A national survey of computed tomography doses in hybrid PET-CT and SPECT-CT examinations in the UK

被引:52
作者
Iball, Gareth R. [1 ]
Bebbington, Natalie A. [2 ,3 ]
Burniston, Maria [4 ]
Edyvean, Sue [5 ]
Fraser, Louise [6 ]
Julyan, Peter [7 ]
Parkar, Nasreen [6 ]
Wood, Tim [8 ,9 ]
机构
[1] Leeds Teaching Hosp NHS Trust, Leeds Gen Infirm, Med Phys Dept, Old Med Sch, Leeds LS1 3EX, W Yorkshire, England
[2] Univ Hosp Birmingham NHS Fdn Trust, Queen Elizabeth Hosp Birmingham, Mindelsohn Way, Birmingham B15 2GW, W Midlands, England
[3] Siemens Healthineers, Borupvang 9, DK-2750 Ballerup, Denmark
[4] Royal Free London NHS Fdn Trust, Pond St, London NW3 2QG, England
[5] Publ Hlth England, Med Radiat Dosimetry Grp, Ctr Radiat Chem & Environm Hazards CRCE, Didcot OX11 0RQ, Oxon, England
[6] Publ Hlth England, ARSAC, Med Exposures Grp, Radiat Assessments Dept,Ctr Radiat Chem & Environ, Didcot OX11 0RQ, Oxon, England
[7] Christie NHS Fdn Trust, Christie Med Phys & Engn, Manchester M20 4BX, Lancs, England
[8] Hull & East Yorkshire Hosp NHS Trust, Castle Hill Hosp, Queens Ctr Oncol & Haematol, Radiat Phys Dept, Kingston Upon Hull HU16 5JQ, N Humberside, England
[9] Univ Hull, Fac Sci, Cottingham Rd, Kingston Upon Hull HU6 7RX, N Humberside, England
关键词
computed tomography; diagnostic reference levels; dosimetry; positron emission tomography; single photon emission computed tomography; SCANS;
D O I
10.1097/MNM.0000000000000672
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
ObjectivesThe aim of this study was to conduct a nationwide survey of computed tomography (CT) doses for a wide range of PET-CT and single photon emission computed tomography-computed tomography (SPECT-CT) imaging procedures, with the aim of generating proposed UK national diagnostic reference levels (NDRLs).MethodsCT protocol and dosimetry data for three PET-CT and seven SPECT-CT examinations were gathered from centres across the UK. Data were divided according to CT purpose (attenuation correction, localization or diagnostic) and third quartile values of scanner average dose metrics were used to generate suggested NDRLs for a range of examination and CT purpose combinations. Achievable doses were also established from the median of the dose distributions.ResultsData were obtained from 47 centres, allowing suggested NDRLs to be produced for fluorine-18-fluorodeoxyglucose half-body PET-CT, and parathyroid, post-thyroid ablation, meta-iodobenzylguanidine/octreotide, cardiac and bone SPECT-CT examinations.Variations in dose of up to a factor of 35 were observed for a given examination/CT purpose combination. For fluorine-18-fluorodeoxyglucose half-body PET-CT examination dose levels for the three CT purposes overlapped, which highlights the variability in the way in which CT purposes are interpreted across the UK. This lack of standardization is believed to be the largest contributor to the dose variations that were observed. The survey highlighted the need for targeted optimization work in many centres.ConclusionSuggested UK NDRLs and achievable doses for six common PET-CT and SPECT-CT examinations have been established as a result of this study.
引用
收藏
页码:459 / 470
页数:12
相关论文
共 32 条
[1]  
Alessio A, 2015, J NUCL MED, V56
[2]  
[Anonymous], 1997, OFF J EUR COMMUN, pL180
[3]  
[Anonymous], ANN ICRP
[4]  
[Anonymous], 2008, IAEATEDDOC1597
[5]  
[Anonymous], 1990, ANN ICRP, DOI DOI 10.1016/0146-6453(91)90066-P
[6]   PATIENT DOSES FROM HYBRID SPECT-CT PROCEDURES [J].
Avramova-Cholakova, S. ;
Dimcheva, M. ;
Petrova, E. ;
Garcheva, M. ;
Dimitrova, M. ;
Palashev, Y. ;
Vassileva, J. .
RADIATION PROTECTION DOSIMETRY, 2015, 165 (1-4) :424-429
[7]   PATIENT DOSES FROM PET-CT PROCEDURES [J].
Avramova-Cholakova, S. ;
Ivanova, S. ;
Petrova, E. ;
Garcheva, M. ;
Vassileva, J. .
RADIATION PROTECTION DOSIMETRY, 2015, 165 (1-4) :430-433
[8]  
Beyer T, 2005, J NUCL MED, V46, P429
[9]   FDG PET/CT: EANM procedure guidelines for tumour imaging: version 2.0 [J].
Boellaard, Ronald ;
Delgado-Bolton, Roberto ;
Oyen, Wim J. G. ;
Giammarile, Francesco ;
Tatsch, Klaus ;
Eschner, Wolfgang ;
Verzijlbergen, Fred J. ;
Barrington, Sally F. ;
Pike, Lucy C. ;
Weber, Wolfgang A. ;
Stroobants, Sigrid ;
Delbeke, Dominique ;
Donohoe, Kevin J. ;
Holbrook, Scott ;
Graham, Michael M. ;
Testanera, Giorgio ;
Hoekstra, Otto S. ;
Zijlstra, Josee ;
Visser, Eric ;
Hoekstra, Corneline J. ;
Pruim, Jan ;
Willemsen, Antoon ;
Arends, Bertjan ;
Kotzerke, Joerg ;
Bockisch, Andreas ;
Beyer, Thomas ;
Chiti, Arturo ;
Krause, Bernd J. .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2015, 42 (02) :328-354
[10]  
Boone JM, 2011, SIZESPECIFIC DOSE ES