franziska
    Aurich, Christine
    Moens, Yves
    RESPIRATORY PHYSIOLOGY & NEUROBIOLOGY, 2012, 180 (2-3) : 258 - 262
  • [50] Complementary use of priors for pulmonary imaging with electrical impedance and ultrasound computed tomography
    Alsaker, Melody
    Cardona Cardenas, Diego Armando
    Furuie, Sergio Shiguemi
    Mueller, Jennifer L.
    JOURNAL OF COMPUTATIONAL AND APPLIED MATHEMATICS, 2021, 395