Comparison of Pumps and Oxygenators With Pulsatile and Nonpulsatile Modes in an Infant Cardiopulmonary Bypass Model

被引:13
作者
Haines, Nikkole M. [1 ]
Wang, Shigang [1 ]
Kunselman, Allen [1 ]
Myers, John L. [1 ,2 ]
Undar, Akif [1 ,2 ,3 ]
机构
[1] Penn State Hershey Childrens Hosp, Penn State Coll Med, Penn State Milton S Hershey Med Ctr, Pediat Cardiac Res Labs,Dept Pediat, Hershey, PA USA
[2] Penn State Hershey Childrens Hosp, Penn State Coll Med, Penn State Milton S Hershey Med Ctr, Dept Surg, Hershey, PA USA
[3] Penn State Hershey Childrens Hosp, Penn State Coll Med, Penn State Milton S Hershey Med Ctr, Dept Bioengn, Hershey, PA USA
关键词
Cardiopulmonary bypass; Pediatric blood pumps; Pediatric oxygenators; Pulsatile flow; VITAL ORGAN RECOVERY; GASEOUS MICROEMBOLI; HEART-SURGERY; PERFUSION; CANNULAE; DELIVERY;
D O I
10.1111/j.1525-1594.2009.00911.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
As the evidence mounts in favor of pulsatile perfusion during CPB, it is necessary to investigate the effect of circuit components on the quality of pulsatility delivered throughout the circuit. We compared two bloodpumps, the Jostra HL-20 heart-lung machine and the MEDOS DELTASTREAM DP1 Bloodpump, and two oxygenators, the Capiox Baby RX05 and the MEDOS HILITE 800LT, in terms of mean arterial pressure, energy equivalent pressure, surplus hemodynamic energy, total hemodynamic energy, and pressure drop over the oxygenators using a blood analog. The pumps and oxygenators were combined in unique circuits and tested in nonpulsatile and pulsatile modes, at two flow rates (500 and 800 mL/min), and three rotational speed differentials when using the MEDOS DELTASTREAM DP1 Bloodpump for 144 trials in total. The Jostra Roller pump produced some pulsatility in nonpulsatile mode and better pulsatility in pulsatile mode than the MEDOS DP1 Bloodpump at a rotational speed differential of 2500 rpm, but not at 3500 or 4500 rpm. The MEDOS DP1 Bloodpump produced almost no pulsatility in nonpulsatile mode. Pressure drops over the Capiox Baby RX05 were markedly higher, at 92.5 +/- 0.4 mm Hg with the MEDOS DP1 Bloodpump at 800 mL/min and 4500 rpm in pulsatile mode, than those of the MEDOS HILITE 800LT oxygenator, which was 67.0 +/- 0.1 mm Hg at the same settings. These results suggest that careful selection of each circuit component, based on the individual clinical case and component specifics, are necessary to achieve the best quality of pulsatility.
引用
收藏
页码:993 / 1001
页数:9
相关论文
共 14 条
[1]   Benefits of pulsatile perfusion on vital organ recovery during and after pediatric open heart surgery [J].
Alkan, Tijen ;
Akcevin, Atif ;
Uendar, Akif ;
Tuerkoglu, Halil ;
Paker, Tufan ;
Aytac, Aydin .
ASAIO JOURNAL, 2007, 53 (06) :651-654
[2]   Effects of pulsatile and nonpulsatile perfusion on vital organ recovery in pediatric heart surgery:: A pilot clinical study [J].
Alkan, Tijen ;
Akcevin, Atif ;
Uendar, Akif ;
Tuerkoglu, Halil ;
Paker, Tufan ;
Aytac, Aydin .
ASAIO JOURNAL, 2006, 52 (05) :530-535
[3]   Mechanism for cavitation in the mechanical heart valve with an artificial heart: Nuclei and viscosity dependence [J].
Lee, H ;
Taenaka, Y ;
Kitamura, S .
ARTIFICIAL ORGANS, 2005, 29 (01) :41-46
[4]  
Littell R.C., 2006, SAS SYSTEM MIXED MOD, V2nd
[5]  
*MED MED AG, INSTR US HILITE 800L
[6]   A performance evaluation of eight geometrically different 10 Fr pediatric arterial cannulae under pulsatile and nonpulsatile perfusion conditions in an infant cardiopulmonary bypass model [J].
Rider, Alan R. ;
Ji, Bingyang ;
Kunselman, Allen R. ;
Weiss, William J. ;
Myers, John L. ;
Uendar, Akif .
ASAIO JOURNAL, 2008, 54 (03) :306-315
[7]  
SHEPARD RB, 1966, ARCH SURG-CHICAGO, V93, P730
[8]  
*TER, INSTR US CAP BAB RX0
[9]   Vascular pulsatility in patients with a pulsatile- or continuous-flow ventricular assist device [J].
Travis, Adam R. ;
Giridharan, Guruprasad A. ;
Pantalos, George M. ;
Dowling, Robert D. ;
Prabhu, Sumanth D. ;
Slaughter, Mark S. ;
Sobieski, Mike ;
Undar, Akif ;
Farrar, David J. ;
Koenig, Steven C. .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2007, 133 (02) :517-524
[10]   Comparison of four different pediatric 10F aortic cannulae during Pulsatile versus nonpulsatile perfusion in a simulated neonatal model of cardiopulmonary bypass [J].
Uendar, Akif ;
Ji, Bingyang ;
Rider, Alan ;
Lukic, Branka ;
Kunselman, Allen R. ;
Weiss, William J. ;
Myers, John L. .
ASAIO JOURNAL, 2007, 53 (06) :778-784